FDA-cleared profile

Radiomics App v1.0

Microsoft Corp.

Microsoft Radiomics App v1.0 is a software-only medical device for trained radiation oncologists, dosimetrists, and physicists to derive organ and tumor contours for radiation-treatment planning. It supports CT and MR images, image and contour review, contour creation and modification, transfer to treatment-planning systems, adaptive-therapy support, archiving, tumor and healthy-structure localization, image fusion, and three-dimensional rendering. Images reviewed with the software should not be used for primary image interpretations, and it is not for digital mammography.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
Microsoft Radiomics App v1.0 is a software-only medical device for trained radiation oncologists, dosimetrists, and physicists to derive organ and tumor contours for radiation-treatment planning. It supports CT and MR images, image and contour review, contour creation and modification, transfer to treatment-planning systems, adaptive-therapy support, archiving, tumor and healthy-structure localization, image fusion, and three-dimensional rendering. Images reviewed with the software should not be used for primary image interpretations, and it is not for digital mammography.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Intended users
Radiation oncologists, medical physicists, dosimetrists, therapists, and other qualified radiation professionals.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Care setting and population
Radiation-oncology settings within the labeled workflow.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Workflow role
Radiotherapy contouring, treatment planning, treatment management, image processing, or session support under qualified review and approval.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Required input
Supported CT or MR images, structures, prescriptions, treatment plans, registrations, and treatment-session data.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Output and human action
Contours, plans, adapted plans, image-processing results, treatment-management records, or session information requiring professional review.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Limitations
The exact labeling excludes primary image interpretation and digital mammography; generated contours require review and professional approval before treatment use.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K173420
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
Radiotherapy Planning Imaging
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Anatomy
CT and MR images; tumors and healthy anatomical structures for radiotherapy planning
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2017-12-27
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
Standards-based DICOM/PACS integration was identified in the FDA submission materials; compatibility with a specific PACS, RIS, EHR, or orchestrator requires confirmation.
  • Public source
  • Exact FDA submission
  • Checked 2026-08-13

Normalized from the prior exact-submission review. Confirm interface versions and local compatibility during technical evaluation.

Deployment and data flow
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-04

The exact public FDA decision material does not establish the complete deployment topology, hosting responsibility, availability, upgrade path, or service contract.

Security and privacy
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-04

Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.

Training and support
Only radiation oncologists, medical physicists, dosimetrists, therapists, and other qualified radiation professionals should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Monitoring and change control
Monitor K173420 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: The exact labeling excludes primary image interpretation and digital mammography; generated contours require review and professional approval before treatment use.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04

This is buyer-facing monitoring guidance derived from the exact cleared scope; it is not a claim that the FDA package provides a live monitoring service.

Release and scope verification

Map installed model, build, options, population, anatomy, inputs, outputs, and enabled functions to K173420 before acceptance.

exact submission · checked 2026-09-04

Later product-family features are not evidence that this cleared or authorized release includes them.

Deployment boundary

The exact public deployment topology remains incomplete; establish hosting responsibility, data flow, availability, upgrade, service, and downtime boundaries locally.

exact submission · checked 2026-09-04

FDA-described topology does not establish current hosting contracts, security controls, uptime, or local qualification.

Interoperability acceptance

The exact public integration contract remains incomplete; establish and test supported input, output, identity, routing, archive, and downtime behavior locally.

exact submission · checked 2026-09-04

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. The exact labeling excludes primary image interpretation and digital mammography; generated contours require review and professional approval before treatment use.

exact submission · checked 2026-09-04

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

exact submission · checked 2026-09-04

No exact-product price or reimbursement determination was promoted.

Lifecycle monitoring

Monitor K173420 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: The exact labeling excludes primary image interpretation and digital mammography; generated contours require review and professional approval before treatment use.

exact submission · checked 2026-09-04

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
K173420 FDA decision material identifies the exact cleared or authorized product, intended use, labeled anatomy and modality, and submission-specific verification. It does not by itself establish product-wide sensitivity, specificity, outcomes, reimbursement, or independent clinical validation for this exact configuration.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Reported sensitivity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-04

Sensitivity is not promoted as a product-wide endpoint. Compare only an exact endpoint, threshold, population, version, study design, and source if separately verified.

Reported specificity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-04

Specificity is not promoted as a product-wide endpoint. Compare only an exact endpoint, threshold, population, version, study design, and source if separately verified.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

K173420 FDA decision evidence package for Radiomics App v1.0

  • DesignFDA decision evidence and product verification described in the exact submission or PMA decision record.
  • PopulationCT and MR images; tumors and healthy anatomical structures for radiotherapy planning
  • Scopeexact submission
  • Tested versionRadiomics App v1.0 configuration described in K173420
  • IndependenceSponsor or vendor study

K173420

FDA decision material is manufacturer-submitted regulatory evidence. It is not independent clinical validation unless the exact source expressly establishes that distinction; local technical, clinical, privacy, security, and workflow acceptance remains necessary.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-04

No public list price for the exact cleared or authorized configuration was found. Obtain a written quote separating license or hardware basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

Reimbursement and coding
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-04

No exact-product reimbursement determination was promoted in this timeboxed review. Verify payer, site-of-service, coding, coverage, and local payment assumptions independently.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K173420 by exact native submission identifier. This does not exclude product-family, later-version, or adverse-event information.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-04

Exact-identifier review does not prove absence of all product-family safety signals.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-04Last searched
26Fields reviewed
7Source classes checked
0Awaiting review PubMed leads
0Awaiting review trial leads
0Unreviewed FDA recall leads

Recovered from the prior exact-submission extraction and normalized under the current provenance rules. Scope was limited to exact FDA decision material and exact-submission recall disposition already present in the checked snapshot. Optional literature, vendor, pricing, and deployment research was not promoted. Automated exact-name discovery found 0 PubMed and 0 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, fda decision material, fda device recall, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA 510(k) decision material K173420: Radiomics App v1.0

    U.S. Food and Drug Administration - Published 2017-12-27 - Accessed 2026-09-04

    Scope: identity, purpose, clinicalUse, modality, anatomy, intendedUsers, careSetting, workflowRole, input, output, limitations, evidenceSummary

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  4. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  5. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

Research history

  1. K173420 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

Alternatives

Related tools to compare

FDA-cleared K220408

AVIEW RT ACS

Coreline Soft Co.,Ltd

AVIEW RT ACS is a medical image-management and processing system for radiation-treatment planning with deep-learning auto-segmentation evidence described in the exact FDA decision...

Modality
Radiotherapy Planning Imaging
Anatomy
Organs at risk across breast, head and neck, abdomen, and pelvis CT validation cases, including lung, heart, eyes, brain, mandible, kidneys, liver, femur, and bladder
Decision
2022-11-10
FDA-cleared K212915

MVision AI Segmentation

MVision AI

MVision AI Segmentation automatically generates initial contours from CT images for radiotherapy planning; the contours are transferred for professional visualization, review,...

Modality
Radiotherapy Planning Imaging
Anatomy
Head and neck, brain, breast, lung and abdomen, male pelvis, and female pelvis structures
Decision
2022-05-03
FDA-cleared K252190

DeepBT Detector-Plus

Aitewan Biomedical Technology, Inc.

DeepBT Detector-Plus provides initial gross-tumor-volume contours on T1-weighted contrast-enhanced and optional T2-weighted MR images for radiation-treatment planning in adults...

Modality
Radiotherapy Planning Imaging
Anatomy
Brain tumors, including metastases, meningiomas, and acoustic neuromas
Decision
2026-04-10
FDA-cleared K252304

syngo.via RT Image Suite VC10

Varian Medical Systems, Inc.

syngo.via RT Image Suite VC10 provides three-dimensional and four-dimensional multimodality visualization, registration, fusion, contouring, and planning support for...

Modality
Radiotherapy Planning Imaging
Anatomy
Brain, pelvis, thorax, abdomen, targets, organs at risk, and moving tumors or organs
Decision
2026-03-18