QUANTRA

HOLOGIC, INC. · Mammography/DBT · FDA-cleared (510(k))

Authorized purpose: Quantra version 2.1 calculates volumetric and area breast-density measures and density categories from supported digital mammography images as an adjunct, not a diagnostic result.

Clinical role
Breast image acquisition, reconstruction, density assessment, or computer-aided interpretation support.
Images and patients
Mammography/DBT · Breast
Key limitation
Keep in mind: Quantra produces adjunctive information; it is not an interpretive or diagnostic aid.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: Quantra version 2.1 calculates volumetric and area breast-density measures and density categories from supported digital mammography images as an adjunct, not a diagnostic result.
  • Which images or clinical context are in scope?Modality: Mammography/DBT Anatomy or clinical context: Breast Input: Images acquired using digital breast x-ray systems.
  • Who uses it?Intended users: Radiologists.
  • What does the software produce?Output: Quantra calculates volumetric breast density as a ratio of fibroglandular tissue and total breast volume estimates. Quantra also provides area breast density as a ratio of fibroglandular tissue area and total breast area estimates. Quantra segregates breast density into categories.
  • What is its most important limitation?Key limitation: Quantra produces adjunctive information; it is not an interpretive or diagnostic aid.

Clinical takeaway. Clinical role: Quantra version 2.1 calculates volumetric and area breast-density measures and density categories from supported digital mammography images as an adjunct, not a diagnostic result.

Evidence takeaway. Evidence from the reviewed source record: K142037 reports bench measurement, reproducibility, and multivendor agreement validation for Quantra 2.1.

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
Quantra version 2.1 calculates volumetric and area breast-density measures and density categories from supported digital mammography images as an adjunct, not a diagnostic result.
  • Exact FDA submission
  • Checked 2026-09-19

Clinical-use display mirrors the exact FDA-supported purpose reconciled from the decision-summary packet.

Workflow role
Breast image acquisition, reconstruction, density assessment, or computer-aided interpretation support.
  • Exact FDA submission
  • Checked 2026-09-03
Required input
Images acquired using digital breast x-ray systems.
  • Exact FDA submission
  • Checked 2026-09-19

This replaces the current DBT-inclusive input wording with the exact indication text.

Output and clinician action
Quantra calculates volumetric breast density as a ratio of fibroglandular tissue and total breast volume estimates. Quantra also provides area breast density as a ratio of fibroglandular tissue area and total breast area estimates. Quantra segregates breast density into categories.
  • Exact FDA submission
  • Checked 2026-09-19

The exact indication directly states the density measures and categories; it does not establish the current reconstructed-image, marks, or scores list.

Care setting
Breast-imaging departments performing supported digital mammography or tomosynthesis workflows.
  • Exact FDA submission
  • Checked 2026-09-03
Intended users
Radiologists.
  • Exact FDA submission
  • Checked 2026-09-19

The exact indication names radiologists and does not establish the current technologist role.

Limitations
Quantra produces adjunctive information; it is not an interpretive or diagnostic aid.
  • Exact FDA submission
  • Checked 2026-09-19

The exact indication states this limitation and does not state the current compatibility-validation or local-reporting-policy language.

Evidence summary
K142037 reports bench measurement, reproducibility, and multivendor agreement validation for Quantra 2.1.
  • Exact FDA submission
  • Checked 2026-09-03

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
K142037 reports bench measurement, reproducibility, and multivendor agreement validation for Quantra 2.1.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Reported sensitivity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-03

Sensitivity is not the primary endpoint for this acquisition, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

Reported specificity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-03

Specificity is not the primary endpoint for this acquisition, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

K142037 FDA performance package for QUANTRA

  • DesignManufacturer measurement, reproducibility, and multivendor agreement validation
  • PopulationSupported digital mammography examinations
  • Scopeexact submission
  • Tested versionQuantra version 2.1
  • IndependenceSponsor or vendor study

K142037

Manufacturer evidence submitted for clearance or approval; local clinical, technical, and workflow acceptance remains necessary.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K142037 by exact submission identifier. This does not exclude product-family or later safety information.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-03

Exact-identifier review does not prove absence of all product-family safety signals.

Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not publicly available: Reviewed sources do not publish an exact-product procurement price with a stated offer scope.

Reimbursement

Reimbursement
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging, procedural, or surgical service line and verify payer, site-of-service, and coding assumptions independently.
  • Service-level, not product-specific
  • Checked 2026-09-03

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

Build a local total-cost and value model; do not infer reimbursement from FDA clearance or approval.

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: Images acquired using digital breast x-ray systems.

    Scope: Exact FDA submission. This replaces the current DBT-inclusive input wording with the exact indication text.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-09-19 · FDA K142037: QUANTRA

  2. Routing

    Source context, not route validation: Validate supported acquisition systems, image type and processing, DICOM transfer, hanging protocols, display requirements, result presentation, and archive retention.

    Scope: Exact FDA submission.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-09-03 · FDA K142037: QUANTRA · Hologic Quantra breast density assessment

  3. AI processing

    Source context, not route validation: Acquisition-system software or networked analysis software on the cleared platform and version.

    Scope: Exact FDA submission.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-09-03 · FDA K142037: QUANTRA · Hologic Quantra breast density assessment

  4. Returned objects

    Source context, not route validation: Quantra calculates volumetric breast density as a ratio of fibroglandular tissue and total breast volume estimates. Quantra also provides area breast density as a ratio of fibroglandular tissue area and total breast area estimates. Quantra segregates breast density into categories.

    Scope: Exact FDA submission. The exact indication directly states the density measures and categories; it does not establish the current reconstructed-image, marks, or scores list.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-19 · FDA K142037: QUANTRA

  5. Viewer/worklist/report

    Unresolved route: Quantra calculates volumetric breast density as a ratio of fibroglandular tissue and total breast volume estimates. Quantra also provides area breast density as a ratio of fibroglandular tissue area and total breast area estimates. Quantra segregates breast density into categories.

    Scope: Exact FDA submission. The exact indication directly states the density measures and categories; it does not establish the current reconstructed-image, marks, or scores list.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-19 · FDA K142037: QUANTRA

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Deployment
Acquisition-system software or networked analysis software on the cleared platform and version.
  • Exact FDA submission
  • Checked 2026-09-03
Integration
Validate supported acquisition systems, image type and processing, DICOM transfer, hanging protocols, display requirements, result presentation, and archive retention.
  • Exact FDA submission
  • Checked 2026-09-03
Security
Hologic's Breast and Skeletal Products Cybersecurity page states patch validation for Hologic Breast and Skeletal Health systems, MDS2 forms availability, and published security advisories with product impact analyses (including Libwebp, MOVEit, Log4j and Axeda advisories).
  • Product family
  • Checked 2026-10-26

Vendor statement, not an attestation. Page scope is Hologic's Breast and Skeletal Health division; this entry is a mammography/DBT product within that division. Exact-release scope is not established.

Training
Only qualified breast-imaging technologists and interpreting physicians. should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
  • Exact FDA submission
  • Checked 2026-09-03
Monitoring
Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track these product-specific limits: Density output is adjunctive and is not an interpretive or diagnostic result; validate acquisition-vendor compatibility and local reporting policy.
  • Exact FDA submission
  • Checked 2026-09-03

Release and scope verification

Map the installed model, build, options, population, anatomy, inputs, and outputs to K142037 before acceptance.

exact submission · checked 2026-09-03

Later product-family features are not evidence that this cleared release includes them.

Interoperability acceptance

Validate supported acquisition systems, image type and processing, DICOM transfer, hanging protocols, display requirements, result presentation, and archive retention.

exact submission · checked 2026-09-03

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Density output is adjunctive and is not an interpretive or diagnostic result; validate acquisition-vendor compatibility and local reporting policy.

exact submission · checked 2026-09-03

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-03

Vendor productivity claims are scenario inputs, not guaranteed local savings.

Lifecycle monitoring

Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track these product-specific limits: Density output is adjunctive and is not an interpretive or diagnostic result; validate acquisition-vendor compatibility and local reporting policy.

exact submission · checked 2026-09-03

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes.

Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K142037
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
Mammography/DBT
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

The exact indication states digital breast x-ray systems and does not state DBT. Controlled directory display reconciled from exact FDA wording: Images acquired using digital breast x-ray systems. -> Mammography/DBT.

Anatomy
Breast
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2014-10-10
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
27Fields reviewed
12Source classes checked
20Screened PubMed leads
10Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains 20 PubMed and 10 ClinicalTrials.gov candidate records. Discovery metadata alone does not establish exact tested-product, version, population, and endpoint applicability, so all candidates remain quarantined and none was promoted. The current discovery snapshot contains 20 PubMed and 10 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, acr ai central product, fda decision material, vendor product page, reimbursement policy, fda device recall, fda decision summary packet, literature index, trial registry, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not publicly available - checked 2026-09-03 - Exact FDA submission

    No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, enabled features, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Installation timeline: Not yet reviewed

9 sources - 3 history items

Source ledger

  1. FDA K142037: QUANTRA

    U.S. Food and Drug Administration - Published 2014-10-10 - Accessed 2026-09-19

    Scope: purpose, anatomy, evidenceSummary, modality, intendedUsers, input, output, limitations, clinicalUse

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. QUANTRA model card (K142037)

    American College of Radiology - Published 2026-06-26 - Accessed 2026-09-01

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. Medicare Coverage Database

    Centers for Medicare & Medicaid Services - Accessed 2026-09-01

    Scope: coverage, codingContext

  7. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  8. Hologic Quantra breast density assessment

    Hologic - Accessed 2026-09-03

    Scope: integration, deployment, monitoring, productFamilyContext

  9. Hologic: Breast & Skeletal Products Cybersecurity

    Hologic - Accessed 2026-10-26

    Scope: security

Research history

  1. K142037 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K082483

QUANTRA

HOLOGIC, INC.

Authorized purpose: Quantra version 1.1 calculates a volumetric breast-density ratio from supported Hologic full-field digital mammography images on the Cenova server as an adjunct, not a diagnostic result.

Modality
Mammography/DBT
Anatomy
Breast
Decision
2008-09-12
FDA-cleared K260714

MammoScreen® (5)

Therapixel

Authorized purpose: Flags calcifications, masses, asymmetries and architectural distortions on compatible FFDM or DBT screening examinations; management must not be based solely on MammoScreen output.

Modality
Mammography/DBT
Anatomy
Breast
Decision
2026-06-26
FDA-cleared K253825

Saige-Dx

DeepHealth, Inc.

Authorized purpose: Concurrent detection and characterization support for screening DBT mammography

Modality
Mammography/DBT
Anatomy
Breast
Decision
2026-06-15
FDA-cleared K252954

MammoSightAI

Neurocareai, Inc. (Dba Savelife.Ai)

Authorized purpose: Tags screening cases as suspicious or blank to aid worklist prioritization; radiologists must review every examination and it does not mark original images.

Modality
Mammography/DBT
Anatomy
Breast
Decision
2026-05-04

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination

Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.

Accepted repository change history

No entry has been added to this scoped change history for this record. This is not proof that the record never changed.

Correction policy and limits · Public repository change history