FDA-cleared profile

Radiomics App v1.0

Microsoft Corp. · Radiotherapy Planning Imaging · FDA-cleared (510(k))

Authorized purpose: Microsoft Radiomics App v1.0 is a software-only medical device for trained radiation oncologists, dosimetrists, and physicists to derive organ and tumor contours for radiation-treatment planning. It supports CT and MR images, image and contour review, contour creation and modification, transfer to treatment-planning systems, adaptive-therapy support, archiving, tumor and healthy-structure localization, image fusion, and three-dimensional rendering. Images reviewed with the software should not be used for primary image interpretations, and it is not for digital mammography.

Clinical role
Radiotherapy contouring, treatment planning, treatment management, image processing, or session support under qualified review and approval.
Images and patients
Radiotherapy Planning Imaging · CT and MR images; tumors and healthy anatomical structures for radiotherapy planning
Key limitation
Keep in mind: The exact labeling excludes primary image interpretation and digital mammography; generated contours require review and professional approval before treatment use.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: Microsoft Radiomics App v1.0 is a software-only medical device for trained radiation oncologists, dosimetrists, and physicists to derive organ and tumor contours for radiation-treatment planning. It supports CT and MR images, image and contour review, contour creation and modification, transfer to treatment-planning systems, adaptive-therapy support, archiving, tumor and healthy-structure localization, image fusion, and three-dimensional rendering. Images reviewed with the software should not be used for primary image interpretations, and it is not for digital mammography.
  • Which images or clinical context are in scope?Modality: Radiotherapy Planning Imaging Anatomy or clinical context: CT and MR images; tumors and healthy anatomical structures for radiotherapy planning Input: Supported CT or MR images, structures, prescriptions, treatment plans, registrations, and treatment-session data.
  • Who uses it?Intended users: Radiation oncologists, medical physicists, dosimetrists, therapists, and other qualified radiation professionals.
  • What does the software produce?Output: Contours, plans, adapted plans, image-processing results, treatment-management records, or session information requiring professional review.
  • What is its most important limitation?Key limitation: The exact labeling excludes primary image interpretation and digital mammography; generated contours require review and professional approval before treatment use.

Clinical takeaway. Clinical role: Microsoft Radiomics App v1.0 is a software-only medical device for trained radiation oncologists, dosimetrists, and physicists to derive organ and tumor contours for radiation-treatment planning. It supports CT and MR images, image and contour review, contour creation and modification, transfer to treatment-planning systems, adaptive-therapy support, archiving, tumor and healthy-structure localization, image fusion, and three-dimensional rendering. Images reviewed with the software should not be used for primary image interpretations, and it is not for digital mammography.

Evidence takeaway. Evidence from the reviewed source record: The exact FDA document reports software verification and validation testing for a moderate level of concern device: manual and automated verification tests, all of which passed against the stated software requirements, and validation tests of measurements, volumetric rendering, multi-planar reconstruction, assisted contouring and automatic contouring showing the software meets user needs and intended uses. It states animal and clinical studies were not necessary to establish substantial equivalence. Results are as reported in the FDA submission, not independent validation.

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
Microsoft Radiomics App v1.0 is a software-only medical device for trained radiation oncologists, dosimetrists, and physicists to derive organ and tumor contours for radiation-treatment planning. It supports CT and MR images, image and contour review, contour creation and modification, transfer to treatment-planning systems, adaptive-therapy support, archiving, tumor and healthy-structure localization, image fusion, and three-dimensional rendering. Images reviewed with the software should not be used for primary image interpretations, and it is not for digital mammography.
  • Exact FDA submission
  • Checked 2026-09-04
Workflow role
Radiotherapy contouring, treatment planning, treatment management, image processing, or session support under qualified review and approval.
  • Exact FDA submission
  • Checked 2026-09-04
Required input
Supported CT or MR images, structures, prescriptions, treatment plans, registrations, and treatment-session data.
  • Exact FDA submission
  • Checked 2026-09-04
Output and clinician action
Contours, plans, adapted plans, image-processing results, treatment-management records, or session information requiring professional review.
  • Exact FDA submission
  • Checked 2026-09-04
Care setting
Radiation-oncology settings within the labeled workflow.
  • Exact FDA submission
  • Checked 2026-09-04
Intended users
Radiation oncologists, medical physicists, dosimetrists, therapists, and other qualified radiation professionals.
  • Exact FDA submission
  • Checked 2026-09-04
Limitations
The exact labeling excludes primary image interpretation and digital mammography; generated contours require review and professional approval before treatment use.
  • Exact FDA submission
  • Checked 2026-09-04
Evidence summary
The exact FDA document reports software verification and validation testing for a moderate level of concern device: manual and automated verification tests, all of which passed against the stated software requirements, and validation tests of measurements, volumetric rendering, multi-planar reconstruction, assisted contouring and automatic contouring showing the software meets user needs and intended uses. It states animal and clinical studies were not necessary to establish substantial equivalence. Results are as reported in the FDA submission, not independent validation.
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "The software for this device was considered as a moderate level of concern." … "Two different types of verification testing were conducted to verify the software requirements: Manual and Automated. All tests passed, demonstrating that the software performance is in accordance with the stated software requirements.". Read from the exact FDA decision document on 2026-11-14.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
The exact FDA document reports software verification and validation testing for a moderate level of concern device: manual and automated verification tests, all of which passed against the stated software requirements, and validation tests of measurements, volumetric rendering, multi-planar reconstruction, assisted contouring and automatic contouring showing the software meets user needs and intended uses. It states animal and clinical studies were not necessary to establish substantial equivalence. Results are as reported in the FDA submission, not independent validation.
  • FDA source
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "The software for this device was considered as a moderate level of concern." … "Two different types of verification testing were conducted to verify the software requirements: Manual and Automated. All tests passed, demonstrating that the software performance is in accordance with the stated software requirements.". Read from the exact FDA decision document on 2026-11-14.

Reported sensitivity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-04

Sensitivity is not promoted as a product-wide endpoint. Compare only an exact endpoint, threshold, population, version, study design, and source if separately verified.

Reported specificity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-04

Specificity is not promoted as a product-wide endpoint. Compare only an exact endpoint, threshold, population, version, study design, and source if separately verified.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

K173420 FDA decision evidence package for Radiomics App v1.0

  • DesignFDA decision evidence and product verification described in the exact submission or PMA decision record.
  • PopulationCT and MR images; tumors and healthy anatomical structures for radiotherapy planning
  • Scopeexact submission
  • Tested versionRadiomics App v1.0 configuration described in K173420
  • IndependenceSponsor or vendor study

K173420

FDA decision material is manufacturer-submitted regulatory evidence. It is not independent clinical validation unless the exact source expressly establishes that distinction; local technical, clinical, privacy, security, and workflow acceptance remains necessary.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K173420 by exact native submission identifier. This does not exclude product-family, later-version, or adverse-event information.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-04

Exact-identifier review does not prove absence of all product-family safety signals.

Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not publicly available: Reviewed sources do not publish an exact-product procurement price with a stated offer scope.

Reimbursement

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: Supported CT or MR images, structures, prescriptions, treatment plans, registrations, and treatment-session data.

    Scope: Exact FDA submission.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-09-04 · FDA 510(k) decision material K173420: Radiomics App v1.0

  2. Routing

    Source context, not route validation: The exact FDA package describes Radiomics App accepting DICOM images generated elsewhere, integrating into the clinical data network of radiation therapy centers to receive data from CT and MR scanners, and exporting segmentations back to that network for use in a radiotherapy treatment planning system. Interface versions and local compatibility require confirmation.

    Scope: Exact FDA submission. Packet wording: "Radiomics App integrates into the clinical data network of radiation therapy treatment centers, receiving data from imaging devices such as CT and MR scanners." … "These segmentations are then exported back into the clinical data network, and subsequently utilized in a radiotherapy treatment planning system to generate a treatment plan for a patient.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-11-14 · FDA 510(k) decision material K173420: Radiomics App v1.0

  3. AI processing

    Unresolved route: Not established for the exact route.

    Scope: Exact FDA submission. The exact public FDA decision material does not establish the complete deployment topology, hosting responsibility, availability, upgrade path, or service contract.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-09-04 ·

  4. Returned objects

    Source context, not route validation: Contours, plans, adapted plans, image-processing results, treatment-management records, or session information requiring professional review.

    Scope: Exact FDA submission.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-04 · FDA 510(k) decision material K173420: Radiomics App v1.0

  5. Viewer/worklist/report

    Unresolved route: Contours, plans, adapted plans, image-processing results, treatment-management records, or session information requiring professional review.

    Scope: Exact FDA submission.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-04 · FDA 510(k) decision material K173420: Radiomics App v1.0

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Integration
The exact FDA package describes Radiomics App accepting DICOM images generated elsewhere, integrating into the clinical data network of radiation therapy centers to receive data from CT and MR scanners, and exporting segmentations back to that network for use in a radiotherapy treatment planning system. Interface versions and local compatibility require confirmation.
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "Radiomics App integrates into the clinical data network of radiation therapy treatment centers, receiving data from imaging devices such as CT and MR scanners." … "These segmentations are then exported back into the clinical data network, and subsequently utilized in a radiotherapy treatment planning system to generate a treatment plan for a patient.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.

Training
Only radiation oncologists, medical physicists, dosimetrists, therapists, and other qualified radiation professionals should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
  • Exact FDA submission
  • Checked 2026-09-04
Monitoring
Monitor K173420 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: The exact labeling excludes primary image interpretation and digital mammography; generated contours require review and professional approval before treatment use.
  • Exact FDA submission
  • Checked 2026-09-04

This is buyer-facing monitoring guidance derived from the exact cleared scope; it is not a claim that the FDA package provides a live monitoring service.

Release and scope verification

Map installed model, build, options, population, anatomy, inputs, outputs, and enabled functions to K173420 before acceptance.

exact submission · checked 2026-09-04

Later product-family features are not evidence that this cleared or authorized release includes them.

Deployment boundary

The exact public deployment topology remains incomplete; establish hosting responsibility, data flow, availability, upgrade, service, and downtime boundaries locally.

exact submission · checked 2026-09-04

FDA-described topology does not establish current hosting contracts, security controls, uptime, or local qualification.

Interoperability acceptance

The exact public integration contract remains incomplete; establish and test supported input, output, identity, routing, archive, and downtime behavior locally.

exact submission · checked 2026-09-04

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. The exact labeling excludes primary image interpretation and digital mammography; generated contours require review and professional approval before treatment use.

exact submission · checked 2026-09-04

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

exact submission · checked 2026-09-04

No exact-product price or reimbursement determination was promoted.

Lifecycle monitoring

Monitor K173420 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: The exact labeling excludes primary image interpretation and digital mammography; generated contours require review and professional approval before treatment use.

exact submission · checked 2026-09-04

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes.

Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K173420
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
Radiotherapy Planning Imaging
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Anatomy
CT and MR images; tumors and healthy anatomical structures for radiotherapy planning
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2017-12-27
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
27Fields reviewed
9Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, fda decision material, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not publicly available - checked 2026-09-04 - Exact FDA submission

    No public list price for the exact cleared or authorized configuration was found. Obtain a written quote separating license or hardware basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Reimbursement: Not publicly available - checked 2026-09-04 - Exact FDA submission

    No exact-product reimbursement determination was promoted in this timeboxed review. Verify payer, site-of-service, coding, coverage, and local payment assumptions independently.

  • Deployment: Not publicly available - checked 2026-09-04 - Exact FDA submission

    The exact public FDA decision material does not establish the complete deployment topology, hosting responsibility, availability, upgrade path, or service contract.

  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Security: Not publicly available - checked 2026-09-04 - Exact FDA submission

    Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.

  • Installation timeline: Not yet reviewed

5 sources - 3 history items

Source ledger

  1. FDA 510(k) decision material K173420: Radiomics App v1.0

    U.S. Food and Drug Administration - Published 2017-12-27 - Accessed 2026-09-04

    Scope: identity, purpose, clinicalUse, modality, anatomy, intendedUsers, careSetting, workflowRole, input, output, limitations, evidenceSummary

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  4. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  5. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

Research history

  1. K173420 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K220408

AVIEW RT ACS

Coreline Soft Co.,Ltd

Authorized purpose: AVIEW RT ACS is a medical image-management and processing system for radiation-treatment planning with deep-learning auto-segmentation evidence described in the exact FDA decision material; the submission compares automated organ segmentation with a predicate using Dice similarity coefficient and 95% Hausdorff distance on CT cases.

Modality
Radiotherapy Planning Imaging
Anatomy
Organs at risk across breast, head and neck, abdomen, and pelvis CT validation cases, including lung, heart, eyes, brain, mandible, kidneys, liver, femur, and bladder
Decision
2022-11-10
FDA-cleared K212915

MVision AI Segmentation

MVision AI

Authorized purpose: MVision AI Segmentation automatically generates initial contours from CT images for radiotherapy planning; the contours are transferred for professional visualization, review, modification, and approval, and the labeled software is not intended for lesion or tumor detection or real-time adaptive planning.

Modality
Radiotherapy Planning Imaging
Anatomy
Head and neck, brain, breast, lung and abdomen, male pelvis, and female pelvis structures
Decision
2022-05-03
FDA-cleared K252190

DeepBT Detector-Plus

Aitewan Biomedical Technology, Inc.

Authorized purpose: DeepBT Detector-Plus provides initial gross-tumor-volume contours on T1-weighted contrast-enhanced and optional T2-weighted MR images for radiation-treatment planning in adults with the labeled brain tumors.

Modality
Radiotherapy Planning Imaging
Anatomy
Brain tumors, including metastases, meningiomas, and acoustic neuromas
Decision
2026-04-10
FDA-cleared K252304

syngo.via RT Image Suite VC10

Varian Medical Systems, Inc.

Authorized purpose: syngo.via RT Image Suite VC10 provides three-dimensional and four-dimensional multimodality visualization, registration, fusion, contouring, and planning support for radiation-treatment preparation.

Modality
Radiotherapy Planning Imaging
Anatomy
Brain, pelvis, thorax, abdomen, targets, organs at risk, and moving tumors or organs
Decision
2026-03-18

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination

Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.

Accepted repository change history

No entry has been added to this scoped change history for this record. This is not proof that the record never changed.

Correction policy and limits · Public repository change history