FDA-cleared profile

Overjet Dental Assist

Overjet, Inc. · X-ray/Radiography · FDA-cleared (510(k))

Overjet Dental Assist is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, dentists and dental hygienists. Intended Patient Population: The intended patient population of the device is patients living in the United States, who are 22 years old or older, and that do not have any remaining primary teeth. Overjet has not evaluated the performance of the device on primary dentition.

Clinical role
Measurement or quantification
Images and patients
X-ray/Radiography · Dental/Maxillofacial
Key limitation
Scanner manufacturers: All

Overview status: Field-derived fact summary This section is assembled from the sourced field records below; it is not a source-reviewed editorial overview.

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • Imaging and anatomyX-ray/Radiography · Dental/Maxillofacial
  • Workflow roleMeasurement or quantification
  • Required inputX-ray/Radiography images or data involving Dental/Maxillofacial; consult the FDA labeling for exact protocol, series, and quality requirements.
  • Output and human actionThe AI model outputs measurements of mesial and distal (interproximal) bone levels associated with each tooth from bitewing and periapical radiographs. The system provides annotations and linear distance measurements of interproximal bone levels, which dental professionals can review, view, and modify.

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
Overjet Dental Assist is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, dentists and dental hygienists. Intended Patient Population: The intended patient population of the device is patients living in the United States, who are 22 years old or older, and that do not have any remaining primary teeth. Overjet has not evaluated the performance of the device on primary dentition.
  • Exact FDA submission
  • Checked 2026-08-31

Clinical-use display mirrors the best exact-submission purpose source because the reviewed FDA packet did not provide a more specific usable abstraction.

Workflow role
Measurement or quantification
  • Exact FDA submission
  • Checked 2026-08-31

Conservative normalization of explicit wording in an exact-submission source.

Required input
X-ray/Radiography images or data involving Dental/Maxillofacial; consult the FDA labeling for exact protocol, series, and quality requirements.
  • Exact FDA submission
  • Checked 2026-08-31

Conservative normalization of explicit wording in an exact-submission source.

Output and clinician action
The AI model outputs measurements of mesial and distal (interproximal) bone levels associated with each tooth from bitewing and periapical radiographs. The system provides annotations and linear distance measurements of interproximal bone levels, which dental professionals can review, view, and modify.
  • Exact FDA submission
  • Checked 2026-09-01

Clinical output reported in the exact-submission ACR model card.

Intended users
Trained dental professionals including, but not limited to, dentists and dental hygienists
  • Exact FDA submission
  • Checked 2026-09-01

Intended-user text reported in the exact-submission ACR model card.

Limitations
Scanner manufacturers: All
  • Exact FDA submission
  • Checked 2026-09-01

Limitations and supported-acquisition details reported in the exact-submission ACR model card.

Evidence summary
Study type: Stand-Alone Performance. Cases: 63. Reference standard: Interpretation by Reviewing Clinician. Reported sensitivity: 98.70%. Reported specificity: 95.00%
  • Exact FDA submission
  • Checked 2026-09-01

Structured study metadata from the exact-submission ACR model card; this is not an independent evidence synthesis.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
Study type: Stand-Alone Performance. Cases: 63. Reference standard: Interpretation by Reviewing Clinician. Reported sensitivity: 98.70%. Reported specificity: 95.00%
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Structured study metadata from the exact-submission ACR model card; this is not an independent evidence synthesis.

Reported sensitivity
98.70%
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Model sensitivity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

Reported specificity
95.00%
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Model specificity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

Overjet Dental Assist regulatory performance study summary

  • DesignStand-Alone Performance
  • PopulationPopulation described in the exact-submission ACR model card
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K210187
  • Samplen=63
  • IndependenceIndependence not established

K210187

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K210187. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-09
Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not verified: Pricing information is recorded but not verified against a resolving source.

Reimbursement

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: X-ray/Radiography images or data involving Dental/Maxillofacial; consult the FDA labeling for exact protocol, series, and quality requirements.

    Scope: Exact FDA submission. Conservative normalization of explicit wording in an exact-submission source.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-08-31 · FDA K210187: Overjet Dental Assist

  2. Routing

    Unresolved route: Not established for the exact route.

    Scope: Exact FDA submission. The exact FDA decision document was reviewed on 2026-11-14 and does not state the subject device's DICOM, PACS, RIS, or EHR interfaces. Confirm supported DICOM services and PACS, RIS, EHR, and interface versions with the vendor.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-11-14 ·

  3. AI processing

    Unresolved route: Not established for the exact route.

    Scope: Exact FDA submission. The reviewed sources do not establish the current exact-release hosting topology, data flow, region, tenancy, or disaster-recovery configuration.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-09-09 ·

  4. Returned objects

    Source context, not route validation: The AI model outputs measurements of mesial and distal (interproximal) bone levels associated with each tooth from bitewing and periapical radiographs. The system provides annotations and linear distance measurements of interproximal bone levels, which dental professionals can review, view, and modify.

    Scope: Exact FDA submission. Clinical output reported in the exact-submission ACR model card.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-01 · Overjet Dental Assist model card (K210187)

  5. Viewer/worklist/report

    Unresolved route: The AI model outputs measurements of mesial and distal (interproximal) bone levels associated with each tooth from bitewing and periapical radiographs. The system provides annotations and linear distance measurements of interproximal bone levels, which dental professionals can review, view, and modify.

    Scope: Exact FDA submission. Clinical output reported in the exact-submission ACR model card.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-01 · Overjet Dental Assist model card (K210187)

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Security
Overjet's trust center states a completed independent HITRUST audit and certification, HIPAA business associate compliance, GDPR and CCPA alignment, third-party penetration tests, need-to-know data access and employee security training. Vendor-wide.
  • Vendor-wide
  • Checked 2026-10-25

Vendor statement, not an attestation; exact-release scope is not established.

Supported acquisition environment

Scanner manufacturers: All

exact submission · checked 2026-09-01

Confirm currently supported software versions, protocols, and site-specific acquisition configurations.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for Overjet Dental Assist; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-09
Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K210187
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
X-ray/Radiography
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Anatomy
Dental/Maxillofacial
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Anatomy normalized from explicit exact-submission ACR model-card wording.

Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2021-05-19
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-19Last searched
25Fields reviewed
11Source classes checked
0Awaiting review PubMed leads
0Awaiting review trial leads
0Unreviewed FDA recall leads

Exact-submission ACR model-card fields were normalized under the current provenance rules; vendor and independent-study confirmation remain distinct. Exact FDA decision-summary packet reviewed on 2026-09-19. Applied FDA-supported reader facts: none. Facts marked unknown in the FDA packet retain any prior non-FDA source without reclassification. Candidate literature and postmarket candidates remain in their existing review state. Automated exact-name discovery found 0 PubMed and 0 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, acr ai central product, vendor documentation, fda device recall, literature index, trial registry, fda decision summary packet, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Care setting: Not publicly available - checked 2026-11-14 - Exact FDA submission

    The exact FDA decision document was reviewed on 2026-11-14 and does not state a care setting or patient population for the subject device. Confirm intended sites and patients with the vendor and local policy.

  • Public pricing: Not verified - checked 2026-09-09 - Exact FDA submission

    No public price or quote for the exact K210187 configuration was established in this review.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Reimbursement: Not publicly available - checked 2026-09-09 - Exact FDA submission

    No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

  • Deployment: Not verified - checked 2026-09-09 - Exact FDA submission

    The reviewed sources do not establish the current exact-release hosting topology, data flow, region, tenancy, or disaster-recovery configuration.

  • Integration: Not publicly available - checked 2026-11-14 - Exact FDA submission

    The exact FDA decision document was reviewed on 2026-11-14 and does not state the subject device's DICOM, PACS, RIS, or EHR interfaces. Confirm supported DICOM services and PACS, RIS, EHR, and interface versions with the vendor.

  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Installation timeline: Not yet reviewed
  • Training: Not verified - checked 2026-09-09 - Exact FDA submission

    Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

  • Monitoring: Not verified - checked 2026-09-09 - Exact FDA submission

    Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

8 sources - 2 history items

Source ledger

  1. FDA K210187: Overjet Dental Assist

    U.S. Food and Drug Administration - Published 2021-05-19 - Accessed 2026-09-09

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. Overjet Dental Assist model card (K210187)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, output, limitations, reportedSensitivity, reportedSpecificity, evidenceSummary, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K210187

    U.S. Food and Drug Administration - Accessed 2026-09-09

    Scope: postmarketSafety

  8. Overjet Trust Center

    Overjet - Accessed 2026-10-25

    Scope: security

Research history

  1. K210187 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

Alternatives

Related tools to compare

FDA-cleared K252086

DTX Studio Assist

Nobel Biocare C/O Medicim NV

Authorized purpose: Analyzes two-dimensional bitewing and periapical radiographs of permanent teeth in patients aged 15 and older to identify and localize caries, calculus, periapical radiolucency, root canal filling deficiency, discrepancy at an existing restoration margin, and bone loss; also supports alveolar bone-level measurement and restoration/dental-anatomy segmentation as an adjunct to dentist review.

Modality
X-ray/Radiography
Anatomy
Dental/Maxillofacial
Decision
2025-11-17
FDA-cleared K242437

Smile Dx®

Cube Click, Inc.

Authorized purpose: Aids dentists reviewing bitewing and periapical radiographs of permanent teeth by detecting/segmenting caries and periapical radiolucencies, identifying restorations and dental anatomy, and measuring mesial/distal bone levels; it does not replace dentist review.

Modality
X-ray/Radiography
Anatomy
Dental/Maxillofacial
Decision
2025-05-14
FDA-cleared K250264

SugarBug (1.x)

Bench7, Inc.

Authorized purpose: Provides automated concurrent-read detection and segmentation of caries on bitewing radiographs in patients 18 and older, adding information for the dentist while requiring complete dentist review and judgment.

Modality
X-ray/Radiography
Anatomy
Dental/Maxillofacial
Decision
2025-11-07
FDA-cleared K243234

Second Opinion® CS

Pearl Inc.

Authorized purpose: Acts as a CADe second reader to detect and segment caries on periapical radiographs of permanent teeth in patients aged 12 years and older; clinicians review the contours and percentage analysis in context.

Modality
X-ray/Radiography
Anatomy
Dental/Maxillofacial
Decision
2025-06-12

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination

Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.

Accepted repository change history

No entry has been added to this scoped change history for this record. This is not proof that the record never changed.

Correction policy and limits · Public repository change history