QP-Brain®

Quibim S.L. · MRI · FDA-cleared (510(k))

Authorized purpose: Automatically label and volumetrically quantify segmentable brain structures and white matter hyperintensities (WMH) from MR images in adults and adolescents aged 18 years and older, as a support tool for properly trained clinicians assessing structural MRIs; patient-management decisions must not rely solely on the device results.

Clinical role
Automated structural-brain MRI segmentation, volumetric quantification, and WMH analysis with clinician review as a support tool; sequence compatibility is checked before analysis and results can be reviewed longitudinally.
Images and patients
MRI · Brain
Key limitation
Keep in mind: The software is a support tool and patient-management decisions must not be based solely on its results. The FDA scope is limited to segmentable brain structures and WMH from the specified T1 and T2 FLAIR MRI inputs; WMH performance varies materially by lesion burden, and the packet does not establish a general diagnostic claim or a specific current PACS, RIS, EHR, or deployment configuration.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: Automatically label and volumetrically quantify segmentable brain structures and white matter hyperintensities (WMH) from MR images in adults and adolescents aged 18 years and older, as a support tool for properly trained clinicians assessing structural MRIs; patient-management decisions must not rely solely on the device results.
  • Which images or clinical context are in scope?Modality: MRI Anatomy or clinical context: Brain Input: DICOM-compatible MR images, including 3D T1-weighted Gradient Echo structural MRI for brain structures and T2 FLAIR MRI for WMH identification, from adults and adolescents aged 18 years and older; the application checks available sequences for compatibility before launching analysis.
  • Who uses it?Intended users: Clinicians with proper training
  • What does the software produce?Output: Volumes for gray matter, white matter, cerebrospinal fluid, and intracranial volume, plus WMH quantification, segmentation color overlays, morphometric reports, comparisons with age- and gender-matched normative percentile data when available, longitudinal reporting, and DICOM-formatted results displayable on DICOM workstations and PACS.
  • What is its most important limitation?Key limitation: The software is a support tool and patient-management decisions must not be based solely on its results. The FDA scope is limited to segmentable brain structures and WMH from the specified T1 and T2 FLAIR MRI inputs; WMH performance varies materially by lesion burden, and the packet does not establish a general diagnostic claim or a specific current PACS, RIS, EHR, or deployment configuration.

Clinical takeaway. Clinical role: Automatically label and volumetrically quantify segmentable brain structures and white matter hyperintensities (WMH) from MR images in adults and adolescents aged 18 years and older, as a support tool for properly trained clinicians assessing structural MRIs; patient-management decisions must not rely solely on the device results.

Evidence takeaway. Evidence from the reviewed source record: FDA K232231 reports software verification and validation against manual expert segmentation. For T1 brain structures, Dice scores and relative-volume differences were GM 0.983 and 2.846, WM 0.990 and 1.643, CSF 0.955 and 7.495, and ICV 0.994 and 0.496. For T2 FLAIR WMH, burden-stratified Dice scores were 0.506, 0.636, 0.774, and 0.885 for low, medium, high, and very high WMH, with corresponding absolute volume errors of 0.675, 2.544, 3.097, and 7.833 mL; the packet also reports Pearson correlation of 0.88 for WMH count. No sample size or clinical-outcome study is stated in the reviewed FDA packet.

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
Automatically label and volumetrically quantify segmentable brain structures and white matter hyperintensities (WMH) from MR images in adults and adolescents aged 18 years and older, as a support tool for properly trained clinicians assessing structural MRIs; patient-management decisions must not rely solely on the device results.
  • Exact FDA submission
  • Checked 2026-09-08
Workflow role
Automated structural-brain MRI segmentation, volumetric quantification, and WMH analysis with clinician review as a support tool; sequence compatibility is checked before analysis and results can be reviewed longitudinally.
  • Exact FDA submission
  • Checked 2026-09-08
Required input
DICOM-compatible MR images, including 3D T1-weighted Gradient Echo structural MRI for brain structures and T2 FLAIR MRI for WMH identification, from adults and adolescents aged 18 years and older; the application checks available sequences for compatibility before launching analysis.
  • Exact FDA submission
  • Checked 2026-09-08
Output and clinician action
Volumes for gray matter, white matter, cerebrospinal fluid, and intracranial volume, plus WMH quantification, segmentation color overlays, morphometric reports, comparisons with age- and gender-matched normative percentile data when available, longitudinal reporting, and DICOM-formatted results displayable on DICOM workstations and PACS.
  • Exact FDA submission
  • Checked 2026-09-08
Care setting
DICOM-compatible clinical imaging workflow on off-the-shelf hardware; the FDA packet does not establish a specific facility type
  • Exact FDA submission
  • Checked 2026-09-08
Intended users
Clinicians with proper training
  • Exact FDA submission
  • Checked 2026-09-08
Limitations
The software is a support tool and patient-management decisions must not be based solely on its results. The FDA scope is limited to segmentable brain structures and WMH from the specified T1 and T2 FLAIR MRI inputs; WMH performance varies materially by lesion burden, and the packet does not establish a general diagnostic claim or a specific current PACS, RIS, EHR, or deployment configuration.
  • Exact FDA submission
  • Checked 2026-09-08
Evidence summary
FDA K232231 reports software verification and validation against manual expert segmentation. For T1 brain structures, Dice scores and relative-volume differences were GM 0.983 and 2.846, WM 0.990 and 1.643, CSF 0.955 and 7.495, and ICV 0.994 and 0.496. For T2 FLAIR WMH, burden-stratified Dice scores were 0.506, 0.636, 0.774, and 0.885 for low, medium, high, and very high WMH, with corresponding absolute volume errors of 0.675, 2.544, 3.097, and 7.833 mL; the packet also reports Pearson correlation of 0.88 for WMH count. No sample size or clinical-outcome study is stated in the reviewed FDA packet.
  • Exact FDA submission
  • Checked 2026-09-08

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
FDA K232231 reports software verification and validation against manual expert segmentation. For T1 brain structures, Dice scores and relative-volume differences were GM 0.983 and 2.846, WM 0.990 and 1.643, CSF 0.955 and 7.495, and ICV 0.994 and 0.496. For T2 FLAIR WMH, burden-stratified Dice scores were 0.506, 0.636, 0.774, and 0.885 for low, medium, high, and very high WMH, with corresponding absolute volume errors of 0.675, 2.544, 3.097, and 7.833 mL; the packet also reports Pearson correlation of 0.88 for WMH count. No sample size or clinical-outcome study is stated in the reviewed FDA packet.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Reported sensitivity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

Reported specificity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

QP-Brain® regulatory performance study summary

  • DesignStand-Alone Performance
  • PopulationAdult
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K232231
  • IndependenceIndependence not established

K232231

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K232231. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-08
Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not verified: Pricing information is recorded but not verified against a resolving source.

Reimbursement

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: DICOM-compatible MR images, including 3D T1-weighted Gradient Echo structural MRI for brain structures and T2 FLAIR MRI for WMH identification, from adults and adolescents aged 18 years and older; the application checks available sequences for compatibility before launching analysis.

    Scope: Exact FDA submission.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-09-08 · FDA K232231: QP-Brain®

  2. Routing

    Source context, not route validation: DICOM-compatible input and return/display of outputs on third-party DICOM workstations and PACS are described. Exact DICOM objects, interfaces, versions, routing, and current local compatibility are not established by the FDA packet.

    Scope: Exact FDA submission.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-09-08 · FDA K232231: QP-Brain®

  3. AI processing

    Source context, not route validation: The exact FDA package describes QP-Brain as software that operates on off-the-shelf hardware from multiple vendors. Current hosting, data flow, and commercial configuration require vendor confirmation.

    Scope: Exact FDA submission. Packet wording: "and operate on off-the-self hardware (multiple vendors).". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-11-14 · FDA K232231: QP-Brain®

  4. Returned objects

    Source context, not route validation: Volumes for gray matter, white matter, cerebrospinal fluid, and intracranial volume, plus WMH quantification, segmentation color overlays, morphometric reports, comparisons with age- and gender-matched normative percentile data when available, longitudinal reporting, and DICOM-formatted results displayable on DICOM workstations and PACS.

    Scope: Exact FDA submission.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-08 · FDA K232231: QP-Brain®

  5. Viewer/worklist/report

    Unresolved route: Volumes for gray matter, white matter, cerebrospinal fluid, and intracranial volume, plus WMH quantification, segmentation color overlays, morphometric reports, comparisons with age- and gender-matched normative percentile data when available, longitudinal reporting, and DICOM-formatted results displayable on DICOM workstations and PACS.

    Scope: Exact FDA submission.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-08 · FDA K232231: QP-Brain®

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Deployment
The exact FDA package describes QP-Brain as software that operates on off-the-shelf hardware from multiple vendors. Current hosting, data flow, and commercial configuration require vendor confirmation.
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "and operate on off-the-self hardware (multiple vendors).". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.

Integration
DICOM-compatible input and return/display of outputs on third-party DICOM workstations and PACS are described. Exact DICOM objects, interfaces, versions, routing, and current local compatibility are not established by the FDA packet.
  • Exact FDA submission
  • Checked 2026-09-08
Security
Quibim's security and privacy statement describes an ISO 27001-certified information security management system and ISO 13485-certified quality management system; QP-Care platform data hosted on Microsoft Azure protected by multi-factor authentication and encryption in transit and at rest; pseudonymization of processed data; regular penetration tests following OSSTMM and OWASP methodologies; staff security training and confidentiality obligations.
  • Vendor-wide
  • Checked 2026-10-27

Vendor statement; certificates were not read. Applicant name variants ('QUIBIM S.L.', 'Quibim S.L.') are bridged via the identical distinctive name and QP product naming. Exact-release scope is not established.

Input or output interoperability

DICOM is explicitly referenced in the described input, output, or workflow.

exact submission · checked 2026-09-01

The model card does not establish every supported DICOM object, transfer method, or local interface.

Result destination

PACS is explicitly referenced as a result destination.

exact submission · checked 2026-09-01

Confirm supported products, versions, routing behavior, and display requirements during local evaluation.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for QP-Brain®; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-08
Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K232231
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
MRI
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Anatomy
Brain
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2023-12-13
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
23Fields reviewed
9Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not verified - checked 2026-09-08 - Exact FDA submission

    No public price or quote for the exact K232231 configuration was established in this review.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Reimbursement: Not publicly available - checked 2026-09-08 - Exact FDA submission

    No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Installation timeline: Not yet reviewed
  • Training: Not verified - checked 2026-09-08 - Exact FDA submission

    Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

  • Monitoring: Not verified - checked 2026-09-08 - Exact FDA submission

    Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

8 sources - 3 history items

Source ledger

  1. FDA K232231: QP-Brain®

    U.S. Food and Drug Administration - Published 2023-12-13 - Accessed 2026-09-08

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation, deployment

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. QP-Brain® model card (K232231)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, evidenceSummary, integration, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K232231

    U.S. Food and Drug Administration - Accessed 2026-09-08

    Scope: postmarketSafety

  8. Quibim: Security and Privacy Statement

    Quibim - Accessed 2026-10-27

    Scope: security

Research history

  1. K232231 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K261405

ANDI 2.2

Imeka Solutions, Inc.

Authorized purpose: ANDI 2.2 processes brain MR images for atlas-assisted visualization, segmentation and volumetric quantification, presenting adjunctive DICOM-readable information for trained medical professionals; it does not predict, diagnose or interpret brain health.

Modality
MRI
Anatomy
Brain
Decision
2026-05-28
FDA-cleared K260032

Morph

BeauBrain Healthcare, Inc.

Authorized purpose: Morph automatically labels, visualizes, and volumetrically quantifies segmentable brain structures from MR images and may compare measurements with reference percentile data; it is adjunctive and not the sole basis for final decisions.

Modality
MRI
Anatomy
Brain
Decision
2026-05-13
FDA-cleared K253057

AI-Rad Companion Brain MR

Siemens Healthcare GmbH

Authorized purpose: Brain MRI morphometry and white matter hyperintensity quantification, including normative comparison and longitudinal change assessment for clinician review.

Modality
MRI
Anatomy
Brain
Decision
2026-01-22
FDA-cleared K252298

ANDI 2.0

Imeka Solutions, Inc.

Authorized purpose: Adjunctive quantitative brain MRI analysis that compares patient measures with normative and longitudinal references; a physician reviews all results and makes the clinical decision.

Modality
MRI
Anatomy
Brain
Decision
2025-10-22

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination

Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.

Accepted repository change history

No entry has been added to this scoped change history for this record. This is not proof that the record never changed.

Correction policy and limits · Public repository change history