Authorized purpose: uDR Aurora CX acquires two-dimensional X-ray images of the labeled skull, spine, chest, abdomen, extremities, limbs, and trunk for review by qualified healthcare professionals; it is not a mammography system.
Clinical role
Digital radiographic image acquisition and release-specific image-processing support.
Images and patients
X-ray/Radiography · Skull, spine, chest, abdomen, extremities, limbs, and trunk
Key limitation
Keep in mind: Use only the cleared anatomy, detector, projection, patient-size, dose, and positioning limits; the system is not a mammography substitute unless explicitly labeled for that use.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: uDR Aurora CX acquires two-dimensional X-ray images of the labeled skull, spine, chest, abdomen, extremities, limbs, and trunk for review by qualified healthcare professionals; it is not a mammography system.
Which images or clinical context are in scope?Modality: X-ray/Radiography Anatomy or clinical context: Skull, spine, chest, abdomen, extremities, limbs, and trunk Input: X-ray exposures acquired with supported detectors, projections, patient positions, and protocols.
Who uses it?Intended users: Qualified radiologic technologists and interpreting physicians.
What does the software produce?Output: Digital radiographic images, processed views, measurements, and procedure records for professional review.
What is its most important limitation?Key limitation: Use only the cleared anatomy, detector, projection, patient-size, dose, and positioning limits; the system is not a mammography substitute unless explicitly labeled for that use.
Clinical takeaway. Clinical role: uDR Aurora CX acquires two-dimensional X-ray images of the labeled skull, spine, chest, abdomen, extremities, limbs, and trunk for review by qualified healthcare professionals; it is not a mammography system.
Evidence takeaway. Evidence from the reviewed source record: The exact FDA document reports non-clinical testing, including bench tests of key features. For uVision automatic positioning, a one-week sample had 14 of 328 chest cases and 0 of 20 stitching cases non-compliant; the summary states that in 95% of positioning processes the automatic settings met clinical requirements. For uAid, sensitivity and specificity for foreign body, lung field, and scapula checks all exceed 0.9. A board certified radiologist judged sample images sufficient for clinical diagnosis. Results are as reported in the FDA submission, not independent validation.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
uDR Aurora CX acquires two-dimensional X-ray images of the labeled skull, spine, chest, abdomen, extremities, limbs, and trunk for review by qualified healthcare professionals; it is not a mammography system.
Use only the cleared anatomy, detector, projection, patient-size, dose, and positioning limits; the system is not a mammography substitute unless explicitly labeled for that use.
The exact FDA document reports non-clinical testing, including bench tests of key features. For uVision automatic positioning, a one-week sample had 14 of 328 chest cases and 0 of 20 stitching cases non-compliant; the summary states that in 95% of positioning processes the automatic settings met clinical requirements. For uAid, sensitivity and specificity for foreign body, lung field, and scapula checks all exceed 0.9. A board certified radiologist judged sample images sufficient for clinical diagnosis. Results are as reported in the FDA submission, not independent validation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "Additional non-clinical tests are conducted for key features to ensure safe and effectiveness when integrated into the system: Feature Bench Testing Performed" … "Date Chest/case Case of Non- Full Spine or Full Case of Non-". Read from the exact FDA decision document on 2026-11-14.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
The exact FDA document reports non-clinical testing, including bench tests of key features. For uVision automatic positioning, a one-week sample had 14 of 328 chest cases and 0 of 20 stitching cases non-compliant; the summary states that in 95% of positioning processes the automatic settings met clinical requirements. For uAid, sensitivity and specificity for foreign body, lung field, and scapula checks all exceed 0.9. A board certified radiologist judged sample images sufficient for clinical diagnosis. Results are as reported in the FDA submission, not independent validation.
FDA source
Exact FDA submission
Checked 2026-11-14
Packet wording: "Additional non-clinical tests are conducted for key features to ensure safe and effectiveness when integrated into the system: Feature Bench Testing Performed" … "Date Chest/case Case of Non- Full Spine or Full Case of Non-". Read from the exact FDA decision document on 2026-11-14.
Sensitivity is not a product-wide endpoint for this system or software function. Compare only an exact endpoint, threshold, population, version, and source if one is later verified.
Reported specificity
Not applicable
Not applicable
Not applicable
Checked 2026-09-04
Specificity is not a product-wide endpoint for this system or software function. Compare only an exact endpoint, threshold, population, version, and source if one is later verified.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
K251167 FDA evidence package for uDR Aurora CX
DesignFDA substantial-equivalence decision with X-ray hardware, detector, image-quality, dose, safety, EMC, DICOM, and usability verification
PopulationAdult and pediatric radiographic examinations of the labeled skull, spine, chest, abdomen, extremities, limbs, and trunk
Scopeexact submission
Tested versionuDR Aurora CX configuration described in K251167
IndependenceSponsor or vendor study
K251167
Manufacturer evidence submitted for the exact FDA decision. It is not independent clinical validation, and local technical, clinical, privacy, security, and workflow acceptance remains necessary.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K251167 by exact native submission identifier. This does not exclude product-family, later, or adverse-event information.
Public source
Exact FDA submission
Checked 2026-09-04
Exact-identifier review does not prove absence of all product-family safety signals.
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging, procedural, or surgical service line and verify payer, site-of-service, and coding assumptions independently.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: X-ray exposures acquired with supported detectors, projections, patient positions, and protocols.
Scope: Exact FDA submission.
Confirm scanner/model, protocol, series completeness and both software builds.
Source context, not route validation: The exact FDA package describes the X-ray Imaging System generating DICOM-standard images by workstation and lists DICOM3 under standards for the proposed device. Interface versions and local compatibility require confirmation.
Scope: Exact FDA submission. Packet wording: "The X-ray Imaging System converts X-ray photons into electrical signals by detectors, and generates DICOM- standard images by workstation" … "Standards DICOM DICOM3 DICOM3 Same Power Source". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Source context, not route validation: The exact FDA package describes the uVision algorithm in the uDR Aurora CX digital medical X-ray imaging system and uAid as a system software function triggered after chest X-ray acquisition. Current hosting, data flow, and commercial configuration require vendor confirmation.
Scope: Exact FDA submission. Packet wording: "The uVision algorithm in the digital medical X-ray imaging system (uDR Aurora CX) aims to optimize the radiographic scanning workflow" … "uAid is triggered after the acquisition of chest X-ray images in patients aged over 20 years, which automatically evaluates image characteristics". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
Confirm processing location, queue ownership, exclusions and exact AI build.
Unresolved route: Digital radiographic images, processed views, measurements, and procedure records for professional review.
Scope: Exact FDA submission.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
The exact FDA package describes the uVision algorithm in the uDR Aurora CX digital medical X-ray imaging system and uAid as a system software function triggered after chest X-ray acquisition. Current hosting, data flow, and commercial configuration require vendor confirmation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "The uVision algorithm in the digital medical X-ray imaging system (uDR Aurora CX) aims to optimize the radiographic scanning workflow" … "uAid is triggered after the acquisition of chest X-ray images in patients aged over 20 years, which automatically evaluates image characteristics". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
The exact FDA package describes the X-ray Imaging System generating DICOM-standard images by workstation and lists DICOM3 under standards for the proposed device. Interface versions and local compatibility require confirmation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "The X-ray Imaging System converts X-ray photons into electrical signals by detectors, and generates DICOM- standard images by workstation" … "Standards DICOM DICOM3 DICOM3 Same Power Source". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
United Imaging Healthcare's cybersecurity page lists a secure development lifecycle and controls: firewall, encrypted connections, OS hardening, antivirus, patches, authentication, audit log, disk and transmission encryption. Company-level, not release-specific.
Vendor-wide
Checked 2026-10-25
Vendor statement, not an attestation; exact-release scope is not established.
Only qualified radiologic technologists and interpreting physicians should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Use only the cleared anatomy, detector, projection, patient-size, dose, and positioning limits; the system is not a mammography substitute unless explicitly labeled for that use.
Product family
Checked 2026-09-04
Current vendor material is product-family context for K251167; confirm exact cleared version, enabled options, deployment, and local compatibility.
Exact public integration details remain incomplete; establish and test the supported input, output, identity, routing, archive, and downtime contract locally.
product family · checked 2026-09-04
Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally. Current vendor material is product-family context; confirm exact release, enabled options, deployment, and local compatibility.
Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Use only the cleared anatomy, detector, projection, patient-size, dose, and positioning limits; the system is not a mammography substitute unless explicitly labeled for that use.
exact submission · checked 2026-09-04
Define stop-use, escalation, rollback, and revalidation triggers before production use.
Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.
product family · checked 2026-09-04
No exact-product price was found in the reviewed public material; productivity or value claims require local evidence.
Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Use only the cleared anatomy, detector, projection, patient-size, dose, and positioning limits; the system is not a mammography substitute unless explicitly labeled for that use.
product family · checked 2026-09-04
Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. Current vendor material is product-family context; confirm exact release, enabled options, deployment, and local compatibility.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Human reviewedStatus
2026-09-20Last searched
27Fields reviewed
14Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, vendor documentation, fda decision material, vendor product page, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not publicly available - checked 2026-09-04 - Exact FDA submission
No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.
Authorized purpose: Opportunistic review of frontal lumbar-spine or KUB radiographs in adults aged 50 years and older to flag possible low bone mineral density at L1-L4 for clinician follow-up; a negative result does not rule out low BMD.
Authorized purpose: Rayvolve supports clinicians by identifying suspected fractures, joint effusions, and dislocations on labeled radiographic views; it does not replace clinician interpretation.
Modality
X-ray/Radiography
Anatomy
Musculoskeletal and chest/ribs radiographs across specified views
Authorized purpose: Concurrent-read CADe to assist dentists in detecting specified tooth and jaw anatomy, bone-level landmarks, and root-canal-treated teeth in digital radiographs of permanent teeth in patients 15 years or older.
Modality
X-ray/Radiography
Anatomy
Permanent teeth and jaw/maxillofacial anatomy, including enamel, dentine, pulp, root, mental foramen, inferior alveolar canal, maxillary sinus, cementoenamel junction, crestal bone, and root-canal-treated teeth
Authorized purpose: qXR-Detect is an adult chest-radiograph concurrent reading aid that identifies, highlights and categorizes suspicious regions of interest across lung, pleura, mediastinum/hila/heart, bone, hardware and other categories; trained clinicians remain responsible for primary interpretation.
Modality
X-ray/Radiography
Anatomy
Chest/Lung; pleura, mediastinum/hila/heart, bone, hardware and other chest-radiograph regions
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.
Accepted repository change history
No entry has been added to this scoped change history for this record. This is not proof that the record never changed.