Authorized purpose: Concurrent reading and reporting aid for physicians interpreting screening mammograms on compatible full-field digital mammography and digital breast tomosynthesis systems, optionally using compatible prior examinations; outputs are adjunctive and must not be the sole basis for patient-management decisions.
Clinical role
Concurrent reading and reporting support that supplies finding marks, suspicion scores, lesion type and location, reporting-system content, and optional patient-management-worklist content while the interpreting physician remains the sole decision-maker.
Images and patients
Mammography/DBT · Breast
Key limitation
Keep in mind: Use requires compatible FFDM/DBT systems and only the packet's supported current/prior combinations. Outputs are complementary information, not a replacement for physician review or judgment, and patient-management decisions must not rely on the analysis alone. Standalone testing used 100% Hologic exams from three US centers, so that cohort does not establish all vendors or local configurations.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: Concurrent reading and reporting aid for physicians interpreting screening mammograms on compatible full-field digital mammography and digital breast tomosynthesis systems, optionally using compatible prior examinations; outputs are adjunctive and must not be the sole basis for patient-management decisions.
Which images or clinical context are in scope?Modality: Mammography/DBT Anatomy or clinical context: Breast Input: Compatible current FFDM, DBT+FFDM, or DBT+2DSM examinations, with optional compatible prior FFDM or DBT+2DSM examination acquired 6 to 60 months before the current study; the packet specifies supported current/prior combinations.
Who uses it?Intended users: Physicians qualified to read mammograms; the FDA reader study used MQSA-qualified and ABR-certified radiologists.
What does the software produce?Output: Graphical marks for soft-tissue lesions or calcifications, a MammoScreen Score from 1 to 10 at finding, breast, and case levels, lesion characterization as mass/asymmetry, distortion, or calcifications, and location including quadrant, depth, and distance from the nipple; outputs can support DICOM viewing, preliminary reporting, and worklist content.
What is its most important limitation?Key limitation: Use requires compatible FFDM/DBT systems and only the packet's supported current/prior combinations. Outputs are complementary information, not a replacement for physician review or judgment, and patient-management decisions must not rely on the analysis alone. Standalone testing used 100% Hologic exams from three US centers, so that cohort does not establish all vendors or local configurations.
Clinical takeaway. Clinical role: Concurrent reading and reporting aid for physicians interpreting screening mammograms on compatible full-field digital mammography and digital breast tomosynthesis systems, optionally using compatible prior examinations; outputs are adjunctive and must not be the sole basis for patient-management decisions.
Evidence takeaway. Evidence from the reviewed source record: FDA testing included a fully crossed multi-reader multi-case study of 240 combined DBT/2D mammograms with a prior, read with and without MammoScreen 3 by 23 MQSA-qualified and ABR-certified radiologists. Reader average AUC increased from 0.797 unaided to 0.871 aided; sensitivity increased from 0.706 to 0.793 and specificity from 0.815 to 0.836. Standalone mammogram-level performance used 7,544 exams from 4,429 patients at three US centers, with 1,053 positive and 6,420 negative cases, and reported AUROC 0.927 (95% CI 0.911-0.942), sensitivity 0.912 (95% CI 0.895-0.928), and specificity 0.727 (95% CI 0.715-0.740). Reader-study and standalone endpoints are distinct.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
Concurrent reading and reporting aid for physicians interpreting screening mammograms on compatible full-field digital mammography and digital breast tomosynthesis systems, optionally using compatible prior examinations; outputs are adjunctive and must not be the sole basis for patient-management decisions.
Concurrent reading and reporting support that supplies finding marks, suspicion scores, lesion type and location, reporting-system content, and optional patient-management-worklist content while the interpreting physician remains the sole decision-maker.
Compatible current FFDM, DBT+FFDM, or DBT+2DSM examinations, with optional compatible prior FFDM or DBT+2DSM examination acquired 6 to 60 months before the current study; the packet specifies supported current/prior combinations.
Graphical marks for soft-tissue lesions or calcifications, a MammoScreen Score from 1 to 10 at finding, breast, and case levels, lesion characterization as mass/asymmetry, distortion, or calcifications, and location including quadrant, depth, and distance from the nipple; outputs can support DICOM viewing, preliminary reporting, and worklist content.
Use requires compatible FFDM/DBT systems and only the packet's supported current/prior combinations. Outputs are complementary information, not a replacement for physician review or judgment, and patient-management decisions must not rely on the analysis alone. Standalone testing used 100% Hologic exams from three US centers, so that cohort does not establish all vendors or local configurations.
FDA testing included a fully crossed multi-reader multi-case study of 240 combined DBT/2D mammograms with a prior, read with and without MammoScreen 3 by 23 MQSA-qualified and ABR-certified radiologists. Reader average AUC increased from 0.797 unaided to 0.871 aided; sensitivity increased from 0.706 to 0.793 and specificity from 0.815 to 0.836. Standalone mammogram-level performance used 7,544 exams from 4,429 patients at three US centers, with 1,053 positive and 6,420 negative cases, and reported AUROC 0.927 (95% CI 0.911-0.942), sensitivity 0.912 (95% CI 0.895-0.928), and specificity 0.727 (95% CI 0.715-0.740). Reader-study and standalone endpoints are distinct.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
FDA testing included a fully crossed multi-reader multi-case study of 240 combined DBT/2D mammograms with a prior, read with and without MammoScreen 3 by 23 MQSA-qualified and ABR-certified radiologists. Reader average AUC increased from 0.797 unaided to 0.871 aided; sensitivity increased from 0.706 to 0.793 and specificity from 0.815 to 0.836. Standalone mammogram-level performance used 7,544 exams from 4,429 patients at three US centers, with 1,053 positive and 6,420 negative cases, and reported AUROC 0.927 (95% CI 0.911-0.942), sensitivity 0.912 (95% CI 0.895-0.928), and specificity 0.727 (95% CI 0.715-0.740). Reader-study and standalone endpoints are distinct.
0.912 (95% CI: 0.895-0.928) at the overall standalone mammogram-level endpoint
FDA source
Exact FDA submission
Checked 2026-09-09
The FDA table supports this point estimate for 1,053 positive and 6,420 negative standalone cases; reader-assisted and standalone sensitivity endpoints must remain separate.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
MammoScreen® (3) regulatory performance study summary
DesignStand-alone performance and reader study (ACR AI Central category)
PopulationAdult
Scopeexact submission
Tested version3
Samplen=7544
Sites3
IndependenceIndependence not established
K240301
ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K240301. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: Compatible current FFDM, DBT+FFDM, or DBT+2DSM examinations, with optional compatible prior FFDM or DBT+2DSM examination acquired 6 to 60 months before the current study; the packet specifies supported current/prior combinations.
Scope: Exact FDA submission.
Confirm scanner/model, protocol, series completeness and both software builds.
Source context, not route validation: Results are accessible through a dedicated interface, can integrate into DICOM viewers using DICOM-SC and DICOM-SR, can populate a preliminary reporting-system report, and can place case scores in a patient-management worklist; specific PACS, RIS, EHR, versions, and local compatibility are not established.
Scope: Exact FDA submission.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Unresolved route: Not established for the exact route.
Scope: Exact FDA submission. The exact FDA decision document was reviewed on 2026-11-14 and does not state where the subject device's software runs or how it is hosted. Confirm the deployment model, hosting, data flow, and current commercial configuration with the vendor.
Confirm processing location, queue ownership, exclusions and exact AI build.
Field checked 2026-11-14 ·
Returned objects
Source context, not route validation: Graphical marks for soft-tissue lesions or calcifications, a MammoScreen Score from 1 to 10 at finding, breast, and case levels, lesion characterization as mass/asymmetry, distortion, or calcifications, and location including quadrant, depth, and distance from the nipple; outputs can support DICOM viewing, preliminary reporting, and worklist content.
Scope: Exact FDA submission.
Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.
Unresolved route: Graphical marks for soft-tissue lesions or calcifications, a MammoScreen Score from 1 to 10 at finding, breast, and case levels, lesion characterization as mass/asymmetry, distortion, or calcifications, and location including quadrant, depth, and distance from the nipple; outputs can support DICOM viewing, preliminary reporting, and worklist content.
Scope: Exact FDA submission.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
Results are accessible through a dedicated interface, can integrate into DICOM viewers using DICOM-SC and DICOM-SR, can populate a preliminary reporting-system report, and can place case scores in a patient-management worklist; specific PACS, RIS, EHR, versions, and local compatibility are not established.
Therapixel's own page says Therapixel and its operations for MammoScreen are ISO/IEC 27001 certified by an independent certification body. The page gives no certificate number, scope statement, issue date or expiry, and the certificate was not read.
Vendor-wide
Checked 2026-10-25
Vendor-wide statement, not an independent attestation. It does not establish certificate or report scope for this exact release, expiry, data location, retention, identity and access, or incident-response terms for a deployed configuration. Obtain these from the vendor.
ACR AI Central provides an exact-submission model card for MammoScreen® (3); FDA labeling remains controlling for clinical use, limitations, and performance.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Exact FDA record reviewedStatus
2026-09-20Last searched
25Fields reviewed
10Source classes checked
1Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains 1 PubMed and 0 ClinicalTrials.gov candidate record. Discovery metadata alone does not establish exact tested-product, version, population, and endpoint applicability, so all candidates remain quarantined and none was promoted. The current discovery snapshot contains 1 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, acr ai central product, vendor documentation, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not verified - checked 2026-09-09 - Exact FDA submission
No public price or quote for the exact K240301 configuration was established in this review.
Pricing model: Not yet reviewed
Setup and integration fees: Not yet reviewed
Recurring fees: Not yet reviewed
Minimum commitment: Not yet reviewed
Contract duration: Not yet reviewed
Renewal terms: Not yet reviewed
Cancellation terms: Not yet reviewed
Support and service terms: Not yet reviewed
Trial or pilot terms: Not yet reviewed
Reimbursement: Not publicly available - checked 2026-09-09 - Exact FDA submission
No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.
Deployment: Not publicly available - checked 2026-11-14 - Exact FDA submission
The exact FDA decision document was reviewed on 2026-11-14 and does not state where the subject device's software runs or how it is hosted. Confirm the deployment model, hosting, data flow, and current commercial configuration with the vendor.
Data hosting location: Not yet reviewed
Data retention: Not yet reviewed
Documented data use: Not yet reviewed
Hardware requirements: Not yet reviewed
Network requirements: Not yet reviewed
Identity and access: Not yet reviewed
Installation timeline: Not yet reviewed
Training: Not verified - checked 2026-09-09 - Exact FDA submission
Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.
Monitoring: Not verified - checked 2026-09-09 - Exact FDA submission
Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.
Authorized purpose: Flags calcifications, masses, asymmetries and architectural distortions on compatible FFDM or DBT screening examinations; management must not be based solely on MammoScreen output.
Authorized purpose: Tags screening cases as suspicious or blank to aid worklist prioritization; radiologists must review every examination and it does not mark original images.
Authorized purpose: Lunit INSIGHT MMG v1.1.10 aids physicians in detection, localization and characterization of suspicious breast-cancer areas on female screening mammograms from compatible FFDM systems, viewed after the physician's initial read as an adjunct to complete clinical interpretation.
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.
Accepted repository change history
No entry has been added to this scoped change history for this record. This is not proof that the record never changed.