Hepatica v1 (Hepatica v1.0.0)

Perspectum Ltd.

Hepatica (Hepatica v1) is a post-processing medical device software that presents quantified metrics which may contribute to the assessment of a patient's liver health. Hepatica (Hepatica v1) uses image visualization and analysis tools to process DICOM 3.0 compliant magnetic resonance image datasets to produce semi-automatic segmented 3D models of the liver based on the work of Couinaud and the Brisbane 2000 terminology. For each identified Couinaud segment, volumetric data is determined and reported. Hepatica (Hepatica v1) may also report iron corrected-T1 (cT1) and PDFF calculated using the IDEAL method from multi-slice acquisitions, on a per segment basis, over the whole liver. Both metrics present numerical values of different fundamental liver tissue characteristics that can be used as measures of liver tissue health. Hepatica (Hepatica v1) provides trained clinicians with additional information to evaluate the volume and health of patient's liver on a segmental basis. It is not intended to replace the established procedures for the assessment of patient's liver health. However, information gathered through existing diagnostic tests, clinical evaluation of the patient, as well Hepatica (Hepatica v1), may support surgical decision making.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
Hepatica (Hepatica v1) is a post-processing medical device software that presents quantified metrics which may contribute to the assessment of a patient's liver health. Hepatica (Hepatica v1) uses image visualization and analysis tools to process DICOM 3.0 compliant magnetic resonance image datasets to produce semi-automatic segmented 3D models of the liver based on the work of Couinaud and the Brisbane 2000 terminology. For each identified Couinaud segment, volumetric data is determined and reported. Hepatica (Hepatica v1) may also report iron corrected-T1 (cT1) and PDFF calculated using the IDEAL method from multi-slice acquisitions, on a per segment basis, over the whole liver. Both metrics present numerical values of different fundamental liver tissue characteristics that can be used as measures of liver tissue health. Hepatica (Hepatica v1) provides trained clinicians with additional information to evaluate the volume and health of patient's liver on a segmental basis. It is not intended to replace the established procedures for the assessment of patient's liver health. However, information gathered through existing diagnostic tests, clinical evaluation of the patient, as well Hepatica (Hepatica v1), may support surgical decision making.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Intended users
The device is intended for use by Perspectum's trained operators. The device output is intended for interpretation by physicians.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Intended-user text reported in the exact-submission ACR model card.

Care setting and population
Adult and Pediatric
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Patient-population text reported in the exact-submission ACR model card; institutional setting may still require vendor confirmation.

Workflow role
Segmentation or contouring
  • Public source
  • Exact FDA submission
  • Checked 2026-08-31

Conservative normalization of explicit wording in an exact-submission source.

Required input
MRI images or data involving Liver; consult the FDA labeling for exact protocol, series, and quality requirements.
  • Public source
  • Exact FDA submission
  • Checked 2026-08-31

Conservative normalization of explicit wording in an exact-submission source.

Output and human action
Per Couinaud segment and across the whole liver: Proton density fat fraction in the liver (PDFF); T1 corrected for liver iron concentration, scanner manufacturer and field strength (cT1); Volume (ml)
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Clinical output reported in the exact-submission ACR model card.

Limitations
Exclusion criteria: non-DICOM 3.0 compliant data
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Limitations and supported-acquisition details reported in the exact-submission ACR model card.

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K203280
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
MRI
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Anatomy
Liver
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2021-01-12
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
The exact-submission ACR model card explicitly references DICOM in the described input, output, or workflow; exact interfaces, versions, and local compatibility require confirmation.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Interoperability terms explicitly reported in the exact-submission ACR model card; interface versions and local compatibility still require confirmation.

Deployment and data flow
On-premise
  • Public source
  • Exact FDA submission
  • Checked 2026-08-13

Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.

Security and privacy
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.

Training and support
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

Monitoring and change control
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

Input or output interoperability

DICOM is explicitly referenced in the described input, output, or workflow.

exact submission · checked 2026-09-01

The model card does not establish every supported DICOM object, transfer method, or local interface.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for Hepatica v1 (Hepatica v1.0.0); FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-08

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
Study type: Both
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Structured study metadata from the exact-submission ACR model card; this is not an independent evidence synthesis.

Reported sensitivity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

Reported specificity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

Hepatica regulatory performance study summary

  • DesignStand-alone performance and reader study (ACR AI Central category)
  • PopulationAdult and Pediatric
  • Scopeexact submission
  • Tested versionv1
  • IndependenceIndependence not established

K203280

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

No public price or quote for the exact K203280 configuration was established in this review.

Reimbursement and coding
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-08

No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K203280. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-08

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-08Last searched
23Fields reviewed
9Source classes checked
20Awaiting review PubMed leads
10Awaiting review trial leads
0Unreviewed FDA recall leads

Exact-submission ACR model-card fields were normalized under the current provenance rules; vendor and independent-study confirmation remain distinct. Exact FDA scope controls clinical claims. ACR and official vendor material provide bounded catalog or product-family context. Candidate literature metadata remains a discovery queue until full-text identity, endpoint, population, and version review is completed. Automated exact-name discovery found 20 PubMed and 10 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA K203280: Hepatica v1 (Hepatica v1.0.0)

    U.S. Food and Drug Administration - Published 2021-01-12 - Accessed 2026-09-08

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. Hepatica model card (K203280)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, limitations, evidenceSummary, integration, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K203280

    U.S. Food and Drug Administration - Accessed 2026-09-08

    Scope: postmarketSafety

Research history

  1. K203280 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

Alternatives

Related tools to compare

FDA-cleared K231459

HepaFatSmart (V2.0.0)

Resonance Health Analysis Services Pty Ltd

Provide quantitative volumetric liver fat fraction, proton-density fat fraction, and steatosis grading to support clinical diagnosis and subsequent management decisions for...

Modality
MRI
Anatomy
Liver
Decision
2023-06-20
FDA-cleared K213776

LiverSmart

Resonance Health Analysis Services Pty Ltd

LiverSmart is indicated to -For Liver Iron Concentration:(i) measure liver iron concentration in individuals with confirmed or suspected systemic iron overload;(ii) monitor liver...

Modality
MRI
Anatomy
Liver
Decision
2021-12-29