Authorized purpose: MAGNETOM Free.Max and MAGNETOM Free.Star produce diagnostic MR images in the labeled scanner, coil, sequence, anatomy, and professional-interpretation workflow.
Clinical role
MRI acquisition, reconstruction, display, and release-specific workflow support.
Images and patients
MRI · Multi-region
Key limitation
Keep in mind: Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release. MAGNETOM Free.Max and MAGNETOM Free.Star produce diagnostic MR images in the labeled scanner, coil, sequence, anatomy, and professional-interpretation workflow.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: MAGNETOM Free.Max and MAGNETOM Free.Star produce diagnostic MR images in the labeled scanner, coil, sequence, anatomy, and professional-interpretation workflow.
Which images or clinical context are in scope?Modality: MRI Anatomy or clinical context: Multi-region Input: Patient MRI signal data acquired with supported coils, pulse sequences, field strengths, and release-specific protocols.
Who uses it?Intended users: Qualified MRI technologists and physicians trained in magnetic resonance imaging.
What does the software produce?Output: Cross-sectional MR images, spectra, derived measurements, and reconstructed series for professional review.
What is its most important limitation?Key limitation: Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release. MAGNETOM Free.Max and MAGNETOM Free.Star produce diagnostic MR images in the labeled scanner, coil, sequence, anatomy, and professional-interpretation workflow.
Clinical takeaway. Clinical role: MAGNETOM Free.Max and MAGNETOM Free.Star produce diagnostic MR images in the labeled scanner, coil, sequence, anatomy, and professional-interpretation workflow.
Evidence takeaway. Evidence from the reviewed source record: The exact FDA document reports non-clinical performance testing on the subject devices: sample clinical images, performance bench testing including heating measurements for coils, and software verification and validation of mainly new and modified features. It states no clinical tests were conducted to support substantial equivalence, although sample clinical images were provided. Results are as reported in the FDA submission, not independent validation.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
MAGNETOM Free.Max and MAGNETOM Free.Star produce diagnostic MR images in the labeled scanner, coil, sequence, anatomy, and professional-interpretation workflow.
Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release. MAGNETOM Free.Max and MAGNETOM Free.Star produce diagnostic MR images in the labeled scanner, coil, sequence, anatomy, and professional-interpretation workflow.
The exact FDA document reports non-clinical performance testing on the subject devices: sample clinical images, performance bench testing including heating measurements for coils, and software verification and validation of mainly new and modified features. It states no clinical tests were conducted to support substantial equivalence, although sample clinical images were provided. Results are as reported in the FDA submission, not independent validation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "The following performance testing was conducted on the subject devices." … "Performance bench test - Heating measurements for coils". Read from the exact FDA decision document on 2026-11-14.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
The exact FDA document reports non-clinical performance testing on the subject devices: sample clinical images, performance bench testing including heating measurements for coils, and software verification and validation of mainly new and modified features. It states no clinical tests were conducted to support substantial equivalence, although sample clinical images were provided. Results are as reported in the FDA submission, not independent validation.
FDA source
Exact FDA submission
Checked 2026-11-14
Packet wording: "The following performance testing was conducted on the subject devices." … "Performance bench test - Heating measurements for coils". Read from the exact FDA decision document on 2026-11-14.
No exact submission and version-linked sensitivity endpoint is published in this dossier; do not substitute a product-family or candidate-study value.
Reported specificity
Not established in reviewed sources
Reviewed sources checked
Exact FDA submission
Checked 2026-09-04
No exact submission and version-linked specificity endpoint is published in this dossier; do not substitute a product-family or candidate-study value.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
FDA submission-specific verification and validation evidence for K220575
DesignRegulatory verification, validation, and labeling evidence reviewed for exact FDA submission K220575.
PopulationThe labeled population and test materials for exact submission K220575; a general clinical study population is not established here.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
FDA recall Z-2648-2023 is associated with native exact submission identifier K220575. Status: Open, Classified. Product: (1) Over-ear headphones used with the MAGNETOM Amira MAGNETOM Free.Star MAGNETOM Sempra, MAGNETOM Free.Max, MAGNETOM Mica MAGNETOM Spectra Siemens Model Number:11060845 Used with the following MAGNETOM MRI Systems: MAGENTOM Amira.... Reason: In some cases, when lubricating grease is present within the headband of the over-ear headphones, it may cause visible, dot- or lineshaped, fat-isointense artifacts during head examinations only. These potential artifacts depend on the position of the headband, the amount of lubricating grease, as well as the sequence parameters used
FDA source
Exact FDA submission
Checked 2026-09-04
Each published event was manually matched to the exact FDA submission and affected product or version; a recall does not establish an AI-algorithm defect or current site exposure unless the event states that scope.
Native exact-submission identifier match from the openFDA candidate snapshot accessed 2026-09-04 with source snapshot date 2026-09-02. Source URL: https://api.fda.gov/device/recall.json?search=product_res_number.exact%3A%22Z-2648-2023%22&limit=1. This does not establish exact installed serial, lot, unit, or software-version applicability.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: Patient MRI signal data acquired with supported coils, pulse sequences, field strengths, and release-specific protocols.
Scope: Exact FDA submission.
Confirm scanner/model, protocol, series completeness and both software builds.
Source context, not route validation: The exact FDA package describes Access-i, an interface enabling connection of a third-party workstation to the MR syngo Acquisition Workplace via a network router and secure local network connection, and lists conformance to the NEMA DICOM standard (PS 3.1–3.20). Interface versions and local compatibility require confirmation.
Scope: Exact FDA submission. Packet wording: "Access-i: Provides an interface to enable the connection of a 3rd party workstation to the MR syngo Acquisition Workplace via a network router and secure local network connection." … "MAGNETOM Free.Max and MAGNETOM Free.Star with software syngo MR XA50A conform to the following FDA recognized and international IEC, ISO and NEMA standards:". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Source context, not route validation: The exact FDA package describes MAGNETOM Free.Max and MAGNETOM Free.Star as two 0.55 Tesla MRI systems supported by the subject device software version syngo MR XA50A. Current hosting, data flow, and commercial configuration require vendor confirmation.
Scope: Exact FDA submission. Packet wording: "The subject device software version, syngo MR XA50A, can support the following two MRI systems:" … "we extend the Free. platform, which consists of two products with a field strength of 0.55 Tesla on our high-value MRI platform.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
Confirm processing location, queue ownership, exclusions and exact AI build.
Unresolved route: Cross-sectional MR images, spectra, derived measurements, and reconstructed series for professional review.
Scope: Exact FDA submission.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
The exact FDA package describes MAGNETOM Free.Max and MAGNETOM Free.Star as two 0.55 Tesla MRI systems supported by the subject device software version syngo MR XA50A. Current hosting, data flow, and commercial configuration require vendor confirmation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "The subject device software version, syngo MR XA50A, can support the following two MRI systems:" … "we extend the Free. platform, which consists of two products with a field strength of 0.55 Tesla on our high-value MRI platform.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
The exact FDA package describes Access-i, an interface enabling connection of a third-party workstation to the MR syngo Acquisition Workplace via a network router and secure local network connection, and lists conformance to the NEMA DICOM standard (PS 3.1–3.20). Interface versions and local compatibility require confirmation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "Access-i: Provides an interface to enable the connection of a 3rd party workstation to the MR syngo Acquisition Workplace via a network router and secure local network connection." … "MAGNETOM Free.Max and MAGNETOM Free.Star with software syngo MR XA50A conform to the following FDA recognized and international IEC, ISO and NEMA standards:". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.
Siemens Healthineers' corporate cybersecurity page states a Secure Development Lifecycle addressing pre-market and post-market phases for products under development and a range of existing offerings; a Cybersecurity Management Service maintaining a product security inventory (over 1,000 product versions and over 21,000 monitored software components) with security whitepapers, SBOMs and vulnerability assessments available to customers via teamplay Fleet; incident and vulnerability management channels; and that Siemens Healthineers AG maintains an independent certification of its global Cybersecurity Management System to ISO/IEC 27001:2022 extended by ISO/IEC 27701:2019, with the certificate referenced to the SGS directory. Vendor statement; no certificate or report was read.
Vendor-wide
Checked 2026-10-26
Vendor statement, not an attestation. Identity bridge: Siemens Healthineers' own corporate-info page names Siemens Medical Solutions USA, Inc. (Malvern, PA), and its 2024 governance press release states Siemens Healthcare GmbH's business was transferred to Siemens Healthineers AG. Corporate program statement; exact-release scope is not established.
Siemens Healthineers' cybersecurity page states that its Cybersecurity Management Service maintains a product security inventory of over 1,000 product versions and over 21,000 monitored software components, with security whitepapers, SBOMs and vulnerability assessments delivered via teamplay Fleet, and that it operates incident and vulnerability management.
Vendor-wide
Checked 2026-10-27
Vendor statement, not an attestation. The page names Siemens Healthineers generally and does not scope these commitments to any specific product release; exact-release scope is not established.
Current vendor material is product-family context for K220575 only; exact release, enabled options, deployment, and local interoperability require confirmation.
product family · checked 2026-09-04
Vendor material is not exact-release evidence unless that linkage is separately proven.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Human reviewedStatus
2026-09-04Last searched
29Fields reviewed
13Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. 1 FDA recall candidate was manually reviewed: 1 published and 0 rejected as a neighboring or otherwise inapplicable version. Native FDA recall identifiers produced 1 postmarket candidate record; 1 has been reviewed (1 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, vendor product page, implementation documentation, reimbursement policy, peer reviewed literature, clinical trial registry, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not publicly available - checked 2026-09-04 - Exact FDA submission
No public exact-release price, license fee, subscription, contract, or implementation cost was established in this review.
Pricing model: Not yet reviewed
Setup and integration fees: Not yet reviewed
Recurring fees: Not yet reviewed
Minimum commitment: Not yet reviewed
Contract duration: Not yet reviewed
Renewal terms: Not yet reviewed
Cancellation terms: Not yet reviewed
Support and service terms: Not yet reviewed
Trial or pilot terms: Not yet reviewed
Reimbursement: Not applicable - checked 2026-09-04 - Not applicable
CMS index context was checked, but no product-specific coverage or reimbursement determination is inferred.
Data hosting location: Not yet reviewed
Data retention: Not yet reviewed
Documented data use: Not yet reviewed
Hardware requirements: Not yet reviewed
Network requirements: Not yet reviewed
Identity and access: Not yet reviewed
Installation timeline: Not yet reviewed
Training: Not verified - checked 2026-10-27 - Exact FDA submission
Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.
Authorized purpose: Vantage Fortian and Orian 1.5T produce diagnostic MR images of the head and body, including labeled noncontrast angiography, for trained technologists and physician interpretation.
Authorized purpose: S-scan Open is an Esaote MR system for diagnostic image acquisition, reconstruction, display, and professional interpretation within the labeled scanner, coil, sequence, and anatomy scope.
Authorized purpose: AIR Recon DL adds a deep-learning phase-correction and reconstruction option to supported GE 1.5T, 3T, and 7T MR workflows for all labeled anatomies and ages to reduce noise or ringing and support image quality or scan-time choices.
Authorized purpose: The listed MAGNETOM 1.5T and 3T MRI systems produce cross-sectional images, spectroscopic images or spectra, and derived physical parameters of the head, body, and extremities for trained-physician interpretation and can support imaging during MR-compatible interventions.