Authorized purpose: Adjunctively assess breast tissue composition and provide an ACR BI-RADS 5th Edition breast-density category for asymptomatic women aged 40 years or older undergoing screening mammography; it is not diagnostic.
Clinical role
Concurrent adjunct to physician mammography interpretation, producing breast-density classification for review and reporting through compatible mammography software, PACS, worklists, or reporting systems.
Images and patients
Mammography/DBT · Breast
Key limitation
Keep in mind: The output is complementary information, not diagnostic software, and patient-management decisions must not be based solely on it. The primary validation data were Hologic examinations from two U.S. screening centers and one French center. Very small subgroup results, including the eight-patient American Indian, Alaska Native, Native Hawaiian, or Pacific Islander group, have wide uncertainty.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: Adjunctively assess breast tissue composition and provide an ACR BI-RADS 5th Edition breast-density category for asymptomatic women aged 40 years or older undergoing screening mammography; it is not diagnostic.
Which images or clinical context are in scope?Modality: Mammography/DBT Anatomy or clinical context: Breast Input: A DICOM image folder from compatible full-field digital mammography or digital breast tomosynthesis systems, including RCC, LCC, RMLO, and LMLO views; the comparison describes FFDM and 2D synthetic mammography from DBT.
Who uses it?Intended users: Interpreting physicians.
What does the software produce?Output: A four-category ACR BI-RADS 5th Edition breast-density result for each examination, delivered as JSON and capable of presentation through a compatible web UI, PACS viewer using DICOM Structured Report or Secondary Capture, patient worklists, or reporting software.
What is its most important limitation?Key limitation: The output is complementary information, not diagnostic software, and patient-management decisions must not be based solely on it. The primary validation data were Hologic examinations from two U.S. screening centers and one French center. Very small subgroup results, including the eight-patient American Indian, Alaska Native, Native Hawaiian, or Pacific Islander group, have wide uncertainty.
Clinical takeaway. Clinical role: Adjunctively assess breast tissue composition and provide an ACR BI-RADS 5th Edition breast-density category for asymptomatic women aged 40 years or older undergoing screening mammography; it is not diagnostic.
Evidence takeaway. Evidence from the reviewed source record: FDA standalone validation used 922 women and 4,274 examinations/images across two U.S. screening centers and one French center, with four views per examination and consensus ground truth from five breast radiologists. Quadratically weighted Cohen kappa was reported as 89.03 (95% CI 87.43-90.56) for the four-class task and 84.50 (95% CI 81.46-87.36) for the dense-versus-non-dense binary task. FFDM-only and 2DSM-only kappa values were 89.11 and 88.89; subgroup kappa values included 87.93, 91.22, and 87.15 by age and 53.85 (95% CI 0.00-83.33) for the eight-patient smallest race group.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
Adjunctively assess breast tissue composition and provide an ACR BI-RADS 5th Edition breast-density category for asymptomatic women aged 40 years or older undergoing screening mammography; it is not diagnostic.
Concurrent adjunct to physician mammography interpretation, producing breast-density classification for review and reporting through compatible mammography software, PACS, worklists, or reporting systems.
A DICOM image folder from compatible full-field digital mammography or digital breast tomosynthesis systems, including RCC, LCC, RMLO, and LMLO views; the comparison describes FFDM and 2D synthetic mammography from DBT.
A four-category ACR BI-RADS 5th Edition breast-density result for each examination, delivered as JSON and capable of presentation through a compatible web UI, PACS viewer using DICOM Structured Report or Secondary Capture, patient worklists, or reporting software.
The output is complementary information, not diagnostic software, and patient-management decisions must not be based solely on it. The primary validation data were Hologic examinations from two U.S. screening centers and one French center. Very small subgroup results, including the eight-patient American Indian, Alaska Native, Native Hawaiian, or Pacific Islander group, have wide uncertainty.
FDA standalone validation used 922 women and 4,274 examinations/images across two U.S. screening centers and one French center, with four views per examination and consensus ground truth from five breast radiologists. Quadratically weighted Cohen kappa was reported as 89.03 (95% CI 87.43-90.56) for the four-class task and 84.50 (95% CI 81.46-87.36) for the dense-versus-non-dense binary task. FFDM-only and 2DSM-only kappa values were 89.11 and 88.89; subgroup kappa values included 87.93, 91.22, and 87.15 by age and 53.85 (95% CI 0.00-83.33) for the eight-patient smallest race group.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
FDA standalone validation used 922 women and 4,274 examinations/images across two U.S. screening centers and one French center, with four views per examination and consensus ground truth from five breast radiologists. Quadratically weighted Cohen kappa was reported as 89.03 (95% CI 87.43-90.56) for the four-class task and 84.50 (95% CI 81.46-87.36) for the dense-versus-non-dense binary task. FFDM-only and 2DSM-only kappa values were 89.11 and 88.89; subgroup kappa values included 87.93, 91.22, and 87.15 by age and 53.85 (95% CI 0.00-83.33) for the eight-patient smallest race group.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
MammoScreen BD regulatory performance study summary
DesignStand-Alone Performance
PopulationAdult
Scopeexact submission
Tested versionExact product version not reported; regulatory study summarized for FDA submission K241561
Samplen=922
IndependenceIndependence not established
K241561
ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K241561. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: A DICOM image folder from compatible full-field digital mammography or digital breast tomosynthesis systems, including RCC, LCC, RMLO, and LMLO views; the comparison describes FFDM and 2D synthetic mammography from DBT.
Scope: Exact FDA submission.
Confirm scanner/model, protocol, series completeness and both software builds.
Source context, not route validation: DICOM folder input and JSON output with described integration through MammoScreen Web-UI, compatible PACS via DICOM Structured Report or Secondary Capture, patient worklists, and reporting software; the device is described as virtual-machine software. Specific local interface versions and compatibility are not established.
Scope: Exact FDA submission.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Source context, not route validation: The exact FDA package describes MammoScreen BD as virtual machine software in the deployment row of its predicate comparison table. Current hosting, data flow, and commercial configuration require vendor confirmation.
Scope: Exact FDA submission. Packet wording: "Deployment Virtual Machine Software Virtual Machine Software" … "Subject Device Predicate Device MammoScreen BD WRDensity by Whiterabbit.ai (K202013)". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
Confirm processing location, queue ownership, exclusions and exact AI build.
Source context, not route validation: A four-category ACR BI-RADS 5th Edition breast-density result for each examination, delivered as JSON and capable of presentation through a compatible web UI, PACS viewer using DICOM Structured Report or Secondary Capture, patient worklists, or reporting software.
Scope: Exact FDA submission.
Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.
Unresolved route: A four-category ACR BI-RADS 5th Edition breast-density result for each examination, delivered as JSON and capable of presentation through a compatible web UI, PACS viewer using DICOM Structured Report or Secondary Capture, patient worklists, or reporting software.
Scope: Exact FDA submission.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
The exact FDA package describes MammoScreen BD as virtual machine software in the deployment row of its predicate comparison table. Current hosting, data flow, and commercial configuration require vendor confirmation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "Deployment Virtual Machine Software Virtual Machine Software" … "Subject Device Predicate Device MammoScreen BD WRDensity by Whiterabbit.ai (K202013)". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
DICOM folder input and JSON output with described integration through MammoScreen Web-UI, compatible PACS via DICOM Structured Report or Secondary Capture, patient worklists, and reporting software; the device is described as virtual-machine software. Specific local interface versions and compatibility are not established.
Therapixel's own page says Therapixel and its operations for MammoScreen are ISO/IEC 27001 certified by an independent certification body. The page gives no certificate number, scope statement, issue date or expiry, and the certificate was not read.
Vendor-wide
Checked 2026-10-25
Vendor-wide statement, not an independent attestation. It does not establish certificate or report scope for this exact release, expiry, data location, retention, identity and access, or incident-response terms for a deployed configuration. Obtain these from the vendor.
ACR AI Central provides an exact-submission model card for MammoScreen BD; FDA labeling remains controlling for clinical use, limitations, and performance.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Exact FDA record reviewedStatus
2026-09-20Last searched
22Fields reviewed
10Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, acr ai central product, vendor documentation, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not verified - checked 2026-09-08 - Exact FDA submission
No public price or quote for the exact K241561 configuration was established in this review.
Pricing model: Not yet reviewed
Setup and integration fees: Not yet reviewed
Recurring fees: Not yet reviewed
Minimum commitment: Not yet reviewed
Contract duration: Not yet reviewed
Renewal terms: Not yet reviewed
Cancellation terms: Not yet reviewed
Support and service terms: Not yet reviewed
Trial or pilot terms: Not yet reviewed
Reimbursement: Not publicly available - checked 2026-09-08 - Exact FDA submission
No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.
Data hosting location: Not yet reviewed
Data retention: Not yet reviewed
Documented data use: Not yet reviewed
Hardware requirements: Not yet reviewed
Network requirements: Not yet reviewed
Identity and access: Not yet reviewed
Installation timeline: Not yet reviewed
Training: Not verified - checked 2026-09-08 - Exact FDA submission
Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.
Monitoring: Not verified - checked 2026-09-08 - Exact FDA submission
Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.
Authorized purpose: Flags calcifications, masses, asymmetries and architectural distortions on compatible FFDM or DBT screening examinations; management must not be based solely on MammoScreen output.
Authorized purpose: Tags screening cases as suspicious or blank to aid worklist prioritization; radiologists must review every examination and it does not mark original images.
Authorized purpose: Lunit INSIGHT MMG v1.1.10 aids physicians in detection, localization and characterization of suspicious breast-cancer areas on female screening mammograms from compatible FFDM systems, viewed after the physician's initial read as an adjunct to complete clinical interpretation.
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.
Accepted repository change history
No entry has been added to this scoped change history for this record. This is not proof that the record never changed.