Authorized purpose: CT post-processing for spine review, labeling, display optimization, and report dictation.
Clinical role
Automatically detects and labels vertebrae and optimizes anatomy display on CT images; users can move, edit, hide or delete labels, with initial vertebra identification supporting clinician visualization of the spine and report dictation.
Images and patients
CT/CTA · Spine
Key limitation
Keep in mind: No disease-specific indication or explicit exclusion list was established in the reviewed packet.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: CT post-processing for spine review, labeling, display optimization, and report dictation.
Which images or clinical context are in scope?Modality: CT/CTA Anatomy or clinical context: Spine Input: CT acquisitions for spine, trauma, oncology, and routine examinations, including single-energy and GSI studies.
Who uses it?Intended users: Clinicians.
What does the software produce?Output: Three-dimensional vertebra positions and labels, editable labeling, MPR/MIP/MinIP/VR views, measurements, and HU tools.
What is its most important limitation?Key limitation: No disease-specific indication or explicit exclusion list was established in the reviewed packet.
Clinical takeaway. Clinical role: CT post-processing for spine review, labeling, display optimization, and report dictation.
Evidence takeaway. Evidence from the reviewed source record: Automated spine-labeling success rate was reported as greater than 90%; three board-certified radiologists used a five-point Likert assessment, and sample size was not stated.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
CT post-processing for spine review, labeling, display optimization, and report dictation.
Exact FDA submission
Checked 2026-09-15
Clinical-use display mirrors the exact FDA-supported purpose reconciled in this remediation finding.
Automatically detects and labels vertebrae and optimizes anatomy display on CT images; users can move, edit, hide or delete labels, with initial vertebra identification supporting clinician visualization of the spine and report dictation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "Bone VCAR automated vertebra detection algorithm returns the list of detected vertebrae position in 3D coordinates and associated labels." … "All labels can be moved, edited, hidden or deleted.". Read from the exact FDA decision document on 2026-11-14.
CT images that include the spine, from multiple care areas and exam types including trauma, oncology, and routine body, not specific to any disease state; the FDA document does not specify a facility type.
Exact FDA submission
Checked 2026-11-14
Packet wording: "The software is intended to support clinicians in the review of images that include the spine by providing tools to label the spine and optimize the display of anatomy within the CT image." … "Bone VCAR may be used for multiple care areas and is not specific to any disease state. It can be utilized during the review of various types of exams including trauma, oncology, and routine body.". Read from the exact FDA decision document on 2026-11-14.
Automated spine-labeling success rate was reported as greater than 90%; three board-certified radiologists used a five-point Likert assessment, and sample size was not stated.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
Automated spine-labeling success rate was reported as greater than 90%; three board-certified radiologists used a five-point Likert assessment, and sample size was not stated.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
Bone VCAR regulatory performance study summary
DesignStand-Alone Performance
PopulationPopulation described in the exact-submission ACR model card
Scopeexact submission
Tested versionExact product version not reported; regulatory study summarized for FDA submission K183204
IndependenceIndependence not established
K183204
ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K183204. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: CT acquisitions for spine, trauma, oncology, and routine examinations, including single-energy and GSI studies.
Scope: Exact FDA submission. FDA packet input scope.
Confirm scanner/model, protocol, series completeness and both software builds.
Unresolved route: Not established for the exact route.
Scope: Exact FDA submission. The exact FDA decision document was reviewed on 2026-11-14 and does not state the subject device's DICOM, PACS, RIS, or EHR interfaces. Confirm supported DICOM services and PACS, RIS, EHR, and interface versions with the vendor.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Field checked 2026-11-14 ·
AI processing
Source context, not route validation: The exact FDA package describes Bone VCAR as a post-processing application option for the Advantage Workstation (AW) platform, CT scanner, cloud or PACS stations. Current hosting, data flow, and commercial configuration require vendor confirmation.
Scope: Exact FDA submission. Packet wording: "Bone VCAR is a post processing application option for the Advantage Workstation (AW) platform, CT Scanner, Cloud or PACS stations which can be used in the analysis of CT images.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
Confirm processing location, queue ownership, exclusions and exact AI build.
Source context, not route validation: Three-dimensional vertebra positions and labels, editable labeling, MPR/MIP/MinIP/VR views, measurements, and HU tools.
Scope: Exact FDA submission. FDA packet output scope.
Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.
Unresolved route: Three-dimensional vertebra positions and labels, editable labeling, MPR/MIP/MinIP/VR views, measurements, and HU tools.
Scope: Exact FDA submission. FDA packet output scope.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
The exact FDA package describes Bone VCAR as a post-processing application option for the Advantage Workstation (AW) platform, CT scanner, cloud or PACS stations. Current hosting, data flow, and commercial configuration require vendor confirmation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "Bone VCAR is a post processing application option for the Advantage Workstation (AW) platform, CT Scanner, Cloud or PACS stations which can be used in the analysis of CT images.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
ACR AI Central provides an exact-submission model card for Bone VCAR (BVCAR); FDA labeling remains controlling for clinical use, limitations, and performance.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Exact FDA record reviewedStatus
2026-09-20Last searched
25Fields reviewed
10Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, fda decision summary packet, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not verified - checked 2026-09-09 - Exact FDA submission
No public price or quote for the exact K183204 configuration was established in this review.
Pricing model: Not yet reviewed
Setup and integration fees: Not yet reviewed
Recurring fees: Not yet reviewed
Minimum commitment: Not yet reviewed
Contract duration: Not yet reviewed
Renewal terms: Not yet reviewed
Cancellation terms: Not yet reviewed
Support and service terms: Not yet reviewed
Trial or pilot terms: Not yet reviewed
Reimbursement: Not publicly available - checked 2026-09-09 - Exact FDA submission
No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.
Integration: Not publicly available - checked 2026-11-14 - Exact FDA submission
The exact FDA decision document was reviewed on 2026-11-14 and does not state the subject device's DICOM, PACS, RIS, or EHR interfaces. Confirm supported DICOM services and PACS, RIS, EHR, and interface versions with the vendor.
Data hosting location: Not yet reviewed
Data retention: Not yet reviewed
Documented data use: Not yet reviewed
Hardware requirements: Not yet reviewed
Network requirements: Not yet reviewed
Identity and access: Not yet reviewed
Security: Not verified - checked 2026-09-09 - Exact FDA submission
Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.
Installation timeline: Not yet reviewed
Training: Not verified - checked 2026-09-09 - Exact FDA submission
Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.
Monitoring: Not verified - checked 2026-09-09 - Exact FDA submission
Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.
Authorized purpose: Auto-Seg and Spine Auto-Seg accept spine and pelvis CT images and provide three-dimensional segmentation and labels to support spine-surgery planning and assessment by spine surgeons.
Authorized purpose: Analyzes existing CT images in DICOM format from adults aged 30 years and older to estimate average areal BMD of spinal bones and flag low bone density below a preset threshold, including opportunistic scans without a phantom; it does not replace DXA or other dedicated BMD testing.
Authorized purpose: Use with CT images to automatically generate anatomically focused multi-planar spine reformats, apply vertebral-body and intervertebral-disc labels, and export the results to predetermined DICOM destinations for review.
AI-Rad Companion (Musculoskeletal) is an image processing software that provides quantitative and qualitative analysis from previously acquired Computed Tomography DICOM images to support radiologists and physicians from emergency medicine, specialty care, urgent care, and general practice in the evaluation and assessment of musculoskeletal disease. It provides the following functionality: • Segmentation of vertebras • Labelling of vertebras • Measurements of heights in each vertebra and indication if they are critically different • Measurement of mean Hounsfield value in volume of interest within vertebra. Only DICOM images of adult patients are considered to be valid input.