Authorized purpose: Provides a softcopy review environment for screening and diagnostic digital breast tomosynthesis, including visualization and image-enhancement aids.
Clinical role
DBT viewing software that displays two views of the same breast and, when the user points the cursor at tissue in one view, indicates the matching tissue in the other; the interpreting physician makes the diagnosis.
Images and patients
Mammography/DBT · Breast
Key limitation
Keep in mind: Supports only DBT input from the specified Hologic and GE systems; MR and ultrasound are not within the reviewed input scope.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: Provides a softcopy review environment for screening and diagnostic digital breast tomosynthesis, including visualization and image-enhancement aids.
Which images or clinical context are in scope?Modality: Mammography/DBT Anatomy or clinical context: Breast Input: Two CC/MLO DBT views of the same breast from specified compatible systems.
Who uses it?Intended users: Qualified interpreting physicians.
What does the software produce?Output: Matched and correlated tissue indication between views with cursor-guided registration.
What is its most important limitation?Key limitation: Supports only DBT input from the specified Hologic and GE systems; MR and ultrasound are not within the reviewed input scope.
Clinical takeaway. Clinical role: Provides a softcopy review environment for screening and diagnostic digital breast tomosynthesis, including visualization and image-enhancement aids.
Evidence takeaway. Evidence from the reviewed source record: No clinical test was conducted for the subject device; FDA reported test data from 28 patients, 34 CC-MLO pairs, and 68 directional matches, with aggregate matching frequency 58/68 (85.3%).
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
Provides a softcopy review environment for screening and diagnostic digital breast tomosynthesis, including visualization and image-enhancement aids.
Exact FDA submission
Checked 2026-09-15
Clinical-use display mirrors the exact FDA-supported purpose reconciled in this remediation finding.
DBT viewing software that displays two views of the same breast and, when the user points the cursor at tissue in one view, indicates the matching tissue in the other; the interpreting physician makes the diagnosis.
Exact FDA submission
Checked 2026-11-14
Packet wording: "ViewFinder is software which displays two Digital Breast Tomosynthesis (DBT) views of the same breast and dynamically indicates correlated (matched) tissue." … "Users operate the device by pointing the cursor at tissue in one view and ViewFinder indicates the matching tissue in the other view.". Read from the exact FDA decision document on 2026-11-14.
Screening and diagnostic digital breast tomosynthesis reading by qualified interpreting physicians; the FDA document does not specify a facility type.
Exact FDA submission
Checked 2026-11-14
Packet wording: "ViewFinder is a dedicated softcopy review environment for both screening and diagnostic digital breast tomosynthesis." … "optimized to support qualified interpreting physicians in both screening and diagnostic reading.". Read from the exact FDA decision document on 2026-11-14.
No clinical test was conducted for the subject device; FDA reported test data from 28 patients, 34 CC-MLO pairs, and 68 directional matches, with aggregate matching frequency 58/68 (85.3%).
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
No clinical test was conducted for the subject device; FDA reported test data from 28 patients, 34 CC-MLO pairs, and 68 directional matches, with aggregate matching frequency 58/68 (85.3%).
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
ViewFinder Software Version 1.1 regulatory performance study summary
DesignStand-Alone Performance
PopulationPopulation described in the exact-submission ACR model card
Scopeexact submission
Tested versionExact product version not reported; regulatory study summarized for FDA submission K223501
Samplen=28
IndependenceIndependence not established
K223501
ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K223501. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: Two CC/MLO DBT views of the same breast from specified compatible systems.
Scope: Exact FDA submission. Exact FDA input and compatibility scope.
Confirm scanner/model, protocol, series completeness and both software builds.
Source context, not route validation: The exact FDA package describes DICOM 3.0 support for ViewFinder, with DICOM-compatible modalities of MG Tomo (mammography tomosynthesis images) and DICOM DSR (Deformable Spatial Registration), an additional conformant modality storing tissue correlation. Interface versions and local compatibility require confirmation.
Scope: Exact FDA submission. Packet wording: "DICOM 3.0 Yes Yes Equivalent None DICOM MG (Digital The subject None compatible Mammography X-Ray) has an" … "modalities MG Tomo MG Tomo additional (Mammography (Mammography conformant Tomosynthesis images) Tomosynthesis images) modality DICOM DSR MR (Magnetic storing tissue (Deformable Spatial Resonance) correlation. Registration) US (Ultrasound)". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Source context, not route validation: The exact FDA package describes ViewFinder as software that runs on a PC in a compatible client/server environment with monitors cleared for mammography, as a standalone application or integrated into medical image management and processing systems. Current hosting, data flow, and commercial configuration require vendor confirmation.
Scope: Exact FDA submission. Packet wording: "ViewFinder runs on a PC and can be used for digital breast tomosynthesis image review together with monitors cleared for mammography diagnostics." … "ViewFinder Software is a software-only solution to run on compatible client server environment together with monitor displays cleared for Mammography.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
Confirm processing location, queue ownership, exclusions and exact AI build.
Unresolved route: Matched and correlated tissue indication between views with cursor-guided registration.
Scope: Exact FDA submission. FDA output description.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
The exact FDA package describes ViewFinder as software that runs on a PC in a compatible client/server environment with monitors cleared for mammography, as a standalone application or integrated into medical image management and processing systems. Current hosting, data flow, and commercial configuration require vendor confirmation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "ViewFinder runs on a PC and can be used for digital breast tomosynthesis image review together with monitors cleared for mammography diagnostics." … "ViewFinder Software is a software-only solution to run on compatible client server environment together with monitor displays cleared for Mammography.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
The exact FDA package describes DICOM 3.0 support for ViewFinder, with DICOM-compatible modalities of MG Tomo (mammography tomosynthesis images) and DICOM DSR (Deformable Spatial Registration), an additional conformant modality storing tissue correlation. Interface versions and local compatibility require confirmation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "DICOM 3.0 Yes Yes Equivalent None DICOM MG (Digital The subject None compatible Mammography X-Ray) has an" … "modalities MG Tomo MG Tomo additional (Mammography (Mammography conformant Tomosynthesis images) Tomosynthesis images) modality DICOM DSR MR (Magnetic storing tissue (Deformable Spatial Resonance) correlation. Registration) US (Ultrasound)". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
ACR AI Central provides an exact-submission model card for ViewFinder Software Version 1.1; FDA labeling remains controlling for clinical use, limitations, and performance.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Exact FDA record reviewedStatus
2026-09-20Last searched
25Fields reviewed
10Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, fda decision summary packet, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not verified - checked 2026-09-09 - Exact FDA submission
No public price or quote for the exact K223501 configuration was established in this review.
Pricing model: Not yet reviewed
Setup and integration fees: Not yet reviewed
Recurring fees: Not yet reviewed
Minimum commitment: Not yet reviewed
Contract duration: Not yet reviewed
Renewal terms: Not yet reviewed
Cancellation terms: Not yet reviewed
Support and service terms: Not yet reviewed
Trial or pilot terms: Not yet reviewed
Reimbursement: Not publicly available - checked 2026-09-09 - Exact FDA submission
No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.
Data hosting location: Not yet reviewed
Data retention: Not yet reviewed
Documented data use: Not yet reviewed
Hardware requirements: Not yet reviewed
Network requirements: Not yet reviewed
Identity and access: Not yet reviewed
Security: Not verified - checked 2026-09-09 - Exact FDA submission
Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.
Installation timeline: Not yet reviewed
Training: Not verified - checked 2026-09-09 - Exact FDA submission
Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.
Monitoring: Not verified - checked 2026-09-09 - Exact FDA submission
Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.
Authorized purpose: Flags calcifications, masses, asymmetries and architectural distortions on compatible FFDM or DBT screening examinations; management must not be based solely on MammoScreen output.
Authorized purpose: Tags screening cases as suspicious or blank to aid worklist prioritization; radiologists must review every examination and it does not mark original images.
Authorized purpose: Lunit INSIGHT MMG v1.1.10 aids physicians in detection, localization and characterization of suspicious breast-cancer areas on female screening mammograms from compatible FFDM systems, viewed after the physician's initial read as an adjunct to complete clinical interpretation.
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.
Accepted repository change history
No entry has been added to this scoped change history for this record. This is not proof that the record never changed.