FDA-cleared profile

SwiftMR

AIRS Medical Inc. · MRI · FDA-cleared (510(k))

Authorized purpose: SwiftMR is image-processing software that enhances supported MRI DICOM images by reducing noise or increasing sharpness within the labeled workflow; exact scanner, sequence, algorithm, and output limits are submission-specific.

Clinical role
MRI image enhancement or post-processing support before routine professional interpretation.
Images and patients
MRI · Multi-region
Key limitation
Keep in mind: Supported image size, anatomy, sequence, input quality, and processing options are release-specific; processed images require review against the source study. SwiftMR is image-processing software that enhances supported MRI DICOM images by reducing noise or increasing sharpness within the labeled workflow; exact scanner, sequence, algorithm, and output limits are submission-specific.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: SwiftMR is image-processing software that enhances supported MRI DICOM images by reducing noise or increasing sharpness within the labeled workflow; exact scanner, sequence, algorithm, and output limits are submission-specific.
  • Which images or clinical context are in scope?Modality: MRI Anatomy or clinical context: Multi-region Input: Supported DICOM MRI images and release-specific processing parameters.
  • Who uses it?Intended users: Qualified MRI technologists and interpreting physicians.
  • What does the software produce?Output: Processed or enhanced MR image objects for review and downstream archival.
  • What is its most important limitation?Key limitation: Supported image size, anatomy, sequence, input quality, and processing options are release-specific; processed images require review against the source study. SwiftMR is image-processing software that enhances supported MRI DICOM images by reducing noise or increasing sharpness within the labeled workflow; exact scanner, sequence, algorithm, and output limits are submission-specific.

Clinical takeaway. Clinical role: SwiftMR is image-processing software that enhances supported MRI DICOM images by reducing noise or increasing sharpness within the labeled workflow; exact scanner, sequence, algorithm, and output limits are submission-specific.

Evidence takeaway. Evidence from the reviewed source record: The exact FDA document reports software verification (unit and integration/system tests, passed) and a validation performance test on retrospective clinical images. Noise reduction met an acceptance criterion of average signal-to-noise ratio (SNR) of SwiftMR-processed series increased by 40% or more versus original series; sharpness increase met a criterion of ROI FWHM decreased by 0.13% or more for at least 90% of test datasets. Both tests passed. Results are as reported in the FDA submission, not independent validation.

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
SwiftMR is image-processing software that enhances supported MRI DICOM images by reducing noise or increasing sharpness within the labeled workflow; exact scanner, sequence, algorithm, and output limits are submission-specific.
  • Exact FDA submission
  • Checked 2026-09-04
Workflow role
MRI image enhancement or post-processing support before routine professional interpretation.
  • Exact FDA submission
  • Checked 2026-09-04
Required input
Supported DICOM MRI images and release-specific processing parameters.
  • Exact FDA submission
  • Checked 2026-09-04
Output and clinician action
Processed or enhanced MR image objects for review and downstream archival.
  • Exact FDA submission
  • Checked 2026-09-04
Care setting
MRI departments or image-processing workstations using supported DICOM MRI workflows.
  • Exact FDA submission
  • Checked 2026-09-04
Intended users
Qualified MRI technologists and interpreting physicians.
  • Exact FDA submission
  • Checked 2026-09-04
Limitations
Supported image size, anatomy, sequence, input quality, and processing options are release-specific; processed images require review against the source study. SwiftMR is image-processing software that enhances supported MRI DICOM images by reducing noise or increasing sharpness within the labeled workflow; exact scanner, sequence, algorithm, and output limits are submission-specific.
  • Exact FDA submission
  • Checked 2026-09-04
Evidence summary
The exact FDA document reports software verification (unit and integration/system tests, passed) and a validation performance test on retrospective clinical images. Noise reduction met an acceptance criterion of average signal-to-noise ratio (SNR) of SwiftMR-processed series increased by 40% or more versus original series; sharpness increase met a criterion of ROI FWHM decreased by 0.13% or more for at least 90% of test datasets. Both tests passed. Results are as reported in the FDA submission, not independent validation.
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "1) Verification testing: Unit test, Integration/system test conducted. These tests passed." … "Validation testing: Performance test was conducted using retrospective clinical images for both noise reduction and sharpness increase functions.". Read from the exact FDA decision document on 2026-11-14.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
The exact FDA document reports software verification (unit and integration/system tests, passed) and a validation performance test on retrospective clinical images. Noise reduction met an acceptance criterion of average signal-to-noise ratio (SNR) of SwiftMR-processed series increased by 40% or more versus original series; sharpness increase met a criterion of ROI FWHM decreased by 0.13% or more for at least 90% of test datasets. Both tests passed. Results are as reported in the FDA submission, not independent validation.
  • FDA source
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "1) Verification testing: Unit test, Integration/system test conducted. These tests passed." … "Validation testing: Performance test was conducted using retrospective clinical images for both noise reduction and sharpness increase functions.". Read from the exact FDA decision document on 2026-11-14.

Reported sensitivity
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-04

No exact submission and version-linked sensitivity endpoint is published in this dossier; do not substitute a product-family or candidate-study value.

Reported specificity
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-04

No exact submission and version-linked specificity endpoint is published in this dossier; do not substitute a product-family or candidate-study value.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

FDA submission-specific verification and validation evidence for K210999

  • DesignRegulatory verification, validation, and labeling evidence reviewed for exact FDA submission K210999.
  • PopulationThe labeled population and test materials for exact submission K210999; a general clinical study population is not established here.
  • Scopeexact submission
  • Tested versionSwiftMR (K210999)
  • IndependenceSponsor or vendor study

K210999

FDA regulatory evidence is submission-specific but is not independent clinical performance evidence.

3 unlinked candidate evidence records are retained for source review and excluded from this evidence display.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-04

No native submission-identifier match was found in the checked openFDA radiology recall snapshot. Product-name, manufacturer, and adverse-event searches remain necessary; this does not establish that no recall or safety report exists.

Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not publicly available: Reviewed sources do not publish an exact-product procurement price with a stated offer scope.

Reimbursement

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: Supported DICOM MRI images and release-specific processing parameters.

    Scope: Exact FDA submission.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-09-04 · FDA decision summary for K210999 - SwiftMR

  2. Routing

    Source context, not route validation: The exact FDA package describes DICOM-format MRI input and enhanced DICOM output, with images saved in PACS automatically uploaded to the server for processing and sent back to PACS, and results retrievable as a worklist after user login. Interface versions and local compatibility require confirmation.

    Scope: Exact FDA submission. Packet wording: "The device’s inputs are standard of care MRI images in DICOM format. The deep learning algorithm produces enhanced images as outputs with reduced noise and increased sharpness in DICOM format." … "the newly acquired images are automatically uploaded to the server and registered in the database (DB) for image processing. Once image processing is complete, the images are sent back to PACS.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-11-14 · FDA decision summary for K210999 - SwiftMR

  3. AI processing

    Source context, not route validation: The exact FDA package describes SwiftMR as a software application loaded into PACS that runs automatically in the background, with newly acquired images uploaded to a server for processing and product settings changed through a client application. Current hosting, data flow, and commercial configuration require vendor confirmation.

    Scope: Exact FDA submission. Packet wording: "The subject device and predicate are both software applications that are loaded into PACS." … "When the user executes MR scans and saves the images in PACS as he/she usually does, the newly acquired images are automatically uploaded to the server and registered in the database (DB) for image processing.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-11-14 · FDA decision summary for K210999 - SwiftMR

  4. Returned objects

    Source context, not route validation: Processed or enhanced MR image objects for review and downstream archival.

    Scope: Exact FDA submission.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-04 · FDA decision summary for K210999 - SwiftMR

  5. Viewer/worklist/report

    Unresolved route: Processed or enhanced MR image objects for review and downstream archival.

    Scope: Exact FDA submission.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-04 · FDA decision summary for K210999 - SwiftMR

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Deployment
The exact FDA package describes SwiftMR as a software application loaded into PACS that runs automatically in the background, with newly acquired images uploaded to a server for processing and product settings changed through a client application. Current hosting, data flow, and commercial configuration require vendor confirmation.
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "The subject device and predicate are both software applications that are loaded into PACS." … "When the user executes MR scans and saves the images in PACS as he/she usually does, the newly acquired images are automatically uploaded to the server and registered in the database (DB) for image processing.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.

Integration
The exact FDA package describes DICOM-format MRI input and enhanced DICOM output, with images saved in PACS automatically uploaded to the server for processing and sent back to PACS, and results retrievable as a worklist after user login. Interface versions and local compatibility require confirmation.
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "The device’s inputs are standard of care MRI images in DICOM format. The deep learning algorithm produces enhanced images as outputs with reduced noise and increased sharpness in DICOM format." … "the newly acquired images are automatically uploaded to the server and registered in the database (DB) for image processing. Once image processing is complete, the images are sent back to PACS.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.

Security
AIRS Medical's SwiftMR information security white paper summary states security practices for cloud and on-premises workflows (transport encryption, network access control, application security, identity management, backup, disaster recovery, monitoring) and lists ISO/IEC 27001:2022, GDPR, HIPAA and SOC 2 Type II. Vendor statement about SwiftMR; the full white paper was not read.
  • Vendor-wide
  • Checked 2026-10-25

Vendor statement, not an attestation; exact-release scope is not established.

Workflow role

MRI image enhancement or post-processing support before routine professional interpretation.

exact submission · checked 2026-09-04

This is a submission-anchored intended-use summary, not a local acceptance result.

Input

Supported DICOM MRI images and release-specific processing parameters.

exact submission · checked 2026-09-04

Exact supported inputs and protocol limits must be confirmed from the complete labeling.

Output

Processed or enhanced MR image objects for review and downstream archival.

exact submission · checked 2026-09-04

Output review, editing, and downstream use remain release- and site-specific.

Vendor family context

Current vendor material is product-family context for K210999 only; exact release, enabled options, deployment, and local interoperability require confirmation.

product family · checked 2026-09-04

Vendor material is not exact-release evidence unless that linkage is separately proven.

Site validation status

Exact local interfaces, deployment topology, security controls, downtime behavior, and support obligations remain unestablished.

exact submission · checked 2026-09-04

Obtain organization-specific implementation evidence before procurement or clinical use.

Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K210999
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
MRI
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Anatomy
Multi-region
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2021-10-14
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
28Fields reviewed
15Source classes checked
3Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains 3 PubMed and 0 ClinicalTrials.gov candidate records. Discovery metadata alone does not establish exact tested-product, version, population, and endpoint applicability, so all candidates remain quarantined and none was promoted. The current discovery snapshot contains 3 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, vendor documentation, vendor product page, implementation documentation, reimbursement policy, peer reviewed literature, clinical trial registry, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not publicly available - checked 2026-09-04 - Exact FDA submission

    No public exact-release price, license fee, subscription, contract, or implementation cost was established in this review.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Reimbursement: Not applicable - checked 2026-09-04 - Not applicable

    CMS index context was checked, but no product-specific coverage or reimbursement determination is inferred.

  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Installation timeline: Not yet reviewed
  • Training: Not verified - checked 2026-10-27 - Exact FDA submission

    Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

  • Monitoring: Not verified - checked 2026-10-27 - Exact FDA submission

    Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

12 sources - 3 history items

Source ledger

  1. FDA decision summary for K210999 - SwiftMR

    U.S. Food and Drug Administration - Published 2021-10-14 - Accessed 2026-09-04

    Scope: regulatoryIdentity, productIdentity, indication, modality, anatomy, workflowRole, input, output, limitations, evidenceSummary

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. SwiftMR model card (K210999)

    American College of Radiology - Published 2026-06-26 - Accessed 2026-09-01

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. Medicare Coverage Database

    Centers for Medicare & Medicaid Services - Accessed 2026-09-01

    Scope: coverage, codingContext

  7. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  8. AIRS SwiftMR family context

    AIRS Medical - Accessed 2026-09-04

    Scope: productFamilyContext, workflowRole, implementationCapabilities

  9. Leveraging a vendor-neutral deep learning reconstruction algorithm to reduce scan time and enhance image quality in T2-weighted liver MRI.

    U.S. National Library of Medicine - Accessed 2026-09-04

    Scope: evidenceStudies, productFamilyContext, evidenceDiscovery

  10. Deep Learning-Based High-Resolution Magnetic Resonance Angiography (MRA) Generation Model for 4D Time-Resolved Angiography with Interleaved Stochastic Trajectories (TWIST) MRA in Fast Stroke Imaging.

    U.S. National Library of Medicine - Accessed 2026-09-04

    Scope: evidenceStudies, productFamilyContext, evidenceDiscovery

  11. Signal-to-noise ratio improvement in postmortem MRI using deep learning reconstruction (SwiftMR).

    U.S. National Library of Medicine - Accessed 2026-09-04

    Scope: evidenceStudies, productFamilyContext, evidenceDiscovery

  12. AIRS Medical SwiftMR Information Security white paper summary

    AIRS Medical - Accessed 2026-10-25

    Scope: security

Research history

  1. K210999 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K260746

S-scan Open (100001800)

Esaote, S.p.A.

Authorized purpose: S-scan Open is an Esaote MR system for diagnostic image acquisition, reconstruction, display, and professional interpretation within the labeled scanner, coil, sequence, and anatomy scope.

Modality
MRI
Anatomy
Multi-region
Decision
2026-03-27
FDA-cleared K252379

AIR Recon DL

Ge Medical Systems, LLC

Authorized purpose: AIR Recon DL adds a deep-learning phase-correction and reconstruction option to supported GE 1.5T, 3T, and 7T MR workflows for all labeled anatomies and ages to reduce noise or ringing and support image quality or scan-time choices.

Modality
MRI
Anatomy
Multi-region
Decision
2025-12-23
FDA-cleared K252838

MAGNETOM Sola; MAGNETOM Altea; MAGNETOM Sola Fit; MAGNETOM Viato.Mobile; MAGNETOM Vida; MAGNETOM Lumina; MAGNETOM Vida Fit; MAGNETOM Flow.Elite; MAGNETOM Flow.Neo; MAGNETOM Flow.Rise

Siemens Healthcare GmbH

Authorized purpose: The listed MAGNETOM 1.5T and 3T MRI systems produce cross-sectional images, spectroscopic images or spectra, and derived physical parameters of the head, body, and extremities for trained-physician interpretation and can support imaging during MR-compatible interventions.

Modality
MRI
Anatomy
Multi-region
Decision
2025-12-19

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination

Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.

Accepted repository change history

No entry has been added to this scoped change history for this record. This is not proof that the record never changed.

Correction policy and limits · Public repository change history