FDA-cleared profile

icobrain

icometrix NV · Radiological images; modality varies or is not specified in FDA list · FDA-cleared (510(k))

Authorized purpose: icobrain segments and quantifies brain structures and lesions from MRI or noncontrast CT using cross-sectional and longitudinal workflows with acquisition-consistency limits.

Clinical role
Brain-image registration, segmentation, quantification, visualization, or procedure-planning support.
Images and patients
Radiological images; modality varies or is not specified in FDA list · Brain
Key limitation
Keep in mind: Longitudinal MRI requires the same scanner, protocol, and contrast conditions; cross-sectional and longitudinal outputs are not interchangeable.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: icobrain segments and quantifies brain structures and lesions from MRI or noncontrast CT using cross-sectional and longitudinal workflows with acquisition-consistency limits.
  • Which images or clinical context are in scope?Modality: Radiological images; modality varies or is not specified in FDA list Anatomy or clinical context: Brain Input: icobrain is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR or NCCT images.
  • Who uses it?Intended users: icobrain is used by trained professionals in hospitals, imaging centers or in image processing labs.
  • What does the software produce?Output: Multiple electronic reports with volumetric information of brain structures and midline shift; annotated DICOM images.
  • What is its most important limitation?Key limitation: Longitudinal MRI requires the same scanner, protocol, and contrast conditions; cross-sectional and longitudinal outputs are not interchangeable.

Clinical takeaway. Clinical role: icobrain segments and quantifies brain structures and lesions from MRI or noncontrast CT using cross-sectional and longitudinal workflows with acquisition-consistency limits.

Evidence takeaway. Evidence from the reviewed source record: K181939 reports CT-pipeline validation on 544 datasets with Pearson correlation 0.95 and ICC 0.94, plus software and MRI workflow validation.

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
icobrain segments and quantifies brain structures and lesions from MRI or noncontrast CT using cross-sectional and longitudinal workflows with acquisition-consistency limits.
  • Exact FDA submission
  • Checked 2026-09-19

Clinical-use display mirrors the exact FDA-supported purpose reconciled from the decision-summary packet.

Workflow role
Brain-image registration, segmentation, quantification, visualization, or procedure-planning support.
  • Exact FDA submission
  • Checked 2026-09-03
Required input
icobrain is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR or NCCT images.
  • Exact FDA submission
  • Checked 2026-09-19

This replaces the broader current input value, including its unsupported associated-workflow-data wording, with the exact indication text.

Output and clinician action
Multiple electronic reports with volumetric information of brain structures and midline shift; annotated DICOM images.
  • Exact FDA submission
  • Checked 2026-09-19

The packet's comparison table directly states these outputs; it does not establish the current registered-images, maps, alerts, or planning-visualization list.

Care setting
Hospital neuroimaging, stroke, or neurosurgical planning workflows.
  • Exact FDA submission
  • Checked 2026-09-03
Intended users
icobrain is used by trained professionals in hospitals, imaging centers or in image processing labs.
  • Exact FDA submission
  • Checked 2026-09-19

The exact packet gives this user and environment wording and does not list the current dossier's specific professional roles.

Limitations
Longitudinal MRI requires the same scanner, protocol, and contrast conditions; cross-sectional and longitudinal outputs are not interchangeable.
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Evidence summary
K181939 reports CT-pipeline validation on 544 datasets with Pearson correlation 0.95 and ICC 0.94, plus software and MRI workflow validation.
  • Exact FDA submission
  • Checked 2026-09-03

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
K181939 reports CT-pipeline validation on 544 datasets with Pearson correlation 0.95 and ICC 0.94, plus software and MRI workflow validation.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Reported sensitivity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-03

Sensitivity is not the primary endpoint for this acquisition, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

Reported specificity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-03

Specificity is not the primary endpoint for this acquisition, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

CT volumetry correlation

0.95 Pearson r

  • EndpointAutomated versus reference brain CT volumetry
  • Populationicobrain
  • Tested versionicobrain release cleared in K181939

CT volumetry agreement

0.94 ICC

  • EndpointAutomated versus reference brain CT volumetry
  • Populationicobrain
  • Tested versionicobrain release cleared in K181939

K181939 FDA performance package for icobrain

  • DesignManufacturer multimodality segmentation, volumetry, and agreement validation
  • Population544 noncontrast brain CT datasets plus supported MRI brain workflows
  • Scopeexact submission
  • Tested versionicobrain release cleared in K181939
  • Samplen=544
  • IndependenceSponsor or vendor study

K181939

Manufacturer evidence submitted for clearance or approval; local clinical, technical, and workflow acceptance remains necessary.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K181939 by exact submission identifier. This does not exclude product-family or later safety information.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-03

Exact-identifier review does not prove absence of all product-family safety signals.

Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not publicly available: Reviewed sources do not publish an exact-product procurement price with a stated offer scope.

Reimbursement

Reimbursement
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging, procedural, or surgical service line and verify payer, site-of-service, and coding assumptions independently.
  • Service-level, not product-specific
  • Checked 2026-09-03

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

Build a local total-cost and value model; do not infer reimbursement from FDA clearance or approval.

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: icobrain is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR or NCCT images.

    Scope: Exact FDA submission. This replaces the broader current input value, including its unsupported associated-workflow-data wording, with the exact indication text.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-09-19 · FDA K181939: icobrain

  2. Routing

    Source context, not route validation: DICOM import and output to PACS, workstations, or surgical systems; validate series selection, orientation, registration, measurements, routing, latency, and failure handling.

    Scope: Exact FDA submission.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-09-03 · FDA K181939: icobrain · icobrain MRI workflow and guidelines

  3. AI processing

    Source context, not route validation: Software workstation, server, virtual appliance, or connected clinical application using the cleared release.

    Scope: Exact FDA submission.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-09-03 · FDA K181939: icobrain · icobrain MRI workflow and guidelines

  4. Returned objects

    Source context, not route validation: Multiple electronic reports with volumetric information of brain structures and midline shift; annotated DICOM images.

    Scope: Exact FDA submission. The packet's comparison table directly states these outputs; it does not establish the current registered-images, maps, alerts, or planning-visualization list.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-19 · FDA K181939: icobrain

  5. Viewer/worklist/report

    Unresolved route: Multiple electronic reports with volumetric information of brain structures and midline shift; annotated DICOM images.

    Scope: Exact FDA submission. The packet's comparison table directly states these outputs; it does not establish the current registered-images, maps, alerts, or planning-visualization list.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-19 · FDA K181939: icobrain

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Deployment
Software workstation, server, virtual appliance, or connected clinical application using the cleared release.
  • Exact FDA submission
  • Checked 2026-09-03
Integration
DICOM import and output to PACS, workstations, or surgical systems; validate series selection, orientation, registration, measurements, routing, latency, and failure handling.
  • Exact FDA submission
  • Checked 2026-09-03
Training
Only qualified radiologists, neurologists, neurosurgeons, and trained imaging or surgical staff. should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
  • Exact FDA submission
  • Checked 2026-09-03
Monitoring
Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track these product-specific limits: Longitudinal MRI requires the same scanner, protocol, and contrast conditions; cross-sectional and longitudinal outputs are not interchangeable.
  • Exact FDA submission
  • Checked 2026-09-03

Release and scope verification

Map the installed model, build, options, population, anatomy, inputs, and outputs to K181939 before acceptance.

exact submission · checked 2026-09-03

Later product-family features are not evidence that this cleared release includes them.

Interoperability acceptance

DICOM import and output to PACS, workstations, or surgical systems; validate series selection, orientation, registration, measurements, routing, latency, and failure handling.

exact submission · checked 2026-09-03

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Longitudinal MRI requires the same scanner, protocol, and contrast conditions; cross-sectional and longitudinal outputs are not interchangeable.

exact submission · checked 2026-09-03

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-03

Vendor productivity claims are scenario inputs, not guaranteed local savings.

Lifecycle monitoring

Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track these product-specific limits: Longitudinal MRI requires the same scanner, protocol, and contrast conditions; cross-sectional and longitudinal outputs are not interchangeable.

exact submission · checked 2026-09-03

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes.

Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K181939
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
Radiological images; modality varies or is not specified in FDA list
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

The exact indication states MR or NCCT images; it does not support the current unspecified modality wording. Controlled directory display reconciled from exact FDA wording: MR or NCCT images. -> Radiological images; modality varies or is not specified in FDA list.

Anatomy
Brain
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2018-11-06
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
27Fields reviewed
13Source classes checked
20Screened PubMed leads
6Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains 20 PubMed and 6 ClinicalTrials.gov candidate records. Discovery metadata alone does not establish exact tested-product, version, population, and endpoint applicability, so all candidates remain quarantined and none was promoted. The current discovery snapshot contains 20 PubMed and 6 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, fda decision material, vendor product page, reimbursement policy, fda device recall, fda decision summary packet, literature index, trial registry, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not publicly available - checked 2026-09-03 - Exact FDA submission

    No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, enabled features, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Security: Not verified - checked 2026-09-03 - Exact FDA submission

    Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.

  • Installation timeline: Not yet reviewed

8 sources - 3 history items

Source ledger

  1. FDA K181939: icobrain

    U.S. Food and Drug Administration - Published 2018-11-06 - Accessed 2026-09-19

    Scope: purpose, anatomy, limitations, evidenceSummary, modality, intendedUsers, input, output, clinicalUse

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. icobrain model card (K181939)

    American College of Radiology - Published 2026-06-26 - Accessed 2026-09-01

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. Medicare Coverage Database

    Centers for Medicare & Medicaid Services - Accessed 2026-09-01

    Scope: coverage, codingContext

  7. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  8. icobrain MRI workflow and guidelines

    icometrix - Accessed 2026-09-03

    Scope: integration, deployment, monitoring, productFamilyContext

Research history

  1. K181939 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K260316

CMN Capillary Function with Oncology Virtual Expert

Cercare Medical A/S

Authorized purpose: Cercare Neurosuite with Capillary Function and Oncology Virtual Expert processes CT or MR information to provide perfusion, tissue-blood-volume, diffusion, labeling, and quantification support for adult brain glioma workflows.

Modality
Radiological images; modality varies or is not specified in FDA list
Anatomy
Brain
Decision
2026-06-22
FDA-cleared K252563

Neurophet SCALE PET

Neurophet., Inc.

Authorized purpose: Brain MRI/PET registration, segmentation and quantitative PET assessment

Modality
Radiological images; modality varies or is not specified in FDA list
Anatomy
Brain
Decision
2026-05-15
FDA-cleared K252496

Neurophet AQUA AD Plus

Neurophet., Inc.

Authorized purpose: Quantitative brain MRI and amyloid-PET analysis, including volumetric reference-percentile comparison and SUVR quantification, to support physician clinical evaluation.

Modality
Radiological images; modality varies or is not specified in FDA list
Anatomy
Brain
Decision
2026-01-29
FDA-cleared K241083

SIS System

Surgical Information Sciences, Inc.

Authorized purpose: SIS System fuses clinical brain MRI with postoperative CT to visualize DBS anatomy, localize STN, GPe, GPi, and Vim, and segment implanted leads for surgical review.

Modality
Radiological images; modality varies or is not specified in FDA list
Anatomy
Brain
Decision
2024-06-14

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination

Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.

Accepted repository change history

No entry has been added to this scoped change history for this record. This is not proof that the record never changed.

Correction policy and limits · Public repository change history