Authorized purpose: Concurrent reading aid for physicians interpreting screening FFDM and DBT to identify suspicious regions and assess malignancy likelihood.
Clinical role
Concurrent reading aid for physicians interpreting screening FFDM and DBT exams, outputting locations of calcification groups and soft-tissue regions with cancer-likelihood scores and an exam score; an adjunct tool, not a replacement for the physician's own review.
Images and patients
Mammography/DBT · Breast
Key limitation
Keep in mind: Adjunctive concurrent-reading aid; does not replace physician review, and patient management must not rely solely on the analysis.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: Concurrent reading aid for physicians interpreting screening FFDM and DBT to identify suspicious regions and assess malignancy likelihood.
Which images or clinical context are in scope?Modality: Mammography/DBT Anatomy or clinical context: Breast Input: Screening FFDM and DBT examinations in DICOM format.
Who uses it?Intended users: Physicians qualified to interpret screening mammography and DBT.
What does the software produce?Output: CAD marks, region scores from 0 to 100, links between corresponding regions and views, and an exam score from 1 to 10.
What is its most important limitation?Key limitation: Adjunctive concurrent-reading aid; does not replace physician review, and patient management must not rely solely on the analysis.
Clinical takeaway. Clinical role: Concurrent reading aid for physicians interpreting screening FFDM and DBT to identify suspicious regions and assess malignancy likelihood.
Evidence takeaway. Evidence from the reviewed source record: FDA-reported verification and independent multi-center testing covered 9,122 examinations, including 7,882 non-cancer and 1,240 cancer examinations across seven EU countries and the US.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
Concurrent reading aid for physicians interpreting screening FFDM and DBT to identify suspicious regions and assess malignancy likelihood.
Exact FDA submission
Checked 2026-09-19
Clinical-use display mirrors the exact FDA-supported purpose reconciled in this remediation finding.
Concurrent reading aid for physicians interpreting screening FFDM and DBT exams, outputting locations of calcification groups and soft-tissue regions with cancer-likelihood scores and an exam score; an adjunct tool, not a replacement for the physician's own review.
Exact FDA submission
Checked 2026-11-14
Packet wording: "Transpara® software is intended for use as a concurrent reading aid for physicians interpreting screening full-field digital mammography exams and digital breast tomosynthesis exams from compatible FFDM and DBT systems, to identify regions suspicious for breast cancer and assess their likelihood of malignancy." … "Output of the device includes locations of calcifications groups and soft-tissue regions, with scores indicating the likelihood that cancer is present, and an exam score". Read from the exact FDA decision document on 2026-11-14.
Women undergoing screening mammography and digital breast tomosynthesis; the device is for use in a healthcare facility or hospital.
Exact FDA submission
Checked 2026-11-14
Packet wording: "Transpara® is a software medical device for use in a healthcare facility or hospital with the following indications for use:" … "The device is intended to be used in the population of women undergoing screening mammography and digital breast tomosynthesis.". Read from the exact FDA decision document on 2026-11-14.
FDA-reported verification and independent multi-center testing covered 9,122 examinations, including 7,882 non-cancer and 1,240 cancer examinations across seven EU countries and the US.
Exact FDA submission
Checked 2026-09-19
Submission evidence only; not a generalized performance claim.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
FDA-reported verification and independent multi-center testing covered 9,122 examinations, including 7,882 non-cancer and 1,240 cancer examinations across seven EU countries and the US.
FDA source
Exact FDA submission
Checked 2026-09-19
Submission evidence only; not a generalized performance claim.
Values reported in the exact-submission materials. Consult the linked source for endpoint, threshold, population, and confidence interval before comparison.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
Transpara 1.7.0 regulatory performance study summary
DesignStand-Alone Performance
PopulationPopulation described in the exact-submission ACR model card
Scopeexact submission
Tested versionExact product version not reported; regulatory study summarized for FDA submission K210404
Samplen=9122
Sites7
IndependenceIndependence not established
K210404
ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K210404. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: Screening FFDM and DBT examinations in DICOM format.
Scope: Exact FDA submission. FDA packet input scope.
Confirm scanner/model, protocol, series completeness and both software builds.
Source context, not route validation: The exact FDA package describes a processing server that accepts mammograms or DBT exams in DICOM format and sends output via the DICOM protocol in a standardized mammography CAD DICOM format, commonly to medical workstations, PACS and RIS. Interface versions and local compatibility require confirmation.
Scope: Exact FDA submission. Packet wording: "which accepts mammograms or DBT exams in DICOM format as input, processes them, and sends the processing output to a destination using the DICOM protocol" … "in a standardized mammography CAD DICOM format. Common destinations are medical workstations, PACS and RIS.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Source context, not route validation: The exact FDA package describes Transpara as offered as a virtual machine that runs on pre-selected standard PC hardware as well as a dedicated virtual machine cluster, with results computed in a processing server. Current hosting, data flow, and commercial configuration require vendor confirmation.
Scope: Exact FDA submission. Packet wording: "Transpara® is offered as a virtual machine and runs on pre-selected standard PC hardware as well as a dedicated virtual machine cluster." … "Results of Transpara® are computed in processing server which accepts mammograms or DBT exams in DICOM format as input". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
Confirm processing location, queue ownership, exclusions and exact AI build.
Source context, not route validation: CAD marks, region scores from 0 to 100, links between corresponding regions and views, and an exam score from 1 to 10.
Scope: Exact FDA submission. FDA packet output scope.
Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.
Unresolved route: CAD marks, region scores from 0 to 100, links between corresponding regions and views, and an exam score from 1 to 10.
Scope: Exact FDA submission. FDA packet output scope.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
The exact FDA package describes Transpara as offered as a virtual machine that runs on pre-selected standard PC hardware as well as a dedicated virtual machine cluster, with results computed in a processing server. Current hosting, data flow, and commercial configuration require vendor confirmation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "Transpara® is offered as a virtual machine and runs on pre-selected standard PC hardware as well as a dedicated virtual machine cluster." … "Results of Transpara® are computed in processing server which accepts mammograms or DBT exams in DICOM format as input". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
The exact FDA package describes a processing server that accepts mammograms or DBT exams in DICOM format and sends output via the DICOM protocol in a standardized mammography CAD DICOM format, commonly to medical workstations, PACS and RIS. Interface versions and local compatibility require confirmation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "which accepts mammograms or DBT exams in DICOM format as input, processes them, and sends the processing output to a destination using the DICOM protocol" … "in a standardized mammography CAD DICOM format. Common destinations are medical workstations, PACS and RIS.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
ScreenPoint Medical's security reporting page states its security program is aligned with ISO/IEC 27001 and describes a vulnerability reporting channel for Transpara. It states alignment, not certification; no controls are listed. Vendor-wide.
Vendor-wide
Checked 2026-10-25
Vendor statement, not an attestation; exact-release scope is not established.
ACR AI Central provides an exact-submission model card for Transpara 1.7.0; FDA labeling remains controlling for clinical use, limitations, and performance.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Exact FDA record reviewedStatus
2026-09-20Last searched
26Fields reviewed
11Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, acr ai central product, vendor documentation, fda device recall, literature index, trial registry, fda decision summary packet, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not verified - checked 2026-09-09 - Exact FDA submission
No public price or quote for the exact K210404 configuration was established in this review.
Pricing model: Not yet reviewed
Setup and integration fees: Not yet reviewed
Recurring fees: Not yet reviewed
Minimum commitment: Not yet reviewed
Contract duration: Not yet reviewed
Renewal terms: Not yet reviewed
Cancellation terms: Not yet reviewed
Support and service terms: Not yet reviewed
Trial or pilot terms: Not yet reviewed
Reimbursement: Not publicly available - checked 2026-09-09 - Exact FDA submission
No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.
Data hosting location: Not yet reviewed
Data retention: Not yet reviewed
Documented data use: Not yet reviewed
Hardware requirements: Not yet reviewed
Network requirements: Not yet reviewed
Identity and access: Not yet reviewed
Installation timeline: Not yet reviewed
Training: Not verified - checked 2026-09-09 - Exact FDA submission
Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.
Monitoring: Not verified - checked 2026-09-09 - Exact FDA submission
Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.
Authorized purpose: Flags calcifications, masses, asymmetries and architectural distortions on compatible FFDM or DBT screening examinations; management must not be based solely on MammoScreen output.
Authorized purpose: Tags screening cases as suspicious or blank to aid worklist prioritization; radiologists must review every examination and it does not mark original images.
Authorized purpose: Lunit INSIGHT MMG v1.1.10 aids physicians in detection, localization and characterization of suspicious breast-cancer areas on female screening mammograms from compatible FFDM systems, viewed after the physician's initial read as an adjunct to complete clinical interpretation.
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.
Accepted repository change history