Authorized purpose: Dental radiograph charting assistance for detecting natural and restorative dental structures and tooth numbering on bitewing, periapical, and panoramic images.
Clinical role
Detection and charting aid; the dental professional remains responsible for complete image review and clinical judgment using the image and patient history.
Images and patients
X-ray/Radiography · Dental/maxillofacial
Key limitation
Keep in mind: The device is not a replacement for complete clinician review or clinical judgment. K241684 adds an FDA-authorized PCCP for manually applied, locked algorithm modifications; the packet does not establish new performance results for this submission.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: Dental radiograph charting assistance for detecting natural and restorative dental structures and tooth numbering on bitewing, periapical, and panoramic images.
Which images or clinical context are in scope?Modality: X-ray/Radiography Anatomy or clinical context: Dental/maxillofacial Input: Two-dimensional bitewing and periapical dental radiographs for patients aged 5 years or older with primary and/or permanent teeth, and panoramic radiographs for patients with only permanent teeth.
Who uses it?Intended users: Dental professionals.
What does the software produce?Output: Detection of natural dental structures including enamel, pulp, and tooth numbering, plus implants, crowns, endodontic treatment, and fillings to support dental charting data.
What is its most important limitation?Key limitation: The device is not a replacement for complete clinician review or clinical judgment. K241684 adds an FDA-authorized PCCP for manually applied, locked algorithm modifications; the packet does not establish new performance results for this submission.
Clinical takeaway. Clinical role: Dental radiograph charting assistance for detecting natural and restorative dental structures and tooth numbering on bitewing, periapical, and panoramic images.
Evidence takeaway. Evidence from the reviewed source record: The FDA packet states that the subject device is identical to Overjet Charting Assist cleared under predicate K233590 except for inclusion of the Predetermined Change Control Plan. No additional performance testing was conducted, and the prior testing remains applicable. The reviewed K241684 packet does not provide numeric sensitivity, specificity, accuracy, sample size, sites, or reference-standard details.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
Dental radiograph charting assistance for detecting natural and restorative dental structures and tooth numbering on bitewing, periapical, and panoramic images.
Detection and charting aid; the dental professional remains responsible for complete image review and clinical judgment using the image and patient history.
Two-dimensional bitewing and periapical dental radiographs for patients aged 5 years or older with primary and/or permanent teeth, and panoramic radiographs for patients with only permanent teeth.
Detection of natural dental structures including enamel, pulp, and tooth numbering, plus implants, crowns, endodontic treatment, and fillings to support dental charting data.
Dental imaging and charting workflow; bitewing and periapical use includes patients aged 5 years or older with primary and/or permanent teeth, while panoramic use is for patients with only permanent teeth.
The device is not a replacement for complete clinician review or clinical judgment. K241684 adds an FDA-authorized PCCP for manually applied, locked algorithm modifications; the packet does not establish new performance results for this submission.
The FDA packet states that the subject device is identical to Overjet Charting Assist cleared under predicate K233590 except for inclusion of the Predetermined Change Control Plan. No additional performance testing was conducted, and the prior testing remains applicable. The reviewed K241684 packet does not provide numeric sensitivity, specificity, accuracy, sample size, sites, or reference-standard details.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
The FDA packet states that the subject device is identical to Overjet Charting Assist cleared under predicate K233590 except for inclusion of the Predetermined Change Control Plan. No additional performance testing was conducted, and the prior testing remains applicable. The reviewed K241684 packet does not provide numeric sensitivity, specificity, accuracy, sample size, sites, or reference-standard details.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
Overjet Charting Assist regulatory performance study summary
DesignStand-Alone Performance
PopulationAdult and Pediatric
Scopeexact submission
Tested versionExact product version not reported; regulatory study summarized for FDA submission K241684
IndependenceIndependence not established
K241684
ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K241684. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: Two-dimensional bitewing and periapical dental radiographs for patients aged 5 years or older with primary and/or permanent teeth, and panoramic radiographs for patients with only permanent teeth.
Scope: Exact FDA submission.
Confirm scanner/model, protocol, series completeness and both software builds.
Unresolved route: Not established for the exact route.
Scope: Exact FDA submission. The exact FDA decision document was reviewed on 2026-11-14 and does not state the subject device's DICOM, PACS, RIS, or EHR interfaces. Confirm supported DICOM services and PACS, RIS, EHR, and interface versions with the vendor.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Field checked 2026-11-14 ·
AI processing
Unresolved route: Not established for the exact route.
Scope: Exact FDA submission. The exact FDA decision document was reviewed on 2026-11-14 and does not state where the subject device's software runs or how it is hosted. Confirm the deployment model, hosting, data flow, and current commercial configuration with the vendor.
Confirm processing location, queue ownership, exclusions and exact AI build.
Field checked 2026-11-14 ·
Returned objects
Source context, not route validation: Detection of natural dental structures including enamel, pulp, and tooth numbering, plus implants, crowns, endodontic treatment, and fillings to support dental charting data.
Scope: Exact FDA submission.
Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.
Unresolved route: Detection of natural dental structures including enamel, pulp, and tooth numbering, plus implants, crowns, endodontic treatment, and fillings to support dental charting data.
Scope: Exact FDA submission.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
Overjet's trust center states a completed independent HITRUST audit and certification, HIPAA business associate compliance, GDPR and CCPA alignment, third-party penetration tests, need-to-know data access and employee security training. Vendor-wide.
Vendor-wide
Checked 2026-10-25
Vendor statement, not an attestation; exact-release scope is not established.
ACR AI Central provides an exact-submission model card for Overjet Charting Assist; FDA labeling remains controlling for clinical use, limitations, and performance.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Exact FDA record reviewedStatus
2026-09-20Last searched
22Fields reviewed
10Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, acr ai central product, vendor documentation, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not verified - checked 2026-09-08 - Exact FDA submission
No public price or quote for the exact K241684 configuration was established in this review.
Pricing model: Not yet reviewed
Setup and integration fees: Not yet reviewed
Recurring fees: Not yet reviewed
Minimum commitment: Not yet reviewed
Contract duration: Not yet reviewed
Renewal terms: Not yet reviewed
Cancellation terms: Not yet reviewed
Support and service terms: Not yet reviewed
Trial or pilot terms: Not yet reviewed
Reimbursement: Not publicly available - checked 2026-09-08 - Exact FDA submission
No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.
Deployment: Not publicly available - checked 2026-11-14 - Exact FDA submission
The exact FDA decision document was reviewed on 2026-11-14 and does not state where the subject device's software runs or how it is hosted. Confirm the deployment model, hosting, data flow, and current commercial configuration with the vendor.
Integration: Not publicly available - checked 2026-11-14 - Exact FDA submission
The exact FDA decision document was reviewed on 2026-11-14 and does not state the subject device's DICOM, PACS, RIS, or EHR interfaces. Confirm supported DICOM services and PACS, RIS, EHR, and interface versions with the vendor.
Data hosting location: Not yet reviewed
Data retention: Not yet reviewed
Documented data use: Not yet reviewed
Hardware requirements: Not yet reviewed
Network requirements: Not yet reviewed
Identity and access: Not yet reviewed
Installation timeline: Not yet reviewed
Training: Not verified - checked 2026-09-08 - Exact FDA submission
Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.
Monitoring: Not verified - checked 2026-09-08 - Exact FDA submission
Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.
Authorized purpose: Provides automated concurrent-read detection and segmentation of caries on bitewing radiographs in patients 18 and older, adding information for the dentist while requiring complete dentist review and judgment.
Authorized purpose: Assists dentists in detecting caries, fillings/restorations, fixed prostheses, and implants on digital bitewing, periapical, and panoramic radiographs of permanent teeth in patients aged 15 years and older; final diagnoses and treatment plans remain the dentist's responsibility.
Authorized purpose: Acts as a CADe second reader to detect and segment caries on periapical radiographs of permanent teeth in patients aged 12 years and older; clinicians review the contours and percentage analysis in context.
Authorized purpose: Aids dentists reviewing bitewing and periapical radiographs of permanent teeth by detecting/segmenting caries and periapical radiolucencies, identifying restorations and dental anatomy, and measuring mesial/distal bone levels; it does not replace dentist review.
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.
Accepted repository change history
No entry has been added to this scoped change history for this record. This is not proof that the record never changed.