Authorized purpose: syngo.via RT Image Suite VC10 provides three-dimensional and four-dimensional multimodality visualization, registration, fusion, contouring, and planning support for radiation-treatment preparation.
Clinical role
Radiotherapy contouring, treatment planning, adaptation, or treatment-session support under professional review.
Images and patients
Radiotherapy Planning Imaging · Brain, pelvis, thorax, abdomen, targets, organs at risk, and moving tumors or organs
Key limitation
Keep in mind: Generated or modified contours and plans require qualified review, editing when needed, independent checks, and final approval before treatment.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: syngo.via RT Image Suite VC10 provides three-dimensional and four-dimensional multimodality visualization, registration, fusion, contouring, and planning support for radiation-treatment preparation.
Which images or clinical context are in scope?Modality: Radiotherapy Planning Imaging Anatomy or clinical context: Brain, pelvis, thorax, abdomen, targets, organs at risk, and moving tumors or organs Input: Supported CT or MR images, structures, prescriptions, plans, registrations, and treatment-session data.
Who uses it?Intended users: Qualified radiation oncologists, medical physicists, dosimetrists, and radiation therapists.
What does the software produce?Output: Contours, treatment plans, adapted plans, dose or evaluation results, or session information requiring review and approval.
What is its most important limitation?Key limitation: Generated or modified contours and plans require qualified review, editing when needed, independent checks, and final approval before treatment.
Clinical takeaway. Clinical role: syngo.via RT Image Suite VC10 provides three-dimensional and four-dimensional multimodality visualization, registration, fusion, contouring, and planning support for radiation-treatment preparation.
Evidence takeaway. Evidence from the reviewed source record: The exact FDA document reports software verification and validation, usability validation, and algorithm testing on independent test sets; no animal studies or clinical tests were included. CT auto-contouring was tested on 469 subjects. The MR brain metastasis model, tested on 30 subjects (60 test cases in its results table) against expert-annotated ground-truth segmentations, had mean Dice 0.74 and mean sensitivity 92.5 percent (lower 95th % confidence bound 85.8). Results are as reported in the FDA submission, not independent validation.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
syngo.via RT Image Suite VC10 provides three-dimensional and four-dimensional multimodality visualization, registration, fusion, contouring, and planning support for radiation-treatment preparation.
The exact FDA document reports software verification and validation, usability validation, and algorithm testing on independent test sets; no animal studies or clinical tests were included. CT auto-contouring was tested on 469 subjects. The MR brain metastasis model, tested on 30 subjects (60 test cases in its results table) against expert-annotated ground-truth segmentations, had mean Dice 0.74 and mean sensitivity 92.5 percent (lower 95th % confidence bound 85.8). Results are as reported in the FDA submission, not independent validation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "Software verification and validation was conducted, and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff" … "No animal studies or clinical tests have been included in this pre-market submission. However, usability validation was successfully performed". Read from the exact FDA decision document on 2026-11-14.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
The exact FDA document reports software verification and validation, usability validation, and algorithm testing on independent test sets; no animal studies or clinical tests were included. CT auto-contouring was tested on 469 subjects. The MR brain metastasis model, tested on 30 subjects (60 test cases in its results table) against expert-annotated ground-truth segmentations, had mean Dice 0.74 and mean sensitivity 92.5 percent (lower 95th % confidence bound 85.8). Results are as reported in the FDA submission, not independent validation.
FDA source
Exact FDA submission
Checked 2026-11-14
Packet wording: "Software verification and validation was conducted, and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff" … "No animal studies or clinical tests have been included in this pre-market submission. However, usability validation was successfully performed". Read from the exact FDA decision document on 2026-11-14.
Sensitivity is not a product-wide endpoint for this system or software function. Compare only an exact endpoint, threshold, population, version, and source if one is later verified.
Reported specificity
Not applicable
Not applicable
Not applicable
Checked 2026-09-04
Specificity is not a product-wide endpoint for this system or software function. Compare only an exact endpoint, threshold, population, version, and source if one is later verified.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
K252304 FDA evidence package for syngo.via RT Image Suite VC10
DesignFDA software substantial-equivalence decision with image-processing, registration, contouring, synthetic-CT, usability, and workflow verification
PopulationRadiation-treatment preparation using CT, PET, PET-CT, MRI, and linac-CBCT studies for labeled targets, organs at risk, and moving anatomy
Scopeexact submission
Tested versionsyngo.via RT Image Suite VC10 described in K252304
IndependenceSponsor or vendor study
K252304
Manufacturer evidence submitted for the exact FDA decision. It is not independent clinical validation, and local technical, clinical, privacy, security, and workflow acceptance remains necessary.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K252304 by exact native submission identifier. This does not exclude product-family, later, or adverse-event information.
Public source
Exact FDA submission
Checked 2026-09-04
Exact-identifier review does not prove absence of all product-family safety signals.
No separate named-product Medicare payment was identified in the reviewed CMS radiation-oncology material. Evaluate net planning or delivery value within the radiotherapy service line.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: Supported CT or MR images, structures, prescriptions, plans, registrations, and treatment-session data.
Scope: Exact FDA submission.
Confirm scanner/model, protocol, series completeness and both software builds.
Unresolved route: Not established for the exact route.
Scope: Exact FDA submission. The public exact-submission material does not establish a complete integration contract. Confirm DICOM, worklist, identity, interfaces, exports, routing, audit behavior, and downtime locally.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Field checked 2026-09-04 ·
AI processing
Unresolved route: Not established for the exact route.
Scope: Exact FDA submission. The public exact-submission material does not establish the complete deployment topology, hosting responsibility, availability, upgrade path, or service contract.
Confirm processing location, queue ownership, exclusions and exact AI build.
Field checked 2026-09-04 ·
Returned objects
Source context, not route validation: Contours, treatment plans, adapted plans, dose or evaluation results, or session information requiring review and approval.
Scope: Exact FDA submission.
Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.
Unresolved route: Contours, treatment plans, adapted plans, dose or evaluation results, or session information requiring review and approval.
Scope: Exact FDA submission.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
Siemens Healthineers' corporate cybersecurity page states a Secure Development Lifecycle addressing pre-market and post-market phases for products under development and a range of existing offerings; a Cybersecurity Management Service maintaining a product security inventory (over 1,000 product versions and over 21,000 monitored software components) with security whitepapers, SBOMs and vulnerability assessments available to customers via teamplay Fleet; incident and vulnerability management channels; and that Siemens Healthineers AG maintains an independent certification of its global Cybersecurity Management System to ISO/IEC 27001:2022 extended by ISO/IEC 27701:2019, with the certificate referenced to the SGS directory. Vendor statement; no certificate or report was read.
Vendor-wide
Checked 2026-10-26
Vendor statement, not an attestation. Identity bridge: Siemens Healthineers' own press release (/press/releases/varian-closing) states it completed the acquisition of Varian Medical Systems, Inc. Corporate program statement; exact-release scope is not established.
Only qualified radiation oncologists, medical physicists, dosimetrists, and radiation therapists should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Generated or modified contours and plans require qualified review, editing when needed, independent checks, and final approval before treatment.
Product family
Checked 2026-09-04
Current vendor material is product-family context for K252304; confirm exact cleared version, enabled options, deployment, and local compatibility.
Exact public integration details remain incomplete; establish and test the supported input, output, identity, routing, archive, and downtime contract locally.
product family · checked 2026-09-04
Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally. Current vendor material is product-family context; confirm exact release, enabled options, deployment, and local compatibility.
Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Generated or modified contours and plans require qualified review, editing when needed, independent checks, and final approval before treatment.
exact submission · checked 2026-09-04
Define stop-use, escalation, rollback, and revalidation triggers before production use.
Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.
product family · checked 2026-09-04
No exact-product price was found in the reviewed public material; productivity or value claims require local evidence.
Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Generated or modified contours and plans require qualified review, editing when needed, independent checks, and final approval before treatment.
product family · checked 2026-09-04
Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. Current vendor material is product-family context; confirm exact release, enabled options, deployment, and local compatibility.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Human reviewedStatus
2026-09-20Last searched
27Fields reviewed
13Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision material, vendor product page, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, literature index, trial registry, fda decision summary, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not publicly available - checked 2026-09-04 - Exact FDA submission
No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.
Pricing model: Not yet reviewed
Setup and integration fees: Not yet reviewed
Recurring fees: Not yet reviewed
Minimum commitment: Not yet reviewed
Contract duration: Not yet reviewed
Renewal terms: Not yet reviewed
Cancellation terms: Not yet reviewed
Support and service terms: Not yet reviewed
Trial or pilot terms: Not yet reviewed
Deployment: Not publicly available - checked 2026-09-04 - Exact FDA submission
The public exact-submission material does not establish the complete deployment topology, hosting responsibility, availability, upgrade path, or service contract.
Integration: Not publicly available - checked 2026-09-04 - Exact FDA submission
The public exact-submission material does not establish a complete integration contract. Confirm DICOM, worklist, identity, interfaces, exports, routing, audit behavior, and downtime locally.
Authorized purpose: DeepBT Detector-Plus provides initial gross-tumor-volume contours on T1-weighted contrast-enhanced and optional T2-weighted MR images for radiation-treatment planning in adults with the labeled brain tumors.
Modality
Radiotherapy Planning Imaging
Anatomy
Brain tumors, including metastases, meningiomas, and acoustic neuromas
Authorized purpose: The EMLA family delivers external-beam radiation therapy for adult and pediatric patients with benign, malignant, or metastatic disease. Authorized techniques vary by model: Synergy and Harmony include conventional and fractionated delivery; Infinity and Harmony Pro add SBRT/SABR; Versa HD and Evo also support SRS and specified functional disorders.
Authorized purpose: TAIMedImg DeepMets assists trained professionals by providing initial gross-tumor-volume contours on axial contrast-enhanced T1 brain MR images for radiation-treatment planning in adults with known brain metastases; the contours require thorough review and are not a replacement for manual contouring.
Authorized purpose: RayStation and RayPlan releases covered by K240398 are configurable radiotherapy and medical-oncology software that proposes or supports treatment plans from user inputs, with authorized users required to review and approve plans before treatment.
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.
Accepted repository change history
No entry has been added to this scoped change history for this record. This is not proof that the record never changed.