Authorized purpose: Overjet Periapical Radiolucency Assist is a concurrent-read aid for detecting periapical radiolucencies on periapical radiographs of permanent teeth in patients aged 12 years and older; it supplements, but does not replace, the dentist's review.
Clinical role
Concurrent review support for labeled intraoral or bitewing radiographic findings.
Images and patients
X-ray/Radiography · Permanent teeth; periapical regions
Key limitation
Keep in mind: Use is limited to periapical radiographs of permanent teeth in patients aged 12 years and older; the aid does not replace the dentist's complete review or judgment.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: Overjet Periapical Radiolucency Assist is a concurrent-read aid for detecting periapical radiolucencies on periapical radiographs of permanent teeth in patients aged 12 years and older; it supplements, but does not replace, the dentist's review.
Which images or clinical context are in scope?Modality: X-ray/Radiography Anatomy or clinical context: Permanent teeth; periapical regions Input: Supported periapical or bitewing radiographs from compatible dental radiography systems.
Who uses it?Intended users: Trained and licensed dentists.
What does the software produce?Output: Candidate findings, localized regions, segmentations, and review displays.
What is its most important limitation?Key limitation: Use is limited to periapical radiographs of permanent teeth in patients aged 12 years and older; the aid does not replace the dentist's complete review or judgment.
Clinical takeaway. Clinical role: Overjet Periapical Radiolucency Assist is a concurrent-read aid for detecting periapical radiolucencies on periapical radiographs of permanent teeth in patients aged 12 years and older; it supplements, but does not replace, the dentist's review.
Evidence takeaway. Evidence from the reviewed source record: K231678 FDA decision material describes the cleared dental radiograph analysis software, labeled use, and submission-specific verification. It does not establish product-wide sensitivity, specificity, outcomes, or independent clinical validation for this exact configuration.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
Overjet Periapical Radiolucency Assist is a concurrent-read aid for detecting periapical radiolucencies on periapical radiographs of permanent teeth in patients aged 12 years and older; it supplements, but does not replace, the dentist's review.
Use is limited to periapical radiographs of permanent teeth in patients aged 12 years and older; the aid does not replace the dentist's complete review or judgment.
K231678 FDA decision material describes the cleared dental radiograph analysis software, labeled use, and submission-specific verification. It does not establish product-wide sensitivity, specificity, outcomes, or independent clinical validation for this exact configuration.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
K231678 FDA decision material describes the cleared dental radiograph analysis software, labeled use, and submission-specific verification. It does not establish product-wide sensitivity, specificity, outcomes, or independent clinical validation for this exact configuration.
Sensitivity is not promoted as a product-wide endpoint. Compare only an exact endpoint, threshold, population, version, and source if later verified.
Reported specificity
Not applicable
Not applicable
Not applicable
Checked 2026-09-04
Specificity is not promoted as a product-wide endpoint. Compare only an exact endpoint, threshold, population, version, and source if later verified.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
K231678 FDA evidence package for Overjet Periapical Radiolucency Assist
DesignFDA 510(k) or PMA decision evidence and product verification described in the exact submission.
PopulationPermanent teeth; periapical regions
Scopeexact submission
Tested versionOverjet Periapical Radiolucency Assist configuration described in K231678
IndependenceSponsor or vendor study
K231678
Manufacturer evidence submitted for the exact FDA decision. It is not independent clinical validation, and local technical, clinical, privacy, security, and workflow acceptance remains necessary.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K231678 by exact native submission identifier. This does not exclude product-family, later, or adverse-event information.
Public source
Exact FDA submission
Checked 2026-09-04
Exact-identifier review does not prove absence of all product-family safety signals.
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging or clinical service line and verify payer, site-of-service, coding, and coverage assumptions independently.
No separate named-product payment identified in reviewed public material
Centers for Medicare & Medicaid Services · United States
This is not a product-specific reimbursement determination. Build a local total-cost and value model and verify payer, site-of-service, coding, and coverage assumptions independently.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: Supported periapical or bitewing radiographs from compatible dental radiography systems.
Scope: Exact FDA submission.
Confirm scanner/model, protocol, series completeness and both software builds.
Unresolved route: Not established for the exact route.
Scope: Exact FDA submission. The exact public FDA decision material does not establish a complete integration contract. Confirm DICOM, worklist, identity, interfaces, exports, routing, audit behavior, and downtime locally.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Field checked 2026-09-04 ·
AI processing
Unresolved route: Not established for the exact route.
Scope: Exact FDA submission. The exact public FDA decision material does not establish the complete deployment topology, hosting responsibility, availability, upgrade path, or service contract.
Confirm processing location, queue ownership, exclusions and exact AI build.
Field checked 2026-09-04 ·
Returned objects
Source context, not route validation: Candidate findings, localized regions, segmentations, and review displays.
Scope: Exact FDA submission.
Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.
Unresolved route: Candidate findings, localized regions, segmentations, and review displays.
Scope: Exact FDA submission.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
Overjet's trust center states a completed independent HITRUST audit and certification, HIPAA business associate compliance, GDPR and CCPA alignment, third-party penetration tests, need-to-know data access and employee security training. Vendor-wide.
Vendor-wide
Checked 2026-10-25
Vendor statement, not an attestation; exact-release scope is not established.
Only trained and licensed dentists should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
Monitor K231678 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Use is limited to periapical radiographs of permanent teeth in patients aged 12 years and older; the aid does not replace the dentist's complete review or judgment.
Exact FDA submission
Checked 2026-09-04
Monitoring guidance is buyer-facing implementation guidance derived from the exact cleared scope; it is not a claim that the FDA package provides a live monitoring service.
The exact public deployment topology remains incomplete; establish hosting responsibility, data flow, availability, upgrade, service, and downtime boundaries locally.
exact submission · checked 2026-09-04
FDA-described deployment boundaries do not establish current hosting contracts, security controls, uptime, or local qualification.
The exact public integration contract remains incomplete; establish and test supported input, output, identity, routing, archive, and downtime behavior locally.
exact submission · checked 2026-09-04
Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.
Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Use is limited to periapical radiographs of permanent teeth in patients aged 12 years and older; the aid does not replace the dentist's complete review or judgment.
exact submission · checked 2026-09-04
Define stop-use, escalation, rollback, and revalidation triggers before production use.
Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.
product family · checked 2026-09-04
No exact-product price was found in the reviewed public material; productivity or value claims require local evidence.
Monitor K231678 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Use is limited to periapical radiographs of permanent teeth in patients aged 12 years and older; the aid does not replace the dentist's complete review or judgment.
exact submission · checked 2026-09-04
Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. dental radiograph analysis software limits remain submission-specific.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Human reviewedStatus
2026-09-20Last searched
27Fields reviewed
14Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, acr ai central product, vendor documentation, fda decision material, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not publicly available - checked 2026-09-04 - Exact FDA submission
No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.
Pricing model: Not yet reviewed
Setup and integration fees: Not yet reviewed
Recurring fees: Not yet reviewed
Minimum commitment: Not yet reviewed
Contract duration: Not yet reviewed
Renewal terms: Not yet reviewed
Cancellation terms: Not yet reviewed
Support and service terms: Not yet reviewed
Trial or pilot terms: Not yet reviewed
Deployment: Not publicly available - checked 2026-09-04 - Exact FDA submission
The exact public FDA decision material does not establish the complete deployment topology, hosting responsibility, availability, upgrade path, or service contract.
Integration: Not publicly available - checked 2026-09-04 - Exact FDA submission
The exact public FDA decision material does not establish a complete integration contract. Confirm DICOM, worklist, identity, interfaces, exports, routing, audit behavior, and downtime locally.
Authorized purpose: Concurrent-read CADe to assist dentists in detecting specified tooth and jaw anatomy, bone-level landmarks, and root-canal-treated teeth in digital radiographs of permanent teeth in patients 15 years or older.
Modality
X-ray/Radiography
Anatomy
Permanent teeth and jaw/maxillofacial anatomy, including enamel, dentine, pulp, root, mental foramen, inferior alveolar canal, maxillary sinus, cementoenamel junction, crestal bone, and root-canal-treated teeth
Authorized purpose: Analyze frontal chest and abdominal X-ray images to detect and localize a nasally or orally inserted enteric tube, its tip and side port if present, and the diaphragm and airways for radiologist review.
Modality
X-ray/Radiography
Anatomy
Chest/lung and abdomen/stomach; enteric tube, tube tip and side port, diaphragm, and airways including carina and left/right bronchi
Authorized purpose: qXR-Detect is an adult chest-radiograph concurrent reading aid that identifies, highlights and categorizes suspicious regions of interest across lung, pleura, mediastinum/hila/heart, bone, hardware and other categories; trained clinicians remain responsible for primary interpretation.
Modality
X-ray/Radiography
Anatomy
Chest/Lung; pleura, mediastinum/hila/heart, bone, hardware and other chest-radiograph regions
Authorized purpose: Provides automated concurrent-read detection and segmentation of caries on bitewing radiographs in patients 18 and older, adding information for the dentist while requiring complete dentist review and judgment.
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.
Accepted repository change history
No entry has been added to this scoped change history for this record. This is not proof that the record never changed.