FDA-cleared profile

Spine Auto Views

GE Medical Systems SCS · CT/CTA · FDA-cleared (510(k))

Authorized purpose: Use with CT images to automatically generate anatomically focused multi-planar spine reformats, apply vertebral-body and intervertebral-disc labels, and export the results to predetermined DICOM destinations for review.

Clinical role
Post-acquisition automated reformat generation, labeling, and export for clinician review, replacing manual generation of the described planes.
Images and patients
CT/CTA · Spine
Key limitation
Keep in mind: The FDA indication is not disease-specific and is intended for multiple care areas. The subject device has no user interface or measurement tools in the described comparison, and its outputs are reformats and labels for interpreting-physician review rather than a standalone diagnostic claim.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: Use with CT images to automatically generate anatomically focused multi-planar spine reformats, apply vertebral-body and intervertebral-disc labels, and export the results to predetermined DICOM destinations for review.
  • Which images or clinical context are in scope?Modality: CT/CTA Anatomy or clinical context: Spine Input: CT images of the spine.
  • Who uses it?Intended users: Clinicians and interpreting physicians.
  • What does the software produce?Output: Oblique axial reformats through each disc space, 3D curved coronal and curved sagittal views, traditional axial/coronal/sagittal planes, vertebral-body and disc-space labels, and automatically exported series to desired DICOM destinations such as PACS.
  • What is its most important limitation?Key limitation: The FDA indication is not disease-specific and is intended for multiple care areas. The subject device has no user interface or measurement tools in the described comparison, and its outputs are reformats and labels for interpreting-physician review rather than a standalone diagnostic claim.

Clinical takeaway. Clinical role: Use with CT images to automatically generate anatomically focused multi-planar spine reformats, apply vertebral-body and intervertebral-disc labels, and export the results to predetermined DICOM destinations for review.

Evidence takeaway. Evidence from the reviewed source record: Engineering validation on retrospective CT examinations showed that the disc-position and orientation algorithm passed defined acceptance criteria. A reader study using clinical CT images assessed acceptance of automatically generated oblique axial reformats and curved coronal/sagittal batch reformats; oblique axial results were reported as user-acceptable greater than 95% of the time for all readers. This acceptance criterion is not sensitivity, specificity, AUC, or a general accuracy rate, and the reviewed text does not state the reader-study sample size.

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
Use with CT images to automatically generate anatomically focused multi-planar spine reformats, apply vertebral-body and intervertebral-disc labels, and export the results to predetermined DICOM destinations for review.
  • Exact FDA submission
  • Checked 2026-09-09
Workflow role
Post-acquisition automated reformat generation, labeling, and export for clinician review, replacing manual generation of the described planes.
  • Exact FDA submission
  • Checked 2026-09-09
Required input
CT images of the spine.
  • Exact FDA submission
  • Checked 2026-09-09
Output and clinician action
Oblique axial reformats through each disc space, 3D curved coronal and curved sagittal views, traditional axial/coronal/sagittal planes, vertebral-body and disc-space labels, and automatically exported series to desired DICOM destinations such as PACS.
  • Exact FDA submission
  • Checked 2026-09-09
Care setting
Multiple CT review care areas, including trauma, oncology, and routine body examinations; the FDA packet does not specify a facility type.
  • Exact FDA submission
  • Checked 2026-09-09
Intended users
Clinicians and interpreting physicians.
  • Exact FDA submission
  • Checked 2026-09-09
Limitations
The FDA indication is not disease-specific and is intended for multiple care areas. The subject device has no user interface or measurement tools in the described comparison, and its outputs are reformats and labels for interpreting-physician review rather than a standalone diagnostic claim.
  • Exact FDA submission
  • Checked 2026-09-09
Evidence summary
Engineering validation on retrospective CT examinations showed that the disc-position and orientation algorithm passed defined acceptance criteria. A reader study using clinical CT images assessed acceptance of automatically generated oblique axial reformats and curved coronal/sagittal batch reformats; oblique axial results were reported as user-acceptable greater than 95% of the time for all readers. This acceptance criterion is not sensitivity, specificity, AUC, or a general accuracy rate, and the reviewed text does not state the reader-study sample size.
  • Exact FDA submission
  • Checked 2026-09-09

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
Engineering validation on retrospective CT examinations showed that the disc-position and orientation algorithm passed defined acceptance criteria. A reader study using clinical CT images assessed acceptance of automatically generated oblique axial reformats and curved coronal/sagittal batch reformats; oblique axial results were reported as user-acceptable greater than 95% of the time for all readers. This acceptance criterion is not sensitivity, specificity, AUC, or a general accuracy rate, and the reviewed text does not state the reader-study sample size.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-09
Reported sensitivity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

Reported specificity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

Spine Auto Views regulatory performance study summary

  • DesignReader Study
  • PopulationPopulation described in the exact-submission ACR model card
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K223424
  • IndependenceIndependence not established

K223424

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K223424. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-09
Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not verified: Pricing information is recorded but not verified against a resolving source.

Reimbursement

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: CT images of the spine.

    Scope: Exact FDA submission.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-09-09 · FDA K223424: Spine Auto Views

  2. Routing

    Source context, not route validation: The FDA packet states that generated series can be automatically exported to predetermined DICOM destinations, including PACS examples; exact supported objects, interface versions, routing, and local compatibility are not established.

    Scope: Exact FDA submission.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-09-09 · FDA K223424: Spine Auto Views

  3. AI processing

    Unresolved route: Not established for the exact route.

    Scope: Exact FDA submission. The reviewed sources do not establish the current exact-release hosting topology, data flow, region, tenancy, or disaster-recovery configuration.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-09-09 ·

  4. Returned objects

    Source context, not route validation: Oblique axial reformats through each disc space, 3D curved coronal and curved sagittal views, traditional axial/coronal/sagittal planes, vertebral-body and disc-space labels, and automatically exported series to desired DICOM destinations such as PACS.

    Scope: Exact FDA submission.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-09 · FDA K223424: Spine Auto Views

  5. Viewer/worklist/report

    Unresolved route: Oblique axial reformats through each disc space, 3D curved coronal and curved sagittal views, traditional axial/coronal/sagittal planes, vertebral-body and disc-space labels, and automatically exported series to desired DICOM destinations such as PACS.

    Scope: Exact FDA submission.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-09 · FDA K223424: Spine Auto Views

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Integration
The FDA packet states that generated series can be automatically exported to predetermined DICOM destinations, including PACS examples; exact supported objects, interface versions, routing, and local compatibility are not established.
  • Exact FDA submission
  • Checked 2026-09-09

Input or output interoperability

DICOM is explicitly referenced in the described input, output, or workflow.

exact submission · checked 2026-09-01

The model card does not establish every supported DICOM object, transfer method, or local interface.

Result destination

PACS is explicitly referenced as a result destination.

exact submission · checked 2026-09-01

Confirm supported products, versions, routing behavior, and display requirements during local evaluation.

Supported acquisition environment

Scanner manufacturers: All

exact submission · checked 2026-09-01

Confirm currently supported software versions, protocols, and site-specific acquisition configurations.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for Spine Auto Views; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-09
Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K223424
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
CT/CTA
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-09
Anatomy
Spine
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-09
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2023-07-13
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
21Fields reviewed
9Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not verified - checked 2026-09-09 - Exact FDA submission

    No public price or quote for the exact K223424 configuration was established in this review.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Reimbursement: Not publicly available - checked 2026-09-09 - Exact FDA submission

    No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

  • Deployment: Not verified - checked 2026-09-09 - Exact FDA submission

    The reviewed sources do not establish the current exact-release hosting topology, data flow, region, tenancy, or disaster-recovery configuration.

  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Security: Not verified - checked 2026-09-09 - Exact FDA submission

    Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.

  • Installation timeline: Not yet reviewed
  • Training: Not verified - checked 2026-09-09 - Exact FDA submission

    Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

  • Monitoring: Not verified - checked 2026-09-09 - Exact FDA submission

    Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

7 sources - 3 history items

Source ledger

  1. FDA K223424: Spine Auto Views

    U.S. Food and Drug Administration - Published 2023-07-13 - Accessed 2026-09-09

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. Spine Auto Views model card (K223424)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, output, limitations, evidenceSummary, integration, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K223424

    U.S. Food and Drug Administration - Accessed 2026-09-09

    Scope: postmarketSafety

Research history

  1. K223424 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K242295

BunkerHill BMD

BunkerHill Health

Authorized purpose: Analyzes existing CT images in DICOM format from adults aged 30 years and older to estimate average areal BMD of spinal bones and flag low bone density below a preset threshold, including opportunistic scans without a phantom; it does not replace DXA or other dedicated BMD testing.

Modality
CT/CTA
Anatomy
Spine
Decision
2025-04-08
FDA-cleared K222361

AI-Rad Companion (Musculoskeletal)

Siemens Medical Solutions USA, Inc.

AI-Rad Companion (Musculoskeletal) is an image processing software that provides quantitative and qualitative analysis from previously acquired Computed Tomography DICOM images to support radiologists and physicians from emergency medicine, specialty care, urgent care, and general practice in the evaluation and assessment of musculoskeletal disease. It provides the following functionality: • Segmentation of vertebras • Labelling of vertebras • Measurements of heights in each vertebra and indication if they are critically different • Measurement of mean Hounsfield value in volume of interest within vertebra. Only DICOM images of adult patients are considered to be valid input.

Modality
CT/CTA
Anatomy
Spine
Decision
2022-10-20
FDA-cleared K193267

Al-Rad Companion (Musculoskeletal)

Siemens Medical Solutions USA, Inc.

Authorized purpose: Provides quantitative and qualitative analysis of previously acquired CT DICOM images to support radiologists and physicians in musculoskeletal-disease assessment.

Modality
CT/CTA
Anatomy
Spine
Decision
2020-03-16

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination

Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.

Accepted repository change history

No entry has been added to this scoped change history for this record. This is not proof that the record never changed.

Correction policy and limits · Public repository change history