Authorized purpose: Venue is a mobile point-of-care diagnostic ultrasound system for broad adult and pediatric applications in emergency, critical-care, office, clinic, and hospital settings; K180599 adds transesophageal support, ECG synchronization, and an automated eFAST navigation and review workflow.
Clinical role
Broad point-of-care ultrasound acquisition, measurement, reporting, and protocol navigation.
Images and patients
Ultrasound · Multi-region
Key limitation
Keep in mind: ECG is not intended for monitoring or diagnosis.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: Venue is a mobile point-of-care diagnostic ultrasound system for broad adult and pediatric applications in emergency, critical-care, office, clinic, and hospital settings; K180599 adds transesophageal support, ECG synchronization, and an automated eFAST navigation and review workflow.
Which images or clinical context are in scope?Modality: Ultrasound Anatomy or clinical context: Multi-region Input: Live ultrasound data from supported probes plus optional ECG input for image synchronization.
Who uses it?Intended users: qualified healthcare professionals
What does the software produce?Output: Ultrasound images, measurements, reports, DICOM studies, and eFAST navigation or review information.
What is its most important limitation?Key limitation: ECG is not intended for monitoring or diagnosis.
Clinical takeaway. Clinical role: Venue is a mobile point-of-care diagnostic ultrasound system for broad adult and pediatric applications in emergency, critical-care, office, clinic, and hospital settings; K180599 adds transesophageal support, ECG synchronization, and an automated eFAST navigation and review workflow.
Evidence takeaway. Evidence from the reviewed source record: K180599 reports system, software, acoustic-output, biocompatibility, cleaning, electrical, electromagnetic, mechanical, DICOM, and workflow verification. Clinical studies were not required and no numerical AI performance result is disclosed.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
Venue is a mobile point-of-care diagnostic ultrasound system for broad adult and pediatric applications in emergency, critical-care, office, clinic, and hospital settings; K180599 adds transesophageal support, ECG synchronization, and an automated eFAST navigation and review workflow.
Exact FDA submission
Checked 2026-09-19
Clinical-use display mirrors the exact FDA-supported purpose reconciled from the decision-summary packet.
The exact indications use the general user class qualified healthcare professionals and do not narrow it to physicians or point-of-care ultrasound professionals.
K180599 reports system, software, acoustic-output, biocompatibility, cleaning, electrical, electromagnetic, mechanical, DICOM, and workflow verification. Clinical studies were not required and no numerical AI performance result is disclosed.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
K180599 reports system, software, acoustic-output, biocompatibility, cleaning, electrical, electromagnetic, mechanical, DICOM, and workflow verification. Clinical studies were not required and no numerical AI performance result is disclosed.
Sensitivity is not the primary applicable endpoint for this acquisition, reconstruction, segmentation, planning, guidance, or quantitative-biomarker function. Use the task-specific endpoints and tested population reported below.
Reported specificity
Not applicable
Not applicable
Not applicable
Checked 2026-09-03
Specificity is not the primary applicable endpoint for this acquisition, reconstruction, segmentation, planning, guidance, or quantitative-biomarker function. Use the task-specific endpoints and tested population reported below.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
K180599 FDA performance package for Venue
DesignManufacturer nonclinical ultrasound-system, software, probe, DICOM, cleaning, and safety verification
PopulationRepresentative emergency, critical-care, and broad point-of-care ultrasound applications
Scopeexact submission
Tested versionVenue release cleared in K180599
IndependenceSponsor or vendor study
K180599
Manufacturer evidence submitted for substantial equivalence; local clinical, technical, and workflow acceptance remains necessary.
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the device within the applicable imaging or surgical service line and verify payer, site-of-service, and coding assumptions independently.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: Live ultrasound data from supported probes plus optional ECG input for image synchronization.
Scope: Exact FDA submission. Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.
Confirm scanner/model, protocol, series completeness and both software builds.
Source context, not route validation: Mobile battery-powered ultrasound system with supported probes, cleanable touch interface, articulated display, and network options.
Scope: Exact FDA submission.
Confirm processing location, queue ownership, exclusions and exact AI build.
Unresolved route: Ultrasound images, measurements, reports, DICOM studies, and eFAST navigation or review information.
Scope: Exact FDA submission. Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
GE HealthCare's Product Security Portal is described as a credentialed global portal giving customers security updates, vulnerability status and required actions for GE HealthCare products potentially impacted by critical vulnerabilities, plus historical patch views; GE HealthCare references its Cybersecurity Risk Management System and publishes a Coordinated Vulnerability Disclosure Statement for reporting vulnerabilities in GE HealthCare products.
Vendor-wide
Checked 2026-10-26
Vendor statement, not an attestation. Identity bridge: GE HealthCare's own Privacy Shield page on gehealthcare.com names GE Medical Systems, Ultrasound & Primary Care Diagnostics LLC as a US entity of the GEHC group. Corporate program statement; portal content is credentialed; exact-release scope is not established.
Only qualified physicians and trained point-of-care ultrasound professionals. should operate or interpret the system. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track the limits that matter for this product: ECG is for image synchronization, not monitoring or diagnosis. K180599 predates later Venue nerve, B-line, EF, and other AI functions; those later capabilities must not be attributed to this release.
Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. ECG is for image synchronization, not monitoring or diagnosis. K180599 predates later Venue nerve, B-line, EF, and other AI functions; those later capabilities must not be attributed to this release.
exact submission · checked 2026-09-03
Define stop-use, escalation, rollback, and revalidation triggers before production use.
Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.
product family · checked 2026-09-03
Vendor productivity claims are scenario inputs, not guaranteed local savings.
Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track the limits that matter for this product: ECG is for image synchronization, not monitoring or diagnosis. K180599 predates later Venue nerve, B-line, EF, and other AI functions; those later capabilities must not be attributed to this release.
exact submission · checked 2026-09-03
Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Exact FDA record reviewedStatus
2026-09-20Last searched
27Fields reviewed
11Source classes checked
20Screened PubMed leads
10Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains 20 PubMed and 10 ClinicalTrials.gov candidate records. Discovery metadata alone does not establish exact tested-product, version, population, and endpoint applicability, so all candidates remain quarantined and none was promoted. The current discovery snapshot contains 20 PubMed and 10 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, vendor product page, reimbursement policy, fda device recall, fda decision summary packet, literature index, trial registry, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not publicly available - checked 2026-09-03 - Exact FDA submission
No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, enabled features, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.
Authorized purpose: Aplio i900, i800, and i700 software V9.0 are mobile diagnostic ultrasound systems for multi-region imaging. K252074 adds deep-learning 3rd Harmonic Imaging for abdominal images; with specified endoscopic transducers, the Aplio i800/E3 also supports imaging of the upper gastrointestinal tract, surrounding organs, airways, tracheobronchial tree, and esophagus.
Ge Medical Systems Ultrasound and Primary Care Diagnostics
Authorized purpose: LOGIQ E10s is a general-purpose diagnostic ultrasound system supporting the labeled adult, pediatric, cardiac, vascular, abdominal, obstetric, small-organ, musculoskeletal, urologic, and procedural applications.
Ge Medical Systems Ultrasound and Primary Care Diagnostics
Authorized purpose: LOGIQ E10 is a general diagnostic ultrasound system for qualified users across fetal and obstetric, abdominal, gynecologic, pediatric, small-organ, breast, cardiac, vascular, musculoskeletal, urologic, transrectal, transvaginal, transesophageal, and intraoperative applications.
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.
Accepted repository change history
No entry has been added to this scoped change history for this record. This is not proof that the record never changed.