FDA-cleared profile

HealthCCS

Zebra Medical Vision Ltd. · CT/CTA · FDA-cleared (510(k))

Authorized purpose: Non-invasive post-processing to evaluate coronary calcified plaques and generate a total coronary-calcium risk category.

Clinical role
Automatic non-invasive post-processing of pre-existing ECG-Gated/Triggered heart or chest CT scans that identifies and quantifies coronary calcification and reports a 4-category Agatston-equivalent risk score for a physician's further analysis and treatment.
Images and patients
CT/CTA · Heart/Cardiovascular
Key limitation
Keep in mind: For patients older than 20; score management depends on physician judgment and may require further testing.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: Non-invasive post-processing to evaluate coronary calcified plaques and generate a total coronary-calcium risk category.
  • Which images or clinical context are in scope?Modality: CT/CTA Anatomy or clinical context: Heart/Cardiovascular Input: Pre-existing ECG-gated or triggered non-contrast heart or chest CT in DICOM format.
  • Who uses it?Intended users: Physicians.
  • What does the software produce?Output: Calcification detection above 130 HU, a CNN probability threshold, Agatston-equivalent and averaged scores, and a four-category report.
  • What is its most important limitation?Key limitation: For patients older than 20; score management depends on physician judgment and may require further testing.

Clinical takeaway. Clinical role: Non-invasive post-processing to evaluate coronary calcified plaques and generate a total coronary-calcium risk category.

Evidence takeaway. Evidence from the reviewed source record: Retrospective study of 249 cases with three-radiologist ground truth; agreement was 0.89 (95% CI 0.85-0.92), with repeatability tested on 150 studies.

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
Non-invasive post-processing to evaluate coronary calcified plaques and generate a total coronary-calcium risk category.
  • Exact FDA submission
  • Checked 2026-09-15

Clinical-use display mirrors the exact FDA-supported purpose reconciled in this remediation finding.

Workflow role
Automatic non-invasive post-processing of pre-existing ECG-Gated/Triggered heart or chest CT scans that identifies and quantifies coronary calcification and reports a 4-category Agatston-equivalent risk score for a physician's further analysis and treatment.
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "The HealthCCS Device is an automatic non-invasive post processing tool that uses cardiac CT images to identify and quantify calcification in the coronary arteries," … "The HealthCCS Device analyses pre-existing heart or chest ECG-Gated/Triggered CT scans. The Device is indicated for use only on patients whose age at the time, when the CT scan was taken, was above 20 years old. This device generates a 4-category Agatston-equivalent risk score,". Read from the exact FDA decision document on 2026-11-14.

Required input
Pre-existing ECG-gated or triggered non-contrast heart or chest CT in DICOM format.
  • Exact FDA submission
  • Checked 2026-09-15

FDA packet acquisition scope.

Output and clinician action
Calcification detection above 130 HU, a CNN probability threshold, Agatston-equivalent and averaged scores, and a four-category report.
  • Exact FDA submission
  • Checked 2026-09-15

FDA packet output scope.

Care setting
Adult
  • Exact FDA submission
  • Checked 2026-09-01

Patient-population text reported in the exact-submission ACR model card; institutional setting may still require vendor confirmation.

Intended users
Physicians.
  • Exact FDA submission
  • Checked 2026-09-15

FDA packet intended-user statement.

Limitations
For patients older than 20; score management depends on physician judgment and may require further testing.
  • Exact FDA submission
  • Checked 2026-09-15

Exact FDA limitation boundary.

Evidence summary
Retrospective study of 249 cases with three-radiologist ground truth; agreement was 0.89 (95% CI 0.85-0.92), with repeatability tested on 150 studies.
  • Exact FDA submission
  • Checked 2026-09-15

Submission-reported study result.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
Retrospective study of 249 cases with three-radiologist ground truth; agreement was 0.89 (95% CI 0.85-0.92), with repeatability tested on 150 studies.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

Submission-reported study result.

Reported sensitivity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

Reported specificity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

HealthCCS regulatory performance study summary

  • DesignStand-Alone Performance
  • PopulationAdult
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K172983
  • Samplen=249
  • IndependenceIndependence not established

K172983

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K172983. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-09
Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not verified: Pricing information is recorded but not verified against a resolving source.

Reimbursement

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: Pre-existing ECG-gated or triggered non-contrast heart or chest CT in DICOM format.

    Scope: Exact FDA submission. FDA packet acquisition scope.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-09-15 · FDA K172983: HealthCCS

  2. Routing

    Unresolved route: Not established for the exact route.

    Scope: Exact FDA submission. The exact FDA decision document was reviewed on 2026-11-14 and does not state the subject device's DICOM, PACS, RIS, or EHR interfaces. Confirm supported DICOM services and PACS, RIS, EHR, and interface versions with the vendor.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-11-14 ·

  3. AI processing

    Unresolved route: Not established for the exact route.

    Scope: Exact FDA submission. The reviewed sources do not establish the current exact-release hosting topology, data flow, region, tenancy, or disaster-recovery configuration.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-09-09 ·

  4. Returned objects

    Source context, not route validation: Calcification detection above 130 HU, a CNN probability threshold, Agatston-equivalent and averaged scores, and a four-category report.

    Scope: Exact FDA submission. FDA packet output scope.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-15 · FDA K172983: HealthCCS

  5. Viewer/worklist/report

    Unresolved route: Calcification detection above 130 HU, a CNN probability threshold, Agatston-equivalent and averaged scores, and a four-category report.

    Scope: Exact FDA submission. FDA packet output scope.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-15 · FDA K172983: HealthCCS

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Supported acquisition environment

Scanner manufacturers: All

exact submission · checked 2026-09-01

Confirm currently supported software versions, protocols, and site-specific acquisition configurations.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for HealthCCS; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-09
Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K172983
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
CT/CTA
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

FDA specifies non-contrast cardiac CT; catalog uses CT/CTA as the CT bucket.

Anatomy
Heart/Cardiovascular
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

FDA packet covers coronary arteries, heart, and chest.

Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2018-06-13
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
24Fields reviewed
10Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, fda decision summary packet, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not verified - checked 2026-09-09 - Exact FDA submission

    No public price or quote for the exact K172983 configuration was established in this review.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Reimbursement: Not publicly available - checked 2026-09-09 - Exact FDA submission

    No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

  • Deployment: Not verified - checked 2026-09-09 - Exact FDA submission

    The reviewed sources do not establish the current exact-release hosting topology, data flow, region, tenancy, or disaster-recovery configuration.

  • Integration: Not publicly available - checked 2026-11-14 - Exact FDA submission

    The exact FDA decision document was reviewed on 2026-11-14 and does not state the subject device's DICOM, PACS, RIS, or EHR interfaces. Confirm supported DICOM services and PACS, RIS, EHR, and interface versions with the vendor.

  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Security: Not verified - checked 2026-09-09 - Exact FDA submission

    Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.

  • Installation timeline: Not yet reviewed
  • Training: Not verified - checked 2026-09-09 - Exact FDA submission

    Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

  • Monitoring: Not verified - checked 2026-09-09 - Exact FDA submission

    Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

7 sources - 3 history items

Source ledger

  1. FDA K172983: HealthCCS

    U.S. Food and Drug Administration - Published 2018-06-13 - Accessed 2026-09-09

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation, purpose, intendedUsers, input, output, limitations, evidenceSummary

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. HealthCCS model card (K172983)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, limitations, evidenceSummary, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K172983

    U.S. Food and Drug Administration - Accessed 2026-09-09

    Scope: postmarketSafety

Research history

  1. K172983 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K253786

syngo.CT Coronary Cockpit

Siemens Medical Solutions USA, Inc.

Authorized purpose: An extension for syngo.via that assists trained physicians with automated and semi-automated analysis, visualization and reporting of cardiac CT angiography volume datasets for coronary-lesion evaluation, documentation, follow-up and treatment planning.

Modality
CT/CTA
Anatomy
Heart/Cardiovascular
Decision
2026-06-25
FDA-cleared K260169

AV Cardiac CT

Philips Medical Systems Nederland B.V.

Authorized purpose: AV Cardiac CT assists radiologists, cardiologists and 3D technologists with viewing, processing and analysis of gated cardiac CT data, including coronary anatomy and cardiac function, to prepare cardiac interventions; physician review and decisions remain required.

Modality
CT/CTA
Anatomy
Heart/Cardiovascular
Decision
2026-03-05
FDA-cleared K252217

CT VScore+

Canon Medical Informatics, Inc.

Authorized purpose: CT VScore+ evaluates calcified coronary lesions on ECG-gated, non-contrast cardiac CT in patients aged 30 years or older, automatically generating coronary calcium scores and highlighted calcium segmentation for qualified clinical review.

Modality
CT/CTA
Anatomy
Heart/Cardiovascular
Decision
2025-11-28

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination

Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.

Accepted repository change history

No entry has been added to this scoped change history for this record. This is not proof that the record never changed.

Correction policy and limits · Public repository change history