FDA-cleared profile

VEA Align; spineEOS

EOS imaging · X-ray/Radiography · FDA-cleared (510(k))

Authorized purpose: VEA Align is cloud-based software for orthopedic applications in adult and pediatric populations using two-dimensional X-ray images from EOS imaging systems to display interactive landmarks and assess patient-specific global alignment, with angle and length measurements for spinal deformities, degenerative diseases, and lower-limb alignment disorders or deformities. It is suitable for patients aged 7 years and older. spineEOS assists healthcare professionals with preoperative planning of spine surgery using EOS images, associated three-dimensional datasets, measurements, and patient-specific surgical-planning tools.

Clinical role
Radiographic image acquisition support, image processing, measurement, planning, or concurrent review under professional control.
Images and patients
X-ray/Radiography · Spine and lower-limb alignment; frontal and lateral EOS images and associated 3D datasets
Key limitation
Keep in mind: Use is limited to orthopedic alignment and spine-planning workflows for patients aged 7 years and older, and clinical judgment and experience are required.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: VEA Align is cloud-based software for orthopedic applications in adult and pediatric populations using two-dimensional X-ray images from EOS imaging systems to display interactive landmarks and assess patient-specific global alignment, with angle and length measurements for spinal deformities, degenerative diseases, and lower-limb alignment disorders or deformities. It is suitable for patients aged 7 years and older. spineEOS assists healthcare professionals with preoperative planning of spine surgery using EOS images, associated three-dimensional datasets, measurements, and patient-specific surgical-planning tools.
  • Which images or clinical context are in scope?Modality: X-ray/Radiography Anatomy or clinical context: Spine and lower-limb alignment; frontal and lateral EOS images and associated 3D datasets Input: Supported X-ray, radiographic, EOS, dental, or other labeled image inputs and associated study data.
  • Who uses it?Intended users: Radiologists, dentists, orthopedic clinicians, and other trained radiographic professionals within the labeled use.
  • What does the software produce?Output: Processed radiographs, measurements, landmarks, annotations, planning views, and review information for professional use.
  • What is its most important limitation?Key limitation: Use is limited to orthopedic alignment and spine-planning workflows for patients aged 7 years and older, and clinical judgment and experience are required.

Clinical takeaway. Clinical role: VEA Align is cloud-based software for orthopedic applications in adult and pediatric populations using two-dimensional X-ray images from EOS imaging systems to display interactive landmarks and assess patient-specific global alignment, with angle and length measurements for spinal deformities, degenerative diseases, and lower-limb alignment disorders or deformities. It is suitable for patients aged 7 years and older. spineEOS assists healthcare professionals with preoperative planning of spine surgery using EOS images, associated three-dimensional datasets, measurements, and patient-specific surgical-planning tools.

Evidence takeaway. Evidence from the reviewed source record: The exact FDA document reports a standalone validation of the VEA Align AI algorithm on a dedicated test data set of 361 patients, separate from the training data, with ground-truth EOS 3DServices reconstructions from sterEOS. Acceptance criteria were Median error ≤ 3 mm and 3rd Quartile ≤ 5 mm; the document states that direct comparison of skeletal landmark locations between the subject and predicate devices met these criteria. Results are as reported in the FDA submission, not independent validation.

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
VEA Align is cloud-based software for orthopedic applications in adult and pediatric populations using two-dimensional X-ray images from EOS imaging systems to display interactive landmarks and assess patient-specific global alignment, with angle and length measurements for spinal deformities, degenerative diseases, and lower-limb alignment disorders or deformities. It is suitable for patients aged 7 years and older. spineEOS assists healthcare professionals with preoperative planning of spine surgery using EOS images, associated three-dimensional datasets, measurements, and patient-specific surgical-planning tools.
  • Exact FDA submission
  • Checked 2026-09-04
Workflow role
Radiographic image acquisition support, image processing, measurement, planning, or concurrent review under professional control.
  • Exact FDA submission
  • Checked 2026-09-04
Required input
Supported X-ray, radiographic, EOS, dental, or other labeled image inputs and associated study data.
  • Exact FDA submission
  • Checked 2026-09-04
Output and clinician action
Processed radiographs, measurements, landmarks, annotations, planning views, and review information for professional use.
  • Exact FDA submission
  • Checked 2026-09-04
Care setting
Radiographic, orthopedic, dental, or clinical image-review settings within the labeled scope.
  • Exact FDA submission
  • Checked 2026-09-04
Intended users
Radiologists, dentists, orthopedic clinicians, and other trained radiographic professionals within the labeled use.
  • Exact FDA submission
  • Checked 2026-09-04
Limitations
Use is limited to orthopedic alignment and spine-planning workflows for patients aged 7 years and older, and clinical judgment and experience are required.
  • Exact FDA submission
  • Checked 2026-09-04
Evidence summary
The exact FDA document reports a standalone validation of the VEA Align AI algorithm on a dedicated test data set of 361 patients, separate from the training data, with ground-truth EOS 3DServices reconstructions from sterEOS. Acceptance criteria were Median error ≤ 3 mm and 3rd Quartile ≤ 5 mm; the document states that direct comparison of skeletal landmark locations between the subject and predicate devices met these criteria. Results are as reported in the FDA submission, not independent validation.
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "To assess the standalone performance of the AI algorithm of the VEA Align, the test was performed with:" … "A dedicated test data set containing different data from the training data set.". Read from the exact FDA decision document on 2026-11-14.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
The exact FDA document reports a standalone validation of the VEA Align AI algorithm on a dedicated test data set of 361 patients, separate from the training data, with ground-truth EOS 3DServices reconstructions from sterEOS. Acceptance criteria were Median error ≤ 3 mm and 3rd Quartile ≤ 5 mm; the document states that direct comparison of skeletal landmark locations between the subject and predicate devices met these criteria. Results are as reported in the FDA submission, not independent validation.
  • FDA source
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "To assess the standalone performance of the AI algorithm of the VEA Align, the test was performed with:" … "A dedicated test data set containing different data from the training data set.". Read from the exact FDA decision document on 2026-11-14.

Reported sensitivity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-04

Sensitivity is not promoted as a product-wide endpoint. Compare only an exact endpoint, threshold, population, version, study design, and source if separately verified.

Reported specificity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-04

Specificity is not promoted as a product-wide endpoint. Compare only an exact endpoint, threshold, population, version, study design, and source if separately verified.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

K251747 FDA decision evidence package for VEA Align; spineEOS

  • DesignFDA decision evidence and product verification described in the exact submission or PMA decision record.
  • PopulationSpine and lower-limb alignment; frontal and lateral EOS images and associated 3D datasets
  • Scopeexact submission
  • Tested versionVEA Align; spineEOS configuration described in K251747
  • IndependenceSponsor or vendor study

K251747

FDA decision material is manufacturer-submitted regulatory evidence. It is not independent clinical validation unless the exact source expressly establishes that distinction; local technical, clinical, privacy, security, and workflow acceptance remains necessary.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K251747 by exact native submission identifier. This does not exclude product-family, later-version, or adverse-event information.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-04

Exact-identifier review does not prove absence of all product-family safety signals.

Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not publicly available: Reviewed sources do not publish an exact-product procurement price with a stated offer scope.

Reimbursement

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: Supported X-ray, radiographic, EOS, dental, or other labeled image inputs and associated study data.

    Scope: Exact FDA submission.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-09-04 · FDA 510(k) decision material K251747: VEA Align; spineEOS

  2. Routing

    Source context, not route validation: The exact FDA package describes EOS 2D X-ray images, associated 3D datasets, landmarks, measurements, and spine-planning outputs; verify image identity, routing, dataset compatibility, and local archive behavior.

    Scope: Exact FDA submission. This is a partial interface boundary described in the exact FDA package; current local compatibility, routing, identity, archive, and downtime behavior still require acceptance testing.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-09-04 · FDA 510(k) decision material K251747: VEA Align; spineEOS

  3. AI processing

    Source context, not route validation: The exact FDA indication identifies VEA Align as cloud-based; current hosting, contracts, security, availability, and site configuration remain unknown.

    Scope: Exact FDA submission. This is a partial release-level topology fact from the exact FDA package; current hosting, contracts, security, availability, and site configuration remain separate evidence.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-09-04 · FDA 510(k) decision material K251747: VEA Align; spineEOS

  4. Returned objects

    Source context, not route validation: Processed radiographs, measurements, landmarks, annotations, planning views, and review information for professional use.

    Scope: Exact FDA submission.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-04 · FDA 510(k) decision material K251747: VEA Align; spineEOS

  5. Viewer/worklist/report

    Unresolved route: Processed radiographs, measurements, landmarks, annotations, planning views, and review information for professional use.

    Scope: Exact FDA submission.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-04 · FDA 510(k) decision material K251747: VEA Align; spineEOS

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Deployment
The exact FDA indication identifies VEA Align as cloud-based; current hosting, contracts, security, availability, and site configuration remain unknown.
  • Exact FDA submission
  • Checked 2026-09-04

This is a partial release-level topology fact from the exact FDA package; current hosting, contracts, security, availability, and site configuration remain separate evidence.

Integration
The exact FDA package describes EOS 2D X-ray images, associated 3D datasets, landmarks, measurements, and spine-planning outputs; verify image identity, routing, dataset compatibility, and local archive behavior.
  • Exact FDA submission
  • Checked 2026-09-04

This is a partial interface boundary described in the exact FDA package; current local compatibility, routing, identity, archive, and downtime behavior still require acceptance testing.

Training
Only radiologists, dentists, orthopedic clinicians, and other trained radiographic professionals within the labeled use should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
  • Exact FDA submission
  • Checked 2026-09-04
Monitoring
Monitor K251747 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Use is limited to orthopedic alignment and spine-planning workflows for patients aged 7 years and older, and clinical judgment and experience are required.
  • Exact FDA submission
  • Checked 2026-09-04

This is buyer-facing monitoring guidance derived from the exact cleared scope; it is not a claim that the FDA package provides a live monitoring service.

Release and scope verification

Map installed model, build, options, population, anatomy, inputs, outputs, and enabled functions to K251747 before acceptance.

exact submission · checked 2026-09-04

Later product-family features are not evidence that this cleared or authorized release includes them.

Deployment boundary

The exact FDA indication identifies VEA Align as cloud-based; current hosting, contracts, security, availability, and site configuration remain unknown.

exact submission · checked 2026-09-04

FDA-described topology does not establish current hosting contracts, security controls, uptime, or local qualification.

Interoperability acceptance

The exact FDA package describes EOS 2D X-ray images, associated 3D datasets, landmarks, measurements, and spine-planning outputs; verify image identity, routing, dataset compatibility, and local archive behavior.

exact submission · checked 2026-09-04

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Use is limited to orthopedic alignment and spine-planning workflows for patients aged 7 years and older, and clinical judgment and experience are required.

exact submission · checked 2026-09-04

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

exact submission · checked 2026-09-04

No exact-product price or reimbursement determination was promoted.

Lifecycle monitoring

Monitor K251747 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Use is limited to orthopedic alignment and spine-planning workflows for patients aged 7 years and older, and clinical judgment and experience are required.

exact submission · checked 2026-09-04

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes.

Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K251747
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
X-ray/Radiography
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Anatomy
Spine and lower-limb alignment; frontal and lateral EOS images and associated 3D datasets
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2025-08-15
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
27Fields reviewed
9Source classes checked
1Screened PubMed leads
1Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains 1 PubMed and 1 ClinicalTrials.gov candidate records. Discovery metadata alone does not establish exact tested-product, version, population, and endpoint applicability, so all candidates remain quarantined and none was promoted. The current discovery snapshot contains 1 PubMed and 1 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, fda decision material, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not publicly available - checked 2026-09-04 - Exact FDA submission

    No public list price for the exact cleared or authorized configuration was found. Obtain a written quote separating license or hardware basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Reimbursement: Not publicly available - checked 2026-09-04 - Exact FDA submission

    No exact-product reimbursement determination was promoted in this timeboxed review. Verify payer, site-of-service, coding, coverage, and local payment assumptions independently.

  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Security: Not publicly available - checked 2026-09-04 - Exact FDA submission

    Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.

  • Installation timeline: Not yet reviewed

5 sources - 3 history items

Source ledger

  1. FDA 510(k) decision material K251747: VEA Align; spineEOS

    U.S. Food and Drug Administration - Published 2025-08-15 - Accessed 2026-09-04

    Scope: identity, purpose, clinicalUse, modality, anatomy, intendedUsers, careSetting, workflowRole, input, output, limitations, evidenceSummary

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  4. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  5. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

Research history

  1. K251747 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K250914

MediAI-BA

Crescom Co., Ltd.

Authorized purpose: View and quantify bone age from 2D posteroanterior left-hand radiographs to aid pediatric radiologists assessing patients aged 2-18 years.

Modality
X-ray/Radiography
Anatomy
Left hand and wrist; skeletal maturity/bone age
Decision
2025-12-18
FDA-cleared K252086

DTX Studio Assist

Nobel Biocare C/O Medicim NV

Authorized purpose: Analyzes two-dimensional bitewing and periapical radiographs of permanent teeth in patients aged 15 and older to identify and localize caries, calculus, periapical radiolucency, root canal filling deficiency, discrepancy at an existing restoration margin, and bone loss; also supports alveolar bone-level measurement and restoration/dental-anatomy segmentation as an adjunct to dentist review.

Modality
X-ray/Radiography
Anatomy
Dental/Maxillofacial
Decision
2025-11-17
FDA-cleared K250367

CoLumboX

Smart Soft Healthcare AD

Authorized purpose: Post-processing segmentation, measurement, threshold labeling, and export of quantitative lumbar-spine measurements to support clinical review; not diagnosis.

Modality
X-ray/Radiography
Anatomy
Lumbar spine, including vertebrae and sacrum
Decision
2025-05-28
FDA-cleared K242437

Smile Dx®

Cube Click, Inc.

Authorized purpose: Aids dentists reviewing bitewing and periapical radiographs of permanent teeth by detecting/segmenting caries and periapical radiolucencies, identifying restorations and dental anatomy, and measuring mesial/distal bone levels; it does not replace dentist review.

Modality
X-ray/Radiography
Anatomy
Dental/Maxillofacial
Decision
2025-05-14

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination

Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.

Accepted repository change history

No entry has been added to this scoped change history for this record. This is not proof that the record never changed.

Correction policy and limits · Public repository change history