EFAI CARDIOSUITE CTA ACUTE AORTIC SYNDROME ASSESSMENT SYSTEM

Ever Fortune.AI, Co., Ltd. · CT/CTA · FDA-cleared (510(k))

Authorized purpose: Assist hospital networks and appropriately trained medical specialists in triaging adults aged 22 years and older by flagging suspected aortic dissection or aortic intramural hematoma on chest or chest-abdomen CTA.

Clinical role
Notification-only parallel triage and worklist prioritization; the device flags cases for earlier review without marking a location or replacing clinical assessment.
Images and patients
CT/CTA · Chest, abdomen, and thoracoabdominal aorta
Key limitation
Keep in mind: The output is for worklist prioritization or triage, not standalone clinical decision-making. It is not intended to rule out acute aortic syndrome or preclude clinical assessment of CTA cases, and it does not provide lesion marking or location-specific guidance.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: Assist hospital networks and appropriately trained medical specialists in triaging adults aged 22 years and older by flagging suspected aortic dissection or aortic intramural hematoma on chest or chest-abdomen CTA.
  • Which images or clinical context are in scope?Modality: CT/CTA Anatomy or clinical context: Chest, abdomen, and thoracoabdominal aorta Input: DICOM chest or chest-abdomen CTA studies from adults aged 22 years and older; the FDA validation set included 380 CTA studies with 156 positive and 224 negative cases.
  • Who uses it?Intended users: Hospital networks and appropriately trained medical specialists
  • What does the software produce?Output: Case-level passive notification or flag to a PACS/workstation indicating suspected aortic dissection or aortic intramural hematoma; the device does not mark, highlight, or direct attention to a specific image location.
  • What is its most important limitation?Key limitation: The output is for worklist prioritization or triage, not standalone clinical decision-making. It is not intended to rule out acute aortic syndrome or preclude clinical assessment of CTA cases, and it does not provide lesion marking or location-specific guidance.

Clinical takeaway. Clinical role: Assist hospital networks and appropriately trained medical specialists in triaging adults aged 22 years and older by flagging suspected aortic dissection or aortic intramural hematoma on chest or chest-abdomen CTA.

Evidence takeaway. Evidence from the reviewed source record: FDA K240291 reports a retrospective, blinded, multisite validation of 380 consecutively collected US CTA studies, including 156 positives and 224 negatives, independent of model development and analytical validation. Three US board-certified radiologists established ground truth by majority agreement. Sensitivity was 0.929 (95% CI 0.878-0.960) and specificity was 0.915 (95% CI 0.871-0.945). The secondary system-processing endpoint was 37.86 seconds per study (95% CI 35.22-40.50).

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
Assist hospital networks and appropriately trained medical specialists in triaging adults aged 22 years and older by flagging suspected aortic dissection or aortic intramural hematoma on chest or chest-abdomen CTA.
  • Exact FDA submission
  • Checked 2026-09-08
Workflow role
Notification-only parallel triage and worklist prioritization; the device flags cases for earlier review without marking a location or replacing clinical assessment.
  • Exact FDA submission
  • Checked 2026-09-08
Required input
DICOM chest or chest-abdomen CTA studies from adults aged 22 years and older; the FDA validation set included 380 CTA studies with 156 positive and 224 negative cases.
  • Exact FDA submission
  • Checked 2026-09-08
Output and clinician action
Case-level passive notification or flag to a PACS/workstation indicating suspected aortic dissection or aortic intramural hematoma; the device does not mark, highlight, or direct attention to a specific image location.
  • Exact FDA submission
  • Checked 2026-09-08
Care setting
Hospital networks / acute-care radiology workflow
  • Exact FDA submission
  • Checked 2026-09-08
Intended users
Hospital networks and appropriately trained medical specialists
  • Exact FDA submission
  • Checked 2026-09-08
Limitations
The output is for worklist prioritization or triage, not standalone clinical decision-making. It is not intended to rule out acute aortic syndrome or preclude clinical assessment of CTA cases, and it does not provide lesion marking or location-specific guidance.
  • Exact FDA submission
  • Checked 2026-09-08
Evidence summary
FDA K240291 reports a retrospective, blinded, multisite validation of 380 consecutively collected US CTA studies, including 156 positives and 224 negatives, independent of model development and analytical validation. Three US board-certified radiologists established ground truth by majority agreement. Sensitivity was 0.929 (95% CI 0.878-0.960) and specificity was 0.915 (95% CI 0.871-0.945). The secondary system-processing endpoint was 37.86 seconds per study (95% CI 35.22-40.50).
  • Exact FDA submission
  • Checked 2026-09-08

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
FDA K240291 reports a retrospective, blinded, multisite validation of 380 consecutively collected US CTA studies, including 156 positives and 224 negatives, independent of model development and analytical validation. Three US board-certified radiologists established ground truth by majority agreement. Sensitivity was 0.929 (95% CI 0.878-0.960) and specificity was 0.915 (95% CI 0.871-0.945). The secondary system-processing endpoint was 37.86 seconds per study (95% CI 35.22-40.50).
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Reported sensitivity
92.90%
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Model sensitivity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

Reported specificity
91.50%
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Model specificity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

EFAI CARDIOSUITE CTA ACUTE AORTIC SYNDROME ASSESSMENT SYSTEM regulatory performance study summary

  • DesignStand-Alone Performance
  • PopulationAdult
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K240291
  • Samplen=380
  • IndependenceIndependence not established

K240291

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K240291. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-08
Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not verified: Pricing information is recorded but not verified against a resolving source.

Reimbursement

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: DICOM chest or chest-abdomen CTA studies from adults aged 22 years and older; the FDA validation set included 380 CTA studies with 156 positive and 224 negative cases.

    Scope: Exact FDA submission.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-09-08 · FDA K240291: EFAI CARDIOSUITE CTA ACUTE AORTIC SYNDROME ASSESSMENT SYSTEM

  2. Routing

    Source context, not route validation: The packet describes DICOM input and passive output to a PACS/RIS workstation for worklist prioritization or triage. Specific interface versions, current deployment, and local compatibility are not established.

    Scope: Exact FDA submission.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-09-08 · FDA K240291: EFAI CARDIOSUITE CTA ACUTE AORTIC SYNDROME ASSESSMENT SYSTEM

  3. AI processing

    Unresolved route: Not established for the exact route.

    Scope: Exact FDA submission. The exact FDA decision document was reviewed on 2026-11-14 and does not state where the subject device's software runs or how it is hosted. Confirm the deployment model, hosting, data flow, and current commercial configuration with the vendor.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-11-14 ·

  4. Returned objects

    Source context, not route validation: Case-level passive notification or flag to a PACS/workstation indicating suspected aortic dissection or aortic intramural hematoma; the device does not mark, highlight, or direct attention to a specific image location.

    Scope: Exact FDA submission.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-08 · FDA K240291: EFAI CARDIOSUITE CTA ACUTE AORTIC SYNDROME ASSESSMENT SYSTEM

  5. Viewer/worklist/report

    Unresolved route: Case-level passive notification or flag to a PACS/workstation indicating suspected aortic dissection or aortic intramural hematoma; the device does not mark, highlight, or direct attention to a specific image location.

    Scope: Exact FDA submission.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-08 · FDA K240291: EFAI CARDIOSUITE CTA ACUTE AORTIC SYNDROME ASSESSMENT SYSTEM

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Integration
The packet describes DICOM input and passive output to a PACS/RIS workstation for worklist prioritization or triage. Specific interface versions, current deployment, and local compatibility are not established.
  • Exact FDA submission
  • Checked 2026-09-08
Security
Ever Fortune.AI's investor information-security policy page states an implemented information security management system covering confidentiality, integrity, and availability, with security training and measurement targets; the page states the company has completed ISO 27001 certification and holds ISO/IEC 27701 PII-processor certification after independent third-party audit. The company's certification-milestones page lists ISO/IEC 27001:2022, ISO 22301:2019, ISO/IEC 27018:2019, and ISO/IEC 27017:2015.
  • Vendor-wide
  • Checked 2026-10-27

Vendor statement; certificates were not read. Applicant name variants (including uppercase and punctuation variants of Ever Fortune.AI Co., Ltd.) are bridged via the vendor's EFAI product naming and the owning domain everfortuneai.com.tw. Exact-release scope is not established.

Input or output interoperability

DICOM is explicitly referenced in the described input, output, or workflow.

exact submission · checked 2026-09-01

The model card does not establish every supported DICOM object, transfer method, or local interface.

Result destination

PACS and workstation are explicitly referenced as a result destination.

exact submission · checked 2026-09-01

Confirm supported products, versions, routing behavior, and display requirements during local evaluation.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for EFAI CARDIOSUITE CTA ACUTE AORTIC SYNDROME ASSESSMENT SYSTEM; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-08
Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K240291
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
CT/CTA
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Anatomy
Chest, abdomen, and thoracoabdominal aorta
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2024-04-08
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
24Fields reviewed
9Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not verified - checked 2026-09-08 - Exact FDA submission

    No public price or quote for the exact K240291 configuration was established in this review.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Reimbursement: Not publicly available - checked 2026-09-08 - Exact FDA submission

    No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

  • Deployment: Not publicly available - checked 2026-11-14 - Exact FDA submission

    The exact FDA decision document was reviewed on 2026-11-14 and does not state where the subject device's software runs or how it is hosted. Confirm the deployment model, hosting, data flow, and current commercial configuration with the vendor.

  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Installation timeline: Not yet reviewed
  • Training: Not verified - checked 2026-09-08 - Exact FDA submission

    Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

  • Monitoring: Not verified - checked 2026-09-08 - Exact FDA submission

    Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

8 sources - 3 history items

Source ledger

  1. FDA K240291: EFAI CARDIOSUITE CTA ACUTE AORTIC SYNDROME ASSESSMENT SYSTEM

    U.S. Food and Drug Administration - Published 2024-04-08 - Accessed 2026-09-08

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. EFAI CARDIOSUITE CTA ACUTE AORTIC SYNDROME ASSESSMENT SYSTEM model card (K240291)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, reportedSensitivity, reportedSpecificity, evidenceSummary, integration, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K240291

    U.S. Food and Drug Administration - Accessed 2026-09-08

    Scope: postmarketSafety

  8. Ever Fortune.AI: Information Security Policy (ISO 27001 / ISO 27701)

    Ever Fortune.AI - Accessed 2026-10-27

    Scope: security

Research history

  1. K240291 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K253560

Enhanced Boundary for PCCT

Ge Medical Systems, LLC

Authorized purpose: Enhanced Boundary for PCCT is a deep-learning CT processing function for improved soft-tissue delineation in routine head examinations with or without contrast.

Modality
CT/CTA
Anatomy
Head
Decision
2026-06-24
FDA-cleared K253649

Spectral CT Verida Family

Philips Medical Systems Technologies, Ltd.

Authorized purpose: Spectral CT Verida Family supports the labeled spectral CT acquisition, reconstruction, material-map, and professional-interpretation workflow; exact models, protocols, and derived outputs remain submission-specific.

Modality
CT/CTA
Anatomy
Multi-region
Decision
2026-03-27
FDA-cleared K253574

SOMATOM X.cite; SOMATOM X.ceed

Siemens Medical Solutions USA, Inc.

Authorized purpose: SOMATOM X.cite and X.ceed with syngo CT VB20 support diagnostic CT, interventions, treatment preparation, radiation-therapy planning, and qualified low-dose lung screening.

Modality
CT/CTA
Anatomy
Multi-region
Decision
2026-03-20
FDA-cleared K252041

Fibresolve (with PCCP)

Imvaria, Inc.

Authorized purpose: Fibresolve analyzes lung CT in patients with pulmonary symptoms to provide an adjunct diagnostic-subtype classification for suspected interstitial lung disease and support referral to a multidisciplinary discussion.

Modality
CT/CTA
Anatomy
Lungs and interstitial lung disease patterns
Decision
2025-11-07

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination

Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.

Accepted repository change history

No entry has been added to this scoped change history for this record. This is not proof that the record never changed.

Correction policy and limits · Public repository change history