Usability-validation subjects
61 subjects
- EndpointAdult TTE acquisition comparison
- PopulationUltraSight AI Guidance
- Tested versionUltraSight AI Guidance
FDA-cleared profile
UltraSight Inc. · Ultrasound · FDA-cleared (510(k))
Authorized purpose: UltraSight AI Guidance assists non-expert healthcare professionals with acquisition of ten standard adult two-dimensional transthoracic echocardiography views by providing real-time transducer-position and orientation guidance on a compatible ultrasound system.
Overview status: Source-reviewed overview
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
Clinical takeaway. Clinical role: UltraSight AI Guidance assists non-expert healthcare professionals with acquisition of ten standard adult two-dimensional transthoracic echocardiography views by providing real-time transducer-position and orientation guidance on a compatible ultrasound system.
Evidence takeaway. Evidence from the reviewed source record: K223347 reports summative usability validation in which 61 subjects were scanned twice, comparing non-expert acquisition with AI guidance against cardiac-sonographer acquisition without AI guidance. This is an acquisition-guidance and usability evaluation, not evidence that the system independently diagnoses cardiac disease.
Evidence context not reviewed for this FDA record.
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.
The exact FDA indication identifies cardiac ultrasound images and a compatible general-purpose diagnostic ultrasound system; it does not support the dossier's radiographic, fluoroscopic, CT, or procedure-input list.
The exact FDA device description states transducer position/orientation guidance, diagnostic-quality heart views, and ten standard cardiac views rather than the dossier's generic image, landmark, measurement, overlay, and navigation list.
The exact FDA indication and comparison table identify medical professionals and exclude expert sonographers.
The replacement retains only the packet's compatibility, adult ten-view scope, excluded-user, and clip-saving statements.
Evidence
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Sensitivity is not the primary endpoint for this system, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.
Specificity is not the primary endpoint for this system, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.
61 subjects
K223347
Manufacturer evidence submitted for FDA clearance or approval; local clinical, technical, and workflow acceptance remains necessary.
2 unlinked candidate evidence records are retained for source review and excluded from this evidence display.
Safety and lifecycle
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Exact-identifier review does not prove absence of all product-family safety signals.
Price not public here; enter your quote
Not publicly available: Reviewed sources do not publish an exact-product procurement price with a stated offer scope.
No separate named-product payment identified in reviewed public material
Centers for Medicare & Medicaid Services · United StatesBuild a local total-cost and value model; do not infer reimbursement from FDA clearance or approval.
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Source context, not route validation: Cardiac ultrasound images acquired using a compatible general-purpose diagnostic ultrasound system.
Scope: Exact FDA submission. The exact FDA indication identifies cardiac ultrasound images and a compatible general-purpose diagnostic ultrasound system; it does not support the dossier's radiographic, fluoroscopic, CT, or procedure-input list.
Confirm scanner/model, protocol, series completeness and both software builds.
Field checked 2026-09-19 · FDA K223347: UltraSight AI Guidance
Source context, not route validation: Validate compatible imaging hardware, acquisition geometry, displays, workstation, exported objects, latency, calibration, and fallback workflow.
Scope: Product family. Current vendor materials are product-family evidence for K223347; confirm the exact cleared version, enabled options, deployment, and local compatibility.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Field checked 2026-09-03 · FDA K223347: UltraSight AI Guidance · UltraSight AI Guidance
Source context, not route validation: Procedure-room hardware and software or a connected workstation with supported imaging equipment and accessories.
Scope: Product family. Current vendor materials are product-family evidence for K223347; confirm the exact cleared version, enabled options, deployment, and local compatibility.
Confirm processing location, queue ownership, exclusions and exact AI build.
Field checked 2026-09-03 · FDA K223347: UltraSight AI Guidance · UltraSight AI Guidance
Source context, not route validation: Dynamic real-time guidance on the position and orientation of the transducer to help non-expert users acquire diagnostic-quality tomographic views of the heart; guidance for ten standard cardiac views.
Scope: Exact FDA submission. The exact FDA device description states transducer position/orientation guidance, diagnostic-quality heart views, and ten standard cardiac views rather than the dossier's generic image, landmark, measurement, overlay, and navigation list.
Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.
Field checked 2026-09-19 · FDA K223347: UltraSight AI Guidance
Unresolved route: Dynamic real-time guidance on the position and orientation of the transducer to help non-expert users acquire diagnostic-quality tomographic views of the heart; guidance for ten standard cardiac views.
Scope: Exact FDA submission. The exact FDA device description states transducer position/orientation guidance, diagnostic-quality heart views, and ten standard cardiac views rather than the dossier's generic image, landmark, measurement, overlay, and navigation list.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Field checked 2026-09-19 · FDA K223347: UltraSight AI Guidance
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
No tests below have been performed or accepted.
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
Current vendor materials are product-family evidence for K223347; confirm the exact cleared version, enabled options, deployment, and local compatibility.
Current vendor materials are product-family evidence for K223347; confirm the exact cleared version, enabled options, deployment, and local compatibility.
Current vendor materials are product-family evidence for K223347; confirm the exact cleared version, enabled options, deployment, and local compatibility.
Map the installed model, build, options, population, anatomy, inputs, outputs, and enabled AI functions to K223347 before acceptance.
exact submission · checked 2026-09-03Later product-family features are not evidence that this cleared release includes them.
Validate compatible imaging hardware, acquisition geometry, displays, workstation, exported objects, latency, calibration, and fallback workflow.
product family · checked 2026-09-03Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally. Vendor statements are current product-family context; confirm the exact cleared release and installed options.
Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Guidance supports image acquisition and does not replace a qualified cardiac examination, image-quality review, interpretation, or clinical decision. Confirm compatible ultrasound host, ten-view scope, adult population, and fallback workflow.
exact submission · checked 2026-09-03Define stop-use, escalation, rollback, and revalidation triggers before production use.
Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.
product family · checked 2026-09-03No exact-product price was found in the reviewed public material; vendor productivity claims are scenario inputs, not guaranteed local savings.
Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Guidance supports image acquisition and does not replace a qualified cardiac examination, image-quality review, interpretation, or clinical decision. Confirm compatible ultrasound host, ten-view scope, adult population, and fallback workflow.
product family · checked 2026-09-03Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. Vendor statements are current product-family context; confirm the exact cleared release and installed options.
Regulatory identity
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.
Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.
Research record
This audit distinguishes completed source review from fields that have not yet been researched.
The refreshed discovery snapshot contains 0 PubMed and 3 ClinicalTrials.gov candidate records. Discovery metadata alone does not establish exact tested-product, version, population, and endpoint applicability, so all candidates remain quarantined and none was promoted. The current discovery snapshot contains 0 PubMed and 3 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, fda decision material, vendor product page, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, fda decision summary packet, literature index, trial registry, pubmed, clinical trials, openfda device recall
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.
Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.
10 sources - 3 history items
Scope: purpose, modality, anatomy, intendedUsers, input, output, limitations, evidenceSummary
Scope: regulatoryIdentity, modalityContext
Scope: evidenceDiscovery
Scope: evidenceDiscovery, studyStatus
Scope: coverage, codingContext
Scope: recallDiscovery, nativeSubmissionMatching
Scope: productFamilyContext, workflowRole, implementationCapabilities
Scope: evidenceStudies, productFamilyContext, evidenceDiscovery
Scope: evidenceStudies, productFamilyContext, evidenceDiscovery
Scope: evidenceStudies, productFamilyContext, evidenceDiscovery
Regulatory record
510(k)
Research record
fda record reviewed
Editorial review
Source-reviewed editorial overlay
Alternatives
Deski
Authorized purpose: Guides medical professionals acquiring ten standard 2D transthoracic echocardiographic views in adults requiring a cardiac ultrasound examination; acquired images require qualified clinical interpretation.
Ligence Uab
Ligence Heart is a fully automated software platform that processes, analyses and makes measurements on acquired transthoracic cardiac ultrasound images, automatically producing a full report with measurements of several key cardiac structural and functional parameters. The data produced by this software is intended to be used to support qualified cardiologists or sonographers for clinical decision making. Ligence Heart is indicated for use in adult patients. Ligence Heart has not been validated for the assessment of congenital heart disease, valve disease, pericardial disease, and/or intra-cardiac lesions (e.g., tumors, thrombi). Limitations: Poor image capture will lead to poor annotations and subsequent measurements. Multiple image quality algorithms are used to filter out images of poor quality. Our software complements good patient care and does not exempt the user from the responsibility to provide supervision, clinically review the patient, and make appropriate clinical decisions. If no gender is present, female referenced guideline values will be used for conclusions. If Body Surface Area (BSA) is not present, indexed values cannot be provided. During image acquisition, inappropriate use of the echo machine, use of non-cardiac ultrasound probes, or use of suboptimal settings (e.g., gain, contrast, depth) may lead to lower accuracy of the software. Ligence Heart has not been validated in patients with prosthetic or surgically altered cardiac valves, implantable intracardiac devices, or irregular (non-sinus) rhythm; these conditions may affect automated measurement accuracy.
Ultrasight, Ltd.
Authorized purpose: PVAD IQ supports qualified-physician clinical decision support by identifying the aortic annulus and PVAD inlet, calculating their mean distance, and identifying clips acceptable for this measurement.
Mycardium AI Limited
Authorized purpose: Transthoracic echocardiography view classification, cardiac structure segmentation, functional measurements and Doppler review and report support.
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Open existing correction destinationPermission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.
No entry has been added to this scoped change history for this record. This is not proof that the record never changed.
Correction policy and limits · Public repository change history