FDA-cleared profile

Second Opinion Periapical Radiolucency Contours

Pearl Inc.

Second Opinion PC assists dentists by detecting and displaying periapical radiolucencies with bounding polygons on dental radiographs of permanent teeth in patients age 12 and older as a concurrent or second reader, not a replacement for complete dental evaluation.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
Second Opinion PC assists dentists by detecting and displaying periapical radiolucencies with bounding polygons on dental radiographs of permanent teeth in patients age 12 and older as a concurrent or second reader, not a replacement for complete dental evaluation.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Intended users
Licensed dentists, dental radiologists, and other trained dental healthcare professionals.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Care setting and population
Dental practices, dental-imaging departments, and clinical workstations within the exact labeled workflow.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Workflow role
Dental image viewing, measurement, second-reader, detection, segmentation, or treatment-planning support.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Required input
Supported dental radiographs, CBCT, DICOM studies, or imported dental-image objects described in the exact submission.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Output and human action
Dental image marks, contours, measurements, three-dimensional views, or planning information for professional review.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Limitations
The exact dental view, patient age, dentition, image quality, and concurrent clinical evaluation limits applicability; outputs do not replace professional judgment.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K242600
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
X-ray/Radiography
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Anatomy
Dental permanent teeth and periapical region
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2025-04-11
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
Standards-based DICOM/PACS integration was identified in the FDA submission materials; compatibility with a specific PACS, RIS, EHR, or orchestrator requires confirmation.
  • Public source
  • Exact FDA submission
  • Checked 2026-08-13

Normalized from the prior exact-submission review. Confirm interface versions and local compatibility during technical evaluation.

Deployment and data flow
On-cloud
  • Public source
  • Exact FDA submission
  • Checked 2026-08-13

Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.

Security and privacy
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-04

Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.

Training and support
Only licensed dentists, dental radiologists, and other trained dental healthcare professionals should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Monitoring and change control
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-04

The exact FDA decision and checked recall snapshot do not establish a complete local postmarket monitoring protocol. Track version, population, anatomy, protocol, input quality, edits, failures, repeats, downstream effects, downtime, incidents, recalls, and updates locally.

Release and scope verification

Map the installed model, build, options, population, anatomy, inputs, outputs, and enabled functions to K242600 before acceptance.

exact submission · checked 2026-09-04

Later product-family features are not evidence that this cleared release includes them.

Interoperability acceptance

Exact public integration details remain incomplete; establish and test the supported input, output, identity, routing, archive, report, and downtime contract locally.

exact submission · checked 2026-09-04

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.

Clinical acceptance

Use a signed local test set for the exact labeled population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. The exact dental view, patient age, dentition, image quality, and concurrent clinical evaluation limits applicability; outputs do not replace professional judgment.

exact submission · checked 2026-09-04

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-04

No exact-product price was found in reviewed public material; productivity or value claims require local evidence.

Lifecycle monitoring

Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates.

exact submission · checked 2026-09-04

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. No exact native-submission recall record was found in the checked snapshot; this does not exclude other product-family or later signals.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
The exact FDA 510(k) decision material establishes the submission identity, labeled use, inputs, population, and configuration boundary summarized here. No product-wide clinical-performance or outcome claim is promoted without an exact endpoint and version linkage.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Reported sensitivity
95%; 88%
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-13

Values reported in the exact-submission materials. Consult the linked source for endpoint, threshold, population, and confidence interval before comparison.

Reported specificity
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-04

No exact submission-linked specificity endpoint with a verified population, threshold, tested version, and source was promoted.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

K242600 FDA evidence package for Second Opinion Periapical Radiolucency Contours

  • DesignFDA 510(k) decision evidence package reviewed for exact submission identity and labeled scope
  • PopulationLabeled population and configuration described in the K242600 FDA decision material
  • Scopeexact submission
  • Tested versionSecond Opinion PC configuration described in K242600
  • IndependenceSponsor or vendor study

K242600

Manufacturer evidence submitted for the exact FDA decision. It is not independent clinical validation, and local technical, clinical, privacy, security, and workflow acceptance remains necessary.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-04

No public list price for the exact cleared configuration was found. Obtain a written quote separating license or hardware basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

Reimbursement and coding
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging, procedural, or surgical service line and verify payer, site-of-service, and coding assumptions independently.
  • Public source
  • Not applicable
  • Checked 2026-09-04

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

Build a local total-cost and value model; do not infer reimbursement from FDA clearance or approval.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K242600 by exact native submission identifier. This does not exclude product-family, later, or adverse-event information.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-04

Exact-identifier review does not prove absence of all product-family safety signals.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-04Last searched
26Fields reviewed
10Source classes checked
0Awaiting review PubMed leads
0Awaiting review trial leads
0Unreviewed FDA recall leads

Recovered from the prior exact-submission extraction and normalized under the current provenance rules. Exact-name and exact-submission searches in the checked 2026-09-01 indexes returned no PubMed records and no ClinicalTrials.gov records for K242600; review status is candidate_review_pending. Any candidate literature or registry record remains quarantined as product-family evidence unless exact submission and version linkage is established. The checked openFDA snapshot has 0 exact native-submission recall records. Pricing, complete security architecture, contracted deployment, exact reimbursement, and local interoperability remain unverified. Automated exact-name discovery found 0 PubMed and 0 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, fda decision material, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA 510(k) decision material K242600: Second Opinion Periapical Radiolucency Contours

    U.S. Food and Drug Administration - Published 2025-04-11 - Accessed 2026-09-04

    Scope: identity, purpose, clinicalUse, modality, anatomy, intendedUsers, careSetting, workflowRole, input, output, limitations, evidenceSummary

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  4. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  5. Medicare Coverage Database

    Centers for Medicare & Medicaid Services - Accessed 2026-09-01

    Scope: coverage, codingContext

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

Research history

  1. K242600 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

Alternatives

Related tools to compare

FDA-cleared K252953

Velmeni for Dentists (V4D) Endo-Perio

Velmeni, Inc.

Concurrent-read CADe to assist dentists in detecting specified tooth and jaw anatomy, bone-level landmarks, and root-canal-treated teeth in digital radiographs of permanent teeth...

Modality
X-ray/Radiography
Anatomy
Permanent teeth and jaw/maxillofacial anatomy, including enamel, dentine, pulp, root, mental foramen, inferior alveolar canal, maxillary sinus, cementoenamel junction, crestal bone, and root-canal-treated teeth
Decision
2026-05-11
FDA-cleared K253502

Critical Care Suite with Enteric Tube Positioning AI Algorithm

Ge Medical Systems, LLC

Analyze frontal chest and abdominal X-ray images to detect and localize a nasally or orally inserted enteric tube, its tip and side port if present, and the diaphragm and airways...

Modality
X-ray/Radiography
Anatomy
Chest/lung and abdomen/stomach; enteric tube, tube tip and side port, diaphragm, and airways including carina and left/right bronchi
Decision
2026-04-14
FDA-cleared K251934

qXR-Detect

Qure.Ai Technologies

qXR-Detect is an adult chest-radiograph concurrent reading aid that identifies, highlights and categorizes suspicious regions of interest across lung, pleura,...

Modality
X-ray/Radiography
Anatomy
Chest/Lung; pleura, mediastinum/hila/heart, bone, hardware and other chest-radiograph regions
Decision
2026-01-16
FDA-cleared K250264

SugarBug (1.x)

Bench7, Inc.

Provides automated concurrent-read detection and segmentation of caries on bitewing radiographs in patients 18 and older, adding information for the dentist while requiring...

Modality
X-ray/Radiography
Anatomy
Dental/Maxillofacial
Decision
2025-11-07