FDA-cleared profile

Overjet Caries Assist

Overjet, Inc. · X-ray/Radiography · FDA-cleared (510(k))

Authorized purpose: Overjet Caries Assist is a concurrent-read aid for detecting and segmenting caries on bitewing radiographs of permanent dentition in patients aged 18 years and older; it provides additional information for a dentist and is not a replacement for complete review or clinical judgment.

Clinical role
Concurrent-read detection, localization, segmentation, and review support for labeled dental radiographs.
Images and patients
X-ray/Radiography · Permanent dentition; tooth surfaces and caries
Key limitation
Keep in mind: Use is limited to bitewing radiographs of permanent dentition in patients aged 18 years and older; the labeling warns of false positives and negatives and states the system is not 100% sensitive.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: Overjet Caries Assist is a concurrent-read aid for detecting and segmenting caries on bitewing radiographs of permanent dentition in patients aged 18 years and older; it provides additional information for a dentist and is not a replacement for complete review or clinical judgment.
  • Which images or clinical context are in scope?Modality: X-ray/Radiography Anatomy or clinical context: Permanent dentition; tooth surfaces and caries Input: Supported intraoral, bitewing, or periapical radiographs.
  • Who uses it?Intended users: Trained and licensed dentists.
  • What does the software produce?Output: Candidate dental findings, localized regions, segmentations, and review displays.
  • What is its most important limitation?Key limitation: Use is limited to bitewing radiographs of permanent dentition in patients aged 18 years and older; the labeling warns of false positives and negatives and states the system is not 100% sensitive.

Clinical takeaway. Clinical role: Overjet Caries Assist is a concurrent-read aid for detecting and segmenting caries on bitewing radiographs of permanent dentition in patients aged 18 years and older; it provides additional information for a dentist and is not a replacement for complete review or clinical judgment.

Evidence takeaway. Evidence from the reviewed source record: The exact FDA document reports standalone testing on 352 images, summarized per surface (7,129 surfaces): overall sensitivity 72.0% (62.9%, 81.1%) and specificity 98.1% (97.7%, 98.5%). In a multi reader fully crossed study, 13 US licensed dentists read 352 radiographs (104 containing caries / 248 without caries) against consensus ground truth of three US licensed dentists; overall reader sensitivity improved from 57.9% (48.9%, 66.0%) to 76.2% (68.4%, 82.6%), and weighted AFROC AUC increased by 0.057 (0.039, 0.098). Results are as reported in the FDA submission, not independent validation.

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
Overjet Caries Assist is a concurrent-read aid for detecting and segmenting caries on bitewing radiographs of permanent dentition in patients aged 18 years and older; it provides additional information for a dentist and is not a replacement for complete review or clinical judgment.
  • Exact FDA submission
  • Checked 2026-09-04
Workflow role
Concurrent-read detection, localization, segmentation, and review support for labeled dental radiographs.
  • Exact FDA submission
  • Checked 2026-09-04
Required input
Supported intraoral, bitewing, or periapical radiographs.
  • Exact FDA submission
  • Checked 2026-09-04
Output and clinician action
Candidate dental findings, localized regions, segmentations, and review displays.
  • Exact FDA submission
  • Checked 2026-09-04
Care setting
Clinical dental imaging workflows.
  • Exact FDA submission
  • Checked 2026-09-04
Intended users
Trained and licensed dentists.
  • Exact FDA submission
  • Checked 2026-09-04
Limitations
Use is limited to bitewing radiographs of permanent dentition in patients aged 18 years and older; the labeling warns of false positives and negatives and states the system is not 100% sensitive.
  • Exact FDA submission
  • Checked 2026-09-04
Evidence summary
The exact FDA document reports standalone testing on 352 images, summarized per surface (7,129 surfaces): overall sensitivity 72.0% (62.9%, 81.1%) and specificity 98.1% (97.7%, 98.5%). In a multi reader fully crossed study, 13 US licensed dentists read 352 radiographs (104 containing caries / 248 without caries) against consensus ground truth of three US licensed dentists; overall reader sensitivity improved from 57.9% (48.9%, 66.0%) to 76.2% (68.4%, 82.6%), and weighted AFROC AUC increased by 0.057 (0.039, 0.098). Results are as reported in the FDA submission, not independent validation.
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "Standalone performance of the Overjet AI algorithm for the 352 images is shown in the tables below. Sensitivity and specificity were summarized based on surfaces, and 95% CIs were provided" … "A total of 7,129 surfaces were available and included in the analysis.". Read from the exact FDA decision document on 2026-11-14.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
The exact FDA document reports standalone testing on 352 images, summarized per surface (7,129 surfaces): overall sensitivity 72.0% (62.9%, 81.1%) and specificity 98.1% (97.7%, 98.5%). In a multi reader fully crossed study, 13 US licensed dentists read 352 radiographs (104 containing caries / 248 without caries) against consensus ground truth of three US licensed dentists; overall reader sensitivity improved from 57.9% (48.9%, 66.0%) to 76.2% (68.4%, 82.6%), and weighted AFROC AUC increased by 0.057 (0.039, 0.098). Results are as reported in the FDA submission, not independent validation.
  • FDA source
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "Standalone performance of the Overjet AI algorithm for the 352 images is shown in the tables below. Sensitivity and specificity were summarized based on surfaces, and 95% CIs were provided" … "A total of 7,129 surfaces were available and included in the analysis.". Read from the exact FDA decision document on 2026-11-14.

Reported sensitivity
Standalone per-surface sensitivity for caries on 352 images (7,129 surfaces): overall 72.0% (95% CI 62.9%-81.1%); primary caries 74.4% (95% CI 64.4%-84.4%); secondary caries 62.5% (95% CI 46.6%-78.4%).
  • FDA source
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "Standalone performance of the Overjet AI algorithm for the 352 images is shown in the tables below. Sensitivity and specificity were summarized based on surfaces, and 95% CIs were provided" … "A total of 7,129 surfaces were available and included in the analysis.". Read from the exact FDA decision document on 2026-11-14; this is the submission-reported standalone result, not independent validation.

Reported specificity
Standalone per-surface specificity for caries on 352 images (7,129 surfaces): overall 98.1% (95% CI 97.7%-98.5%).
  • FDA source
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "Standalone performance of the Overjet AI algorithm for the 352 images is shown in the tables below. Sensitivity and specificity were summarized based on surfaces, and 95% CIs were provided" … "A total of 7,129 surfaces were available and included in the analysis.". Read from the exact FDA decision document on 2026-11-14; this is the submission-reported standalone result, not independent validation.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

K212519 FDA evidence package for Overjet Caries Assist

  • DesignFDA 510(k) or PMA decision evidence and product verification described in the exact submission.
  • PopulationPermanent dentition; tooth surfaces and caries
  • Scopeexact submission
  • Tested versionOverjet Caries Assist configuration described in K212519
  • IndependenceSponsor or vendor study

K212519

Manufacturer evidence submitted for the exact FDA decision. It is not independent clinical validation, and local technical, clinical, privacy, security, and workflow acceptance remains necessary.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K212519 by exact native submission identifier. This does not exclude product-family, later, or adverse-event information.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-04

Exact-identifier review does not prove absence of all product-family safety signals.

Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not publicly available: Reviewed sources do not publish an exact-product procurement price with a stated offer scope.

Reimbursement

Reimbursement
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging or clinical service line and verify payer, site-of-service, coding, and coverage assumptions independently.
  • Service-level, not product-specific
  • Checked 2026-09-04

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

This is not a product-specific reimbursement determination. Build a local total-cost and value model and verify payer, site-of-service, coding, and coverage assumptions independently.

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: Supported intraoral, bitewing, or periapical radiographs.

    Scope: Exact FDA submission.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-09-04 · FDA 510(k) decision material K212519: Overjet Caries Assist

  2. Routing

    Source context, not route validation: The FDA package describes supported dental radiograph systems and images with minimum resolution of 500 by 500 pixels; the complete local interface contract remains unverified.

    Scope: Exact FDA submission. This is the interface boundary described in the exact FDA package; current local compatibility, routing, identity, archive, and downtime behavior still require acceptance testing.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-09-04 · FDA 510(k) decision material K212519: Overjet Caries Assist

  3. AI processing

    Source context, not route validation: The exact FDA package describes Overjet Caries Assist as a cloud native application on a web platform (Edge, Chrome, Firefox) for any operating system, with images pulled in from a clinic or dental office and results pushed to a dashboard. Current hosting, data flow, and commercial configuration require vendor confirmation.

    Scope: Exact FDA submission. Packet wording: "Logicon is a local software while Overjet Caries Assist is a cloud native application." … "Platform Windows PC Web - Edge, Chrome, Firefox OS Microsoft Window 7, 8, 10 Any". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-11-14 · FDA 510(k) decision material K212519: Overjet Caries Assist

  4. Returned objects

    Source context, not route validation: Candidate dental findings, localized regions, segmentations, and review displays.

    Scope: Exact FDA submission.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-04 · FDA 510(k) decision material K212519: Overjet Caries Assist

  5. Viewer/worklist/report

    Unresolved route: Candidate dental findings, localized regions, segmentations, and review displays.

    Scope: Exact FDA submission.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-04 · FDA 510(k) decision material K212519: Overjet Caries Assist

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Deployment
The exact FDA package describes Overjet Caries Assist as a cloud native application on a web platform (Edge, Chrome, Firefox) for any operating system, with images pulled in from a clinic or dental office and results pushed to a dashboard. Current hosting, data flow, and commercial configuration require vendor confirmation.
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "Logicon is a local software while Overjet Caries Assist is a cloud native application." … "Platform Windows PC Web - Edge, Chrome, Firefox OS Microsoft Window 7, 8, 10 Any". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.

Integration
The FDA package describes supported dental radiograph systems and images with minimum resolution of 500 by 500 pixels; the complete local interface contract remains unverified.
  • Exact FDA submission
  • Checked 2026-09-04

This is the interface boundary described in the exact FDA package; current local compatibility, routing, identity, archive, and downtime behavior still require acceptance testing.

Security
Overjet's trust center states a completed independent HITRUST audit and certification, HIPAA business associate compliance, GDPR and CCPA alignment, third-party penetration tests, need-to-know data access and employee security training. Vendor-wide.
  • Vendor-wide
  • Checked 2026-10-25

Vendor statement, not an attestation; exact-release scope is not established.

Training
Only trained and licensed dentists should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
  • Exact FDA submission
  • Checked 2026-09-04
Monitoring
Monitor K212519 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Use is limited to bitewing radiographs of permanent dentition in patients aged 18 years and older; the labeling warns of false positives and negatives and states the system is not 100% sensitive.
  • Exact FDA submission
  • Checked 2026-09-04

Monitoring guidance is buyer-facing implementation guidance derived from the exact cleared scope; it is not a claim that the FDA package provides a live monitoring service.

Release and scope verification

Map installed model, build, options, population, anatomy, inputs, outputs, and enabled functions to K212519 before acceptance.

exact submission · checked 2026-09-04

Later product-family features are not evidence that this cleared release includes them.

Deployment boundary

The exact public deployment topology remains incomplete; establish hosting responsibility, data flow, availability, upgrade, service, and downtime boundaries locally.

exact submission · checked 2026-09-04

FDA-described deployment boundaries do not establish current hosting contracts, security controls, uptime, or local qualification.

Interoperability acceptance

The FDA package describes supported dental radiograph systems and images with minimum resolution of 500 by 500 pixels; the complete local interface contract remains unverified.

exact submission · checked 2026-09-04

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Use is limited to bitewing radiographs of permanent dentition in patients aged 18 years and older; the labeling warns of false positives and negatives and states the system is not 100% sensitive.

exact submission · checked 2026-09-04

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-04

No exact-product price was found in the reviewed public material; productivity or value claims require local evidence.

Lifecycle monitoring

Monitor K212519 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Use is limited to bitewing radiographs of permanent dentition in patients aged 18 years and older; the labeling warns of false positives and negatives and states the system is not 100% sensitive.

exact submission · checked 2026-09-04

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. dental radiograph analysis software limits remain submission-specific.

Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K212519
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
X-ray/Radiography
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Anatomy
Permanent dentition; tooth surfaces and caries
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2022-05-10
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Human reviewedStatus
2026-09-20Last searched
27Fields reviewed
14Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, vendor documentation, fda decision material, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not publicly available - checked 2026-09-04 - Exact FDA submission

    No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Installation timeline: Not yet reviewed

8 sources - 3 history items

Source ledger

  1. FDA 510(k) decision material K212519: Overjet Caries Assist

    U.S. Food and Drug Administration - Published 2022-05-10 - Accessed 2026-09-04

    Scope: identity, purpose, clinicalUse, modality, anatomy, intendedUsers, careSetting, workflowRole, input, output, limitations, evidenceSummary

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. Overjet Caries Assist model card (K212519)

    American College of Radiology - Published 2026-06-26 - Accessed 2026-09-01

    Scope: anatomy

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. Medicare Coverage Database

    Centers for Medicare & Medicaid Services - Accessed 2026-09-01

    Scope: coverage, codingContext

  7. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  8. Overjet Trust Center

    Overjet - Accessed 2026-10-25

    Scope: security

Research history

  1. K212519 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    human reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K250264

SugarBug (1.x)

Bench7, Inc.

Authorized purpose: Provides automated concurrent-read detection and segmentation of caries on bitewing radiographs in patients 18 and older, adding information for the dentist while requiring complete dentist review and judgment.

Modality
X-ray/Radiography
Anatomy
Dental/Maxillofacial
Decision
2025-11-07
FDA-cleared K243234

Second Opinion® CS

Pearl Inc.

Authorized purpose: Acts as a CADe second reader to detect and segment caries on periapical radiographs of permanent teeth in patients aged 12 years and older; clinicians review the contours and percentage analysis in context.

Modality
X-ray/Radiography
Anatomy
Dental/Maxillofacial
Decision
2025-06-12
FDA-cleared K242437

Smile Dx®

Cube Click, Inc.

Authorized purpose: Aids dentists reviewing bitewing and periapical radiographs of permanent teeth by detecting/segmenting caries and periapical radiolucencies, identifying restorations and dental anatomy, and measuring mesial/distal bone levels; it does not replace dentist review.

Modality
X-ray/Radiography
Anatomy
Dental/Maxillofacial
Decision
2025-05-14
FDA-cleared K233738

Overjet Caries Assist-Pediatric

Overjet, Inc

Authorized purpose: Aid dentists in detecting and segmenting caries on bitewing and periapical radiographs for patients aged 4-11 with primary or permanent teeth; the output is additional information and does not replace a complete dental review or clinical judgment.

Modality
X-ray/Radiography
Anatomy
Dental/Maxillofacial (carious tooth surfaces)
Decision
2024-03-04

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination

Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.

Accepted repository change history

No entry has been added to this scoped change history for this record. This is not proof that the record never changed.

Correction policy and limits · Public repository change history