FDA-cleared profile

CoLumbo

Smart Soft Healthcare AD

CoLumbo is an image post-processing and measurement software tool that provides quantitative spine measurements from previously-acquired DICOM lumbar spine Magnetic Resonance (MR) images for users' review, analysis, and interpretation. It provides the following functionality to assist users in visualizing, measuring and documenting out-of range measurements:- Feature segmentation;- Feature measurement;- Threshold-based labeling of out-of-range measurement; and- Export of measurement results to a written report for user's review, revise and approval. Columbo does not produce or recommend any type of medical diagnosis or treatment. Instead, it simply helps users to more easily identify and classify features in lumbar MR images and compile a report. The user is responsible forconfirming/modifying settings, reviewing and verifying the software-generated measurements, inspecting out-of-range measurements, and approving draft report content using their medical judgment and discretion. The device is intended to be used only by hospitals and other medical institutions. Only DICOM images of MRI acquired from lumbar spine exams of patients aged 18 and above are considered to be valid input. CoLumbo does not support DICOM images of patients that are pregnant, undergo MRI scan with contrast media, or have post-operational complications, scoliosis, tumors, infections, fractures.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
CoLumbo is an image post-processing and measurement software tool that provides quantitative spine measurements from previously-acquired DICOM lumbar spine Magnetic Resonance (MR) images for users' review, analysis, and interpretation. It provides the following functionality to assist users in visualizing, measuring and documenting out-of range measurements:- Feature segmentation;- Feature measurement;- Threshold-based labeling of out-of-range measurement; and- Export of measurement results to a written report for user's review, revise and approval. Columbo does not produce or recommend any type of medical diagnosis or treatment. Instead, it simply helps users to more easily identify and classify features in lumbar MR images and compile a report. The user is responsible forconfirming/modifying settings, reviewing and verifying the software-generated measurements, inspecting out-of-range measurements, and approving draft report content using their medical judgment and discretion. The device is intended to be used only by hospitals and other medical institutions. Only DICOM images of MRI acquired from lumbar spine exams of patients aged 18 and above are considered to be valid input. CoLumbo does not support DICOM images of patients that are pregnant, undergo MRI scan with contrast media, or have post-operational complications, scoliosis, tumors, infections, fractures.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Intended users
Radiologist
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Intended-user text reported in the exact-submission ACR model card.

Care setting and population
Adult
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Patient-population text reported in the exact-submission ACR model card; institutional setting may still require vendor confirmation.

Workflow role
Segmentation or contouring
  • Public source
  • Exact FDA submission
  • Checked 2026-08-31

Conservative normalization of explicit wording in an exact-submission source.

Required input
MRI images or data involving Spine; consult the FDA labeling for exact protocol, series, and quality requirements.
  • Public source
  • Exact FDA submission
  • Checked 2026-08-31

Conservative normalization of explicit wording in an exact-submission source.

Output and human action
The software is a quantitative imaging tool that assists radiologists and neuro- and spine surgeons ( users ) to identify and measure lumbar spine features in medical images and record their observations in a report. The users then confirm whether the out-of-range measurements represent any true abnormality versus a spurious finding, such as an artifact or normal variation of the anatomy. The segmentation and measurements are classified using modifiers based on rule-based algorithms and thresholds set by each software user and stored in the user's individualized software settings. The user also identifies and classifies any other observations that the software may not annotate.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Clinical output reported in the exact-submission ACR model card.

Limitations
Scanner manufacturers: All
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Limitations and supported-acquisition details reported in the exact-submission ACR model card.

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K220497
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
MRI
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Anatomy
Spine
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2022-06-23
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
The exact-submission ACR model card explicitly references DICOM in the described input, output, or workflow; exact interfaces, versions, and local compatibility require confirmation.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Interoperability terms explicitly reported in the exact-submission ACR model card; interface versions and local compatibility still require confirmation.

Deployment and data flow
On-premise
  • Public source
  • Exact FDA submission
  • Checked 2026-08-13

Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.

Security and privacy
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.

Training and support
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

Monitoring and change control
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

Input or output interoperability

DICOM is explicitly referenced in the described input, output, or workflow.

exact submission · checked 2026-09-01

The model card does not establish every supported DICOM object, transfer method, or local interface.

Supported acquisition environment

Scanner manufacturers: All

exact submission · checked 2026-09-01

Confirm currently supported software versions, protocols, and site-specific acquisition configurations.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for CoLumbo; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-08

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
Study type: Stand-Alone Performance. Cases: 101. Sites: 7. Reference standard: Interpretation by Reviewing Clinician
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Structured study metadata from the exact-submission ACR model card; this is not an independent evidence synthesis.

Reported sensitivity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

Reported specificity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

CoLumbo regulatory performance study summary

  • DesignStand-Alone Performance
  • PopulationAdult
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K220497
  • Samplen=101
  • Sites7
  • IndependenceIndependence not established

K220497

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

No public price or quote for the exact K220497 configuration was established in this review.

Reimbursement and coding
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-08

No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K220497. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-08

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-08Last searched
23Fields reviewed
9Source classes checked
16Awaiting review PubMed leads
10Awaiting review trial leads
0Unreviewed FDA recall leads

Exact-submission ACR model-card fields were normalized under the current provenance rules; vendor and independent-study confirmation remain distinct. Exact FDA scope controls clinical claims. ACR and official vendor material provide bounded catalog or product-family context. Candidate literature metadata remains a discovery queue until full-text identity, endpoint, population, and version review is completed. Automated exact-name discovery found 16 PubMed and 10 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA K220497: CoLumbo

    U.S. Food and Drug Administration - Published 2022-06-23 - Accessed 2026-09-08

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. CoLumbo model card (K220497)

    American College of Radiology - Published 2026-06-26 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, limitations, evidenceSummary, integration, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K220497

    U.S. Food and Drug Administration - Accessed 2026-09-08

    Scope: postmarketSafety

Research history

  1. K220497 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

Alternatives

Related tools to compare

FDA-cleared K233030

BoneMRI

MRIguidance B.V

MRI-based BoneMRI image enhancement for visualizing pelvic and cervical, thoracic, lumbar and S1 spinal bone structures with CT-like radiodensity contrast; not for tumor diagnosis...

Modality
MRI
Anatomy
Spine
Decision
2024-03-01
FDA-cleared K230197

BoneMRI v1.6

MRIguidance B.V.

MRI-based BoneMRI v1.6 image enhancement for visualizing pelvic and L3-S1 lumbar bone structures with enhanced contrast; not for tumor diagnosis or monitoring and not a CT...

Modality
MRI
Anatomy
Spine
Decision
2023-02-22