FDA-cleared profile

Swoop Point-of-Care Magnetic Resonance Imaging (POC MRI) Scanner System

Hyperfine, Inc. · MRI · FDA-cleared (510(k))

Authorized purpose: Swoop Point-of-Care Magnetic Resonance Imaging Scanner System provides point-of-care MR images of the head for the labeled adult and pediatric workflow, with interpretation and clinical decisions remaining with trained physicians.

Clinical role
MRI acquisition, reconstruction, display, review, and release-specific workflow support.
Images and patients
MRI · Head
Key limitation
Keep in mind: Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release; map the installed configuration to exact clearance and labeling.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: Swoop Point-of-Care Magnetic Resonance Imaging Scanner System provides point-of-care MR images of the head for the labeled adult and pediatric workflow, with interpretation and clinical decisions remaining with trained physicians.
  • Which images or clinical context are in scope?Modality: MRI Anatomy or clinical context: Head Input: Patient MRI signal data acquired with supported coils, pulse sequences, field strengths, contrast, and release-specific protocols.
  • Who uses it?Intended users: Qualified MRI technologists and physicians trained in magnetic resonance imaging.
  • What does the software produce?Output: Cross-sectional MR images, spectra, derived measurements, and reconstructed series for professional review.
  • What is its most important limitation?Key limitation: Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release; map the installed configuration to exact clearance and labeling.

Clinical takeaway. Clinical role: Swoop Point-of-Care Magnetic Resonance Imaging Scanner System provides point-of-care MR images of the head for the labeled adult and pediatric workflow, with interpretation and clinical decisions remaining with trained physicians.

Evidence takeaway. Evidence from the reviewed source record: K212456 FDA decision material describes the cleared magnetic resonance imaging system, labeled use, and submission-specific verification. It does not establish product-wide sensitivity, specificity, outcomes, or independent clinical validation for this exact configuration.

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
Swoop Point-of-Care Magnetic Resonance Imaging Scanner System provides point-of-care MR images of the head for the labeled adult and pediatric workflow, with interpretation and clinical decisions remaining with trained physicians.
  • Exact FDA submission
  • Checked 2026-09-04
Workflow role
MRI acquisition, reconstruction, display, review, and release-specific workflow support.
  • Exact FDA submission
  • Checked 2026-09-04
Required input
Patient MRI signal data acquired with supported coils, pulse sequences, field strengths, contrast, and release-specific protocols.
  • Exact FDA submission
  • Checked 2026-09-04
Output and clinician action
Cross-sectional MR images, spectra, derived measurements, and reconstructed series for professional review.
  • Exact FDA submission
  • Checked 2026-09-04
Care setting
Hospital and outpatient MRI departments operating the cleared field strength, coils, protocols, and accessories.
  • Exact FDA submission
  • Checked 2026-09-04
Intended users
Qualified MRI technologists and physicians trained in magnetic resonance imaging.
  • Exact FDA submission
  • Checked 2026-09-04
Limitations
Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release; map the installed configuration to exact clearance and labeling.
  • Exact FDA submission
  • Checked 2026-09-04
Evidence summary
K212456 FDA decision material describes the cleared magnetic resonance imaging system, labeled use, and submission-specific verification. It does not establish product-wide sensitivity, specificity, outcomes, or independent clinical validation for this exact configuration.
  • Exact FDA submission
  • Checked 2026-09-04

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
K212456 FDA decision material describes the cleared magnetic resonance imaging system, labeled use, and submission-specific verification. It does not establish product-wide sensitivity, specificity, outcomes, or independent clinical validation for this exact configuration.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Reported sensitivity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-04

Sensitivity is not promoted as a product-wide endpoint for this device or software function. Compare only an exact endpoint, threshold, population, version, and source if later verified.

Reported specificity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-04

Specificity is not promoted as a product-wide endpoint for this device or software function. Compare only an exact endpoint, threshold, population, version, and source if later verified.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

K212456 FDA evidence package for Swoop Point-of-Care Magnetic Resonance Imaging (POC MRI) Scanner System

  • DesignFDA 510(k) substantial-equivalence decision evidence and product verification described in the exact submission.
  • PopulationHead
  • Scopeexact submission
  • Tested versionSwoop Point-of-Care Magnetic Resonance Imaging (POC MRI) Scanner System configuration described in K212456
  • IndependenceSponsor or vendor study

K212456

Manufacturer evidence submitted for the exact FDA decision. It is not independent clinical validation, and local technical, clinical, privacy, security, and workflow acceptance remains necessary.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K212456 by exact native submission identifier. This does not exclude product-family, later, or adverse-event information.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-04

Exact-identifier review does not prove absence of all product-family safety signals.

Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not publicly available: Reviewed sources do not publish an exact-product procurement price with a stated offer scope.

Reimbursement

Reimbursement
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging, procedural, or surgical service line and verify payer, site-of-service, and coding assumptions independently.
  • Service-level, not product-specific
  • Checked 2026-09-04

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

This is not a product-specific reimbursement determination. Build a local total-cost and value model and verify payer, site-of-service, coding, and coverage assumptions independently.

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: Patient MRI signal data acquired with supported coils, pulse sequences, field strengths, contrast, and release-specific protocols.

    Scope: Exact FDA submission.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-09-04 · FDA 510(k) decision material K212456: Swoop Point-of-Care Magnetic Resonance Imaging (POC MRI) Scanner System

  2. Routing

    Unresolved route: Not established for the exact route.

    Scope: Exact FDA submission. The exact public FDA decision material does not establish a complete integration contract. Confirm DICOM, worklist, identity, interfaces, exports, routing, audit behavior, and downtime locally.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-09-04 ·

  3. AI processing

    Unresolved route: Not established for the exact route.

    Scope: Exact FDA submission. The exact public FDA decision material does not establish the complete deployment topology, hosting responsibility, availability, upgrade path, or service contract.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-09-04 ·

  4. Returned objects

    Source context, not route validation: Cross-sectional MR images, spectra, derived measurements, and reconstructed series for professional review.

    Scope: Exact FDA submission.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-04 · FDA 510(k) decision material K212456: Swoop Point-of-Care Magnetic Resonance Imaging (POC MRI) Scanner System

  5. Viewer/worklist/report

    Unresolved route: Cross-sectional MR images, spectra, derived measurements, and reconstructed series for professional review.

    Scope: Exact FDA submission.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-04 · FDA 510(k) decision material K212456: Swoop Point-of-Care Magnetic Resonance Imaging (POC MRI) Scanner System

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Security
Hyperfine's current certifications page says the Swoop System device services, security services and Hyperfine PACS services (hosted on AWS, Ashburn, Virginia) meet HITRUST CSF v9.4 r2 criteria, and that a SOC 2 Type II report on the Swoop system (security criteria only) is available under NDA. The page gives no certification dates; neither report was read.
  • Vendor-wide
  • Checked 2026-10-25

Vendor-wide statement, not an independent attestation. It does not establish certificate or report scope for this exact release, expiry, data location, retention, identity and access, or incident-response terms for a deployed configuration. Obtain these from the vendor.

Training
Only qualified mri technologists and physicians trained in magnetic resonance imaging should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
  • Exact FDA submission
  • Checked 2026-09-04
Monitoring
Monitor K212456 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release; map the installed configuration to exact clearance and labeling.
  • Exact FDA submission
  • Checked 2026-09-04

Monitoring guidance is buyer-facing implementation guidance derived from the exact cleared scope; it is not a claim that the FDA package provides a live monitoring service.

Release and scope verification

Map installed model, build, options, population, anatomy, inputs, outputs, and enabled functions to K212456 before acceptance.

exact submission · checked 2026-09-04

Later product-family features are not evidence that this cleared release includes them.

Deployment boundary

The exact public deployment topology remains incomplete; establish hosting responsibility, data flow, availability, upgrade, service, and downtime boundaries locally.

exact submission · checked 2026-09-04

FDA-described deployment boundaries do not establish current hosting contracts, security controls, uptime, or local qualification.

Interoperability acceptance

The exact public integration contract remains incomplete; establish and test supported input, output, identity, routing, archive, and downtime behavior locally.

exact submission · checked 2026-09-04

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release; map the installed configuration to exact clearance and labeling.

exact submission · checked 2026-09-04

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-04

No exact-product price was found in the reviewed public material; productivity or value claims require local evidence.

Lifecycle monitoring

Monitor K212456 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release; map the installed configuration to exact clearance and labeling.

exact submission · checked 2026-09-04

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. magnetic resonance imaging system limits remain submission-specific.

Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Human reviewedStatus
2026-09-20Last searched
27Fields reviewed
13Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, vendor product page, fda decision material, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not publicly available - checked 2026-09-04 - Exact FDA submission

    No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Deployment: Not publicly available - checked 2026-09-04 - Exact FDA submission

    The exact public FDA decision material does not establish the complete deployment topology, hosting responsibility, availability, upgrade path, or service contract.

  • Integration: Not publicly available - checked 2026-09-04 - Exact FDA submission

    The exact public FDA decision material does not establish a complete integration contract. Confirm DICOM, worklist, identity, interfaces, exports, routing, audit behavior, and downtime locally.

  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Installation timeline: Not yet reviewed

7 sources - 3 history items

Source ledger

  1. FDA 510(k) decision material K212456: Swoop Point-of-Care Magnetic Resonance Imaging (POC MRI) Scanner System

    U.S. Food and Drug Administration - Published 2021-11-17 - Accessed 2026-09-04

    Scope: identity, purpose, clinicalUse, modality, anatomy, intendedUsers, careSetting, workflowRole, input, output, limitations, evidenceSummary

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  4. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  5. Medicare Coverage Database

    Centers for Medicare & Medicaid Services - Accessed 2026-09-01

    Scope: coverage, codingContext

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. Hyperfine Certifications

    Hyperfine, Inc. - Accessed 2026-10-25

    Scope: security

Research history

  1. K212456 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    human reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K241121

MICSI-RMT

Microstructure Imaging, Inc.

Authorized purpose: MICSI-RMT is image-processing software for MRI image enhancement and noise reduction in head MRI images. The FDA material describes analysis of functional and diffusion MRI images and use by radiologists in imaging centers, clinics, or hospitals.

Modality
MRI
Anatomy
Head
Decision
2024-07-17
FDA-cleared K232760

Swoop® Portable MR Imaging System®

Hyperfine, Inc.

Authorized purpose: The Swoop Portable MR Imaging System is a portable, ultra-low-field, bedside magnetic-resonance device for producing images that display the internal structure of the head where a full diagnostic examination is not clinically practical; a trained physician interprets the images.

Modality
MRI
Anatomy
Head
Decision
2023-10-06
FDA-cleared K230264

Ezra Flash

Ezra AI, Inc.

Authorized purpose: Ezra Flash is MR image-processing software used to reduce image noise in non-contrast MRI examinations acquired on 3-Tesla Siemens and GE scanners for sagittal T1, axial T2, and axial FLAIR sequences of the head in patients older than 18 years.

Modality
MRI
Anatomy
Head
Decision
2023-04-13
FDA-cleared K230208

Swoop® Portable MR Imaging System™

Hyperfine, Inc.

Authorized purpose: The Swoop Portable MR Imaging System is a bedside magnetic-resonance device for producing images that display the internal structure of the head where a full diagnostic examination is not clinically practical; a trained physician interprets the images for information useful in determining a diagnosis.

Modality
MRI
Anatomy
Head
Decision
2023-02-22

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination