Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems

Philips Medical Systems Nederland B.V. · MRI · FDA-cleared (510(k))

Authorized purpose: The Philips Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300, and MR 7700 systems provide head, body, and extremity MRI with SmartSpeed AI reconstruction or acceleration in supported configurations.

Clinical role
MRI acquisition, reconstruction, review, and release-specific workflow support.
Images and patients
MRI · Multi-region
Key limitation
Keep in mind: Accuracy of physical parameters depends on system and scan parameters and must be controlled and validated by the clinical user; MR interventional procedures require MR Conditional or MR Safe instrumentation selected for the specific hospital configuration.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: The Philips Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300, and MR 7700 systems provide head, body, and extremity MRI with SmartSpeed AI reconstruction or acceleration in supported configurations.
  • Which images or clinical context are in scope?Modality: MRI Anatomy or clinical context: Multi-region Input: Patient MRI signal data acquired with supported coils, pulse sequences, field strengths, and release-specific protocols.
  • Who uses it?Intended users: Trained physicians and trained clinical users; interpretation of images is the responsibility of trained physicians.
  • What does the software produce?Output: Cross-sectional MR images, spectra, derived measurements, and reconstructed series for professional review.
  • What is its most important limitation?Key limitation: Accuracy of physical parameters depends on system and scan parameters and must be controlled and validated by the clinical user; MR interventional procedures require MR Conditional or MR Safe instrumentation selected for the specific hospital configuration.

Clinical takeaway. Clinical role: The Philips Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300, and MR 7700 systems provide head, body, and extremity MRI with SmartSpeed AI reconstruction or acceleration in supported configurations.

Evidence takeaway. Evidence from the reviewed source record: The exact FDA document reports non-clinical verification and validation tests complying with international and FDA-recognized consensus standards, and simulated use validation of Precise image by qualitative review performed by application specialists. A reader evaluation by ABR board certified radiologists analyzed SNR, artifact level, sharpness and CNR; the proposed device was assessed as equivalent for diagnosis with significantly better SNR and sharpness than the predicate reconstruction technology. No technological characteristics would require clinical testing. Results are as reported in the FDA submission, not independent validation.

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
The Philips Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300, and MR 7700 systems provide head, body, and extremity MRI with SmartSpeed AI reconstruction or acceleration in supported configurations.
  • Exact FDA submission
  • Checked 2026-09-19

Clinical-use display mirrors the exact FDA-supported purpose reconciled from the decision-summary packet.

Workflow role
MRI acquisition, reconstruction, review, and release-specific workflow support.
  • Exact FDA submission
  • Checked 2026-09-03
Required input
Patient MRI signal data acquired with supported coils, pulse sequences, field strengths, and release-specific protocols.
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Output and clinician action
Cross-sectional MR images, spectra, derived measurements, and reconstructed series for professional review.
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Care setting
Hospital and outpatient MRI departments operating the cleared field strength, coils, protocols, and accessories.
  • Exact FDA submission
  • Checked 2026-09-03
Intended users
Trained physicians and trained clinical users; interpretation of images is the responsibility of trained physicians.
  • Exact FDA submission
  • Checked 2026-09-19

The exact indication and comparison summary establish trained physicians and clinical users, not MRI technologists as a named class.

Limitations
Accuracy of physical parameters depends on system and scan parameters and must be controlled and validated by the clinical user; MR interventional procedures require MR Conditional or MR Safe instrumentation selected for the specific hospital configuration.
  • Exact FDA submission
  • Checked 2026-09-19

This uses the exact configuration and validation boundaries stated in the indication.

Evidence summary
The exact FDA document reports non-clinical verification and validation tests complying with international and FDA-recognized consensus standards, and simulated use validation of Precise image by qualitative review performed by application specialists. A reader evaluation by ABR board certified radiologists analyzed SNR, artifact level, sharpness and CNR; the proposed device was assessed as equivalent for diagnosis with significantly better SNR and sharpness than the predicate reconstruction technology. No technological characteristics would require clinical testing. Results are as reported in the FDA submission, not independent validation.
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "substantial equivalence is demonstrated with non-clinical performance (verification and validation) tests, which complied with the requirements specified in the international and FDA-recognized consensus standards and device-specific guidance." … "Non-Clinical verification and validation tests have been performed with regards to the intended use, the technical claims, the requirement specifications and the risk management results.". Read from the exact FDA decision document on 2026-11-14.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
The exact FDA document reports non-clinical verification and validation tests complying with international and FDA-recognized consensus standards, and simulated use validation of Precise image by qualitative review performed by application specialists. A reader evaluation by ABR board certified radiologists analyzed SNR, artifact level, sharpness and CNR; the proposed device was assessed as equivalent for diagnosis with significantly better SNR and sharpness than the predicate reconstruction technology. No technological characteristics would require clinical testing. Results are as reported in the FDA submission, not independent validation.
  • FDA source
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "substantial equivalence is demonstrated with non-clinical performance (verification and validation) tests, which complied with the requirements specified in the international and FDA-recognized consensus standards and device-specific guidance." … "Non-Clinical verification and validation tests have been performed with regards to the intended use, the technical claims, the requirement specifications and the risk management results.". Read from the exact FDA decision document on 2026-11-14.

Reported sensitivity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-03

Sensitivity is not the primary endpoint for this system, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

Reported specificity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-03

Specificity is not the primary endpoint for this system, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

K223458 FDA performance package for Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems

  • DesignManufacturer MR-system, software, image-quality, and reader validation
  • PopulationSupported head, body, extremity, and release-specific accelerated MRI applications
  • Scopeexact submission
  • Tested versionIngenia/Ingenia CX/Ingenia Elition/Ingenia Ambition/MR 5300/MR 7700 MR Systems
  • IndependenceSponsor or vendor study

K223458

Manufacturer evidence submitted for FDA clearance or approval; local clinical, technical, and workflow acceptance remains necessary.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
8 exact-submission FDA recall records in the 2026-09-02 openFDA snapshot identify K223458. Affected model, version, unit, correction, and current manufacturer instructions still require local confirmation. FDA recall Z-1954-2026 is Open, Classified for Philips Evolution Upgrade 1.5T with MR Elastography (MRE). 1. Model Number (REF): 782116 2. Model Number (REF): 782148. 3. Model Number (REF): 782166.. FDA reason: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small. FDA action: Philips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the provided response form. FDA recall Z-1955-2026 is Open, Classified for Philips Evolution upgrade 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782143. 2. Model Number (REF): 782162.. FDA reason: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small. FDA action: Philips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the provided response form. FDA recall Z-1958-2026 is Open, Classified for Philips Ingenia 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781342. 2. Model Number (REF): 781377. 3. Model Number (REF): 782103. 510(k) Numbers: K163116, K173079, K183063, K193215, K213516, K213583, K223458, K251397, K251808. FDA reason: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small. FDA action: Philips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the provided response form. FDA recall Z-1961-2026 is Open, Classified for Philips Ingenia Ambition X with MR Elastography (MRE). 1. Model Number (REF): 781356. 2. Model Number (REF): 782109. 3. Model Number (REF): 782138. 4. Model Number (REF): 782160.. FDA reason: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small. FDA action: Philips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the provided response form. FDA recall Z-1962-2026 is Open, Classified for Philips Ingenia Elition S with MR Elastography (MRE). 1. Model Number (REF): 781357. 2. Model Number (REF): 782106. 3. Model Number (REF): 782137.. FDA reason: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small. FDA action: Philips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the provided response form. FDA recall Z-1964-2026 is Open, Classified for Philips MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782120. 2. Model Number (REF): 782153.. FDA reason: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small. FDA action: Philips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the provided response form. FDA recall Z-1967-2026 is Open, Classified for Philips SmartPath to dStream for XR and 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781270. 2. Model Number (REF): 782113.. FDA reason: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small. FDA action: Philips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the provided response form. FDA recall Z-1969-2026 is Open, Classified for Philips Upgrade to MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782130.. FDA reason: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small. FDA action: Philips Medical notified consignees on about 04/14/2026 via letter. Consignees were made aware of the issue, affected units, and instructed to continue to use MDE as per established guidelines, manually update the scan protocol settings according to the directions provided to prevent the issue from occurring, and continue using the systems in accordance with their intended use. Consignees were also instructed to circulate the notice to all users for awareness, retain the Urgent Medical Device letter with affected systems until the software upgrade is installed and complete and return the provided response form.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04

Each published event was manually matched to the exact FDA submission and affected product or version; a recall does not establish an AI-algorithm defect or current site exposure unless the event states that scope.

recall: Z-1954-2026

Open, Classified

2026-04-27

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; do not infer an AI-algorithm defect unless the FDA record says so. Confirm affected model, version, units, corrective action, and completion status locally.

recall: Z-1955-2026

Open, Classified

2026-04-27

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; do not infer an AI-algorithm defect unless the FDA record says so. Confirm affected model, version, units, corrective action, and completion status locally.

recall: Z-1958-2026

Open, Classified

2026-04-27

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; do not infer an AI-algorithm defect unless the FDA record says so. Confirm affected model, version, units, corrective action, and completion status locally.

recall: Z-1961-2026

Open, Classified

2026-04-27

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; do not infer an AI-algorithm defect unless the FDA record says so. Confirm affected model, version, units, corrective action, and completion status locally.

recall: Z-1962-2026

Open, Classified

2026-04-27

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; do not infer an AI-algorithm defect unless the FDA record says so. Confirm affected model, version, units, corrective action, and completion status locally.

recall: Z-1964-2026

Open, Classified

2026-04-27

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; do not infer an AI-algorithm defect unless the FDA record says so. Confirm affected model, version, units, corrective action, and completion status locally.

recall: Z-1967-2026

Open, Classified

2026-04-27

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; do not infer an AI-algorithm defect unless the FDA record says so. Confirm affected model, version, units, corrective action, and completion status locally.

recall: Z-1969-2026

Open, Classified

2026-04-27

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; do not infer an AI-algorithm defect unless the FDA record says so. Confirm affected model, version, units, corrective action, and completion status locally.

Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not publicly available: Reviewed sources do not publish an exact-product procurement price with a stated offer scope.

Reimbursement

Reimbursement
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging, procedural, or surgical service line and verify payer, site-of-service, and coding assumptions independently.
  • Service-level, not product-specific
  • Checked 2026-09-03

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

Build a local total-cost and value model; do not infer reimbursement from FDA clearance or approval.

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: Patient MRI signal data acquired with supported coils, pulse sequences, field strengths, and release-specific protocols.

    Scope: Exact FDA submission. Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-09-19 · FDA K223458: Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems

  2. Routing

    Source context, not route validation: Scanner-console and DICOM workflow to PACS or downstream applications; validate worklist, identifiers, series labels, geometry, measurements, routing, and archive completeness.

    Scope: Product family. Current vendor materials are product-family evidence for K223458; confirm the exact cleared version, enabled options, deployment, and local compatibility.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-09-03 · FDA K223458: Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems · Philips SmartSpeed AI

  3. AI processing

    Source context, not route validation: Installed MRI hardware and release-specific scanner software with licensed coils, sequences, reconstruction options, and accessories.

    Scope: Product family. Current vendor materials are product-family evidence for K223458; confirm the exact cleared version, enabled options, deployment, and local compatibility.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-09-03 · FDA K223458: Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems · Philips SmartSpeed AI

  4. Returned objects

    Source context, not route validation: Cross-sectional MR images, spectra, derived measurements, and reconstructed series for professional review.

    Scope: Exact FDA submission. Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-19 · FDA K223458: Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems

  5. Viewer/worklist/report

    Unresolved route: Cross-sectional MR images, spectra, derived measurements, and reconstructed series for professional review.

    Scope: Exact FDA submission. Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-19 · FDA K223458: Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Deployment
Installed MRI hardware and release-specific scanner software with licensed coils, sequences, reconstruction options, and accessories.
  • Product family
  • Checked 2026-09-03

Current vendor materials are product-family evidence for K223458; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Integration
Scanner-console and DICOM workflow to PACS or downstream applications; validate worklist, identifiers, series labels, geometry, measurements, routing, and archive completeness.
  • Product family
  • Checked 2026-09-03

Current vendor materials are product-family evidence for K223458; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Security
Philips' Responsible Disclosure Statement states that Philips maintains a global network of product security officers for developing and deploying security and privacy features for its products and services and for managing security events; that vulnerability reports to productsecurity@philips.com (PGP-encrypted) are acknowledged within two business days with a tracking number and an assigned contact; status updates approximately every 2-3 weeks; a 90-day resolution aim; and patch or advisory notification through existing customer notification processes. Vendor statement; no certification or test report was read.
  • Vendor-wide
  • Checked 2026-10-26

Vendor statement, not an attestation. Identity bridge: the statement is on philips.com (Royal Philips) and covers Philips products and services; the FDA applicant entity appears in Philips' own BCR members list on documents.philips.com (Philips Ultrasound LLC, Philips Medical Systems Nederland B.V., Philips Healthcare (Suzhou), Philips Medical Systems DMC GmbH, Philips Medical Systems Technologies Ltd., Philips Oy). Corporate policy; exact-release scope is not established.

Training
Only qualified mri technologists and physicians trained in magnetic resonance imaging should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
  • Exact FDA submission
  • Checked 2026-09-03
Monitoring
Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release; map the installed configuration to the exact clearance and labeling.
  • Product family
  • Checked 2026-09-03

Current vendor materials are product-family evidence for K223458; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Release and scope verification

Map the installed model, build, options, population, anatomy, inputs, outputs, and enabled AI functions to K223458 before acceptance.

exact submission · checked 2026-09-03

Later product-family features are not evidence that this cleared release includes them.

Interoperability acceptance

Scanner-console and DICOM workflow to PACS or downstream applications; validate worklist, identifiers, series labels, geometry, measurements, routing, and archive completeness.

product family · checked 2026-09-03

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally. Vendor statements are current product-family context; confirm the exact cleared release and installed options.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release; map the installed configuration to the exact clearance and labeling.

exact submission · checked 2026-09-03

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-03

No exact-product price was found in the reviewed public material; vendor productivity claims are scenario inputs, not guaranteed local savings.

Lifecycle monitoring

Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release; map the installed configuration to the exact clearance and labeling.

product family · checked 2026-09-03

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. Vendor statements are current product-family context; confirm the exact cleared release and installed options.

Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K223458
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
MRI
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Anatomy
Multi-region
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2023-04-06
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Human reviewedStatus
2026-09-04Last searched
28Fields reviewed
14Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. 8 FDA recall candidates were manually reviewed: 8 published and 0 rejected as a neighboring or otherwise inapplicable version. Native FDA recall identifiers produced 8 postmarket candidate records; 8 have been reviewed (8 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, fda decision material, vendor product page, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, fda decision summary packet, literature index, trial registry, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not publicly available - checked 2026-09-03 - Exact FDA submission

    No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Installation timeline: Not yet reviewed

16 sources - 3 history items

Source ledger

  1. FDA K223458: Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems

    U.S. Food and Drug Administration - Published 2023-04-06 - Accessed 2026-09-19

    Scope: modality, anatomy, input, output, purpose, intendedUsers, limitations, evidenceSummary, clinicalUse

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. FDA recall Z-1954-2026: Philips Evolution Upgrade 1.5T with MR Elastography (MRE). 1. Model Number (REF): 782116 2. Model Number (REF): 782148. 3. Model Number (REF): 782166.

    U.S. Food and Drug Administration - Published 2026-04-27 - Accessed 2026-09-04

    Scope: postmarketSafety, affectedProduct, recallStatus, correctiveAction

  4. FDA recall Z-1955-2026: Philips Evolution upgrade 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782143. 2. Model Number (REF): 782162.

    U.S. Food and Drug Administration - Published 2026-04-27 - Accessed 2026-09-04

    Scope: postmarketSafety, affectedProduct, recallStatus, correctiveAction

  5. FDA recall Z-1958-2026: Philips Ingenia 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781342. 2. Model Number (REF): 781377. 3. Model Number (REF): 782103. 510(k) Numbers: K163116, K173079, K183063, K193215, K213516, K213583, K223458, K251397, K251808

    U.S. Food and Drug Administration - Published 2026-04-27 - Accessed 2026-09-04

    Scope: postmarketSafety, affectedProduct, recallStatus, correctiveAction

  6. FDA recall Z-1961-2026: Philips Ingenia Ambition X with MR Elastography (MRE). 1. Model Number (REF): 781356. 2. Model Number (REF): 782109. 3. Model Number (REF): 782138. 4. Model Number (REF): 782160.

    U.S. Food and Drug Administration - Published 2026-04-27 - Accessed 2026-09-04

    Scope: postmarketSafety, affectedProduct, recallStatus, correctiveAction

  7. FDA recall Z-1962-2026: Philips Ingenia Elition S with MR Elastography (MRE). 1. Model Number (REF): 781357. 2. Model Number (REF): 782106. 3. Model Number (REF): 782137.

    U.S. Food and Drug Administration - Published 2026-04-27 - Accessed 2026-09-04

    Scope: postmarketSafety, affectedProduct, recallStatus, correctiveAction

  8. FDA recall Z-1964-2026: Philips MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782120. 2. Model Number (REF): 782153.

    U.S. Food and Drug Administration - Published 2026-04-27 - Accessed 2026-09-04

    Scope: postmarketSafety, affectedProduct, recallStatus, correctiveAction

  9. FDA recall Z-1967-2026: Philips SmartPath to dStream for XR and 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781270. 2. Model Number (REF): 782113.

    U.S. Food and Drug Administration - Published 2026-04-27 - Accessed 2026-09-04

    Scope: postmarketSafety, affectedProduct, recallStatus, correctiveAction

  10. FDA recall Z-1969-2026: Philips Upgrade to MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782130.

    U.S. Food and Drug Administration - Published 2026-04-27 - Accessed 2026-09-04

    Scope: postmarketSafety, affectedProduct, recallStatus, correctiveAction

  11. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  12. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  13. Medicare Coverage Database

    Centers for Medicare & Medicaid Services - Accessed 2026-09-01

    Scope: coverage, codingContext

  14. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  15. Philips SmartSpeed AI

    Philips - Accessed 2026-09-03

    Scope: productFamilyContext, workflowRole, implementationCapabilities

  16. Philips: Responsible Disclosure Statement

    Philips - Accessed 2026-10-26

    Scope: security

Research history

  1. K223458 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    human reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K260746

S-scan Open (100001800)

Esaote, S.p.A.

Authorized purpose: S-scan Open is an Esaote MR system for diagnostic image acquisition, reconstruction, display, and professional interpretation within the labeled scanner, coil, sequence, and anatomy scope.

Modality
MRI
Anatomy
Multi-region
Decision
2026-03-27
FDA-cleared K252379

AIR Recon DL

Ge Medical Systems, LLC

Authorized purpose: AIR Recon DL adds a deep-learning phase-correction and reconstruction option to supported GE 1.5T, 3T, and 7T MR workflows for all labeled anatomies and ages to reduce noise or ringing and support image quality or scan-time choices.

Modality
MRI
Anatomy
Multi-region
Decision
2025-12-23
FDA-cleared K252838

MAGNETOM Sola; MAGNETOM Altea; MAGNETOM Sola Fit; MAGNETOM Viato.Mobile; MAGNETOM Vida; MAGNETOM Lumina; MAGNETOM Vida Fit; MAGNETOM Flow.Elite; MAGNETOM Flow.Neo; MAGNETOM Flow.Rise

Siemens Healthcare GmbH

Authorized purpose: The listed MAGNETOM 1.5T and 3T MRI systems produce cross-sectional images, spectroscopic images or spectra, and derived physical parameters of the head, body, and extremities for trained-physician interpretation and can support imaging during MR-compatible interventions.

Modality
MRI
Anatomy
Multi-region
Decision
2025-12-19

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination

Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.

Accepted repository change history

No entry has been added to this scoped change history for this record. This is not proof that the record never changed.

Correction policy and limits · Public repository change history