FDA-cleared profile

DrAid™ for Liver Segmentation

VinBrain Joint Stock Company · CT/CTA · FDA-cleared (510(k))

Authorized purpose: Support trained professionals in visualizing, evaluating, measuring, and reporting liver and physician-identified liver lesions on multiphase CT; it does not generate diagnoses or potential findings directly.

Clinical role
Semi-automated liver segmentation and quantitative measurement aid with manual seed-point and boundary editing and report generation.
Images and patients
CT/CTA · Liver and physician-identified liver lesions.
Key limitation
Keep in mind: Does not generate diagnoses or potential findings directly; qualified professionals make image interpretations and clinical decisions. The stated multiphase CT, slice-thickness <= 3.0 mm, adult, and DICOM scope applies. Installation and deployment are described as being carried out by VinBrain and trained technicians.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: Support trained professionals in visualizing, evaluating, measuring, and reporting liver and physician-identified liver lesions on multiphase CT; it does not generate diagnoses or potential findings directly.
  • Which images or clinical context are in scope?Modality: CT/CTA Anatomy or clinical context: Liver and physician-identified liver lesions. Input: Multiphase DICOM CT images with slice thickness <= 3.0 mm from patients older than 22 years; contrast-enhanced CT scans were used in the reported validation.
  • Who uses it?Intended users: Trained professionals including physicians and technicians; VinBrain and trained technicians carry out installation and deployment.
  • What does the software produce?Output: Editable liver segmentation mask; manual or physician-identified lesion segmentation; liver volume, lesion volume, maximum lesion diameter, lesion position, visualization, multiplanar reconstruction, and reporting outputs.
  • What is its most important limitation?Key limitation: Does not generate diagnoses or potential findings directly; qualified professionals make image interpretations and clinical decisions. The stated multiphase CT, slice-thickness <= 3.0 mm, adult, and DICOM scope applies. Installation and deployment are described as being carried out by VinBrain and trained technicians.

Clinical takeaway. Clinical role: Support trained professionals in visualizing, evaluating, measuring, and reporting liver and physician-identified liver lesions on multiphase CT; it does not generate diagnoses or potential findings directly.

Evidence takeaway. Evidence from the reviewed source record: FDA K241543 reports validation on 450 contrast-enhanced CT scans from 150 patients (age 60 +/- 12) at U.S. medical institutions, with ground truth from three U.S. board-certified radiologists. Liver segmentation had mean Dice 0.9649 (95% CI 0.9631-0.9667) and HD95 1.7061 (95% CI 1.5595-1.8526). Liver-volume measurement mean volume error was 2.7269% (95% CI 2.4308%-3.0230%). Subgroup and scanner analyses were also reported; no sensitivity or specificity disease-classification endpoint was reported.

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
Support trained professionals in visualizing, evaluating, measuring, and reporting liver and physician-identified liver lesions on multiphase CT; it does not generate diagnoses or potential findings directly.
  • Exact FDA submission
  • Checked 2026-09-08
Workflow role
Semi-automated liver segmentation and quantitative measurement aid with manual seed-point and boundary editing and report generation.
  • Exact FDA submission
  • Checked 2026-09-08
Required input
Multiphase DICOM CT images with slice thickness <= 3.0 mm from patients older than 22 years; contrast-enhanced CT scans were used in the reported validation.
  • Exact FDA submission
  • Checked 2026-09-08
Output and clinician action
Editable liver segmentation mask; manual or physician-identified lesion segmentation; liver volume, lesion volume, maximum lesion diameter, lesion position, visualization, multiplanar reconstruction, and reporting outputs.
  • Exact FDA submission
  • Checked 2026-09-08
Care setting
Healthcare facility or hospital using standard hospital/clinic workstations; the FDA-labeled population is older than 22 years.
  • Exact FDA submission
  • Checked 2026-09-08
Intended users
Trained professionals including physicians and technicians; VinBrain and trained technicians carry out installation and deployment.
  • Exact FDA submission
  • Checked 2026-09-08
Limitations
Does not generate diagnoses or potential findings directly; qualified professionals make image interpretations and clinical decisions. The stated multiphase CT, slice-thickness <= 3.0 mm, adult, and DICOM scope applies. Installation and deployment are described as being carried out by VinBrain and trained technicians.
  • Exact FDA submission
  • Checked 2026-09-08
Evidence summary
FDA K241543 reports validation on 450 contrast-enhanced CT scans from 150 patients (age 60 +/- 12) at U.S. medical institutions, with ground truth from three U.S. board-certified radiologists. Liver segmentation had mean Dice 0.9649 (95% CI 0.9631-0.9667) and HD95 1.7061 (95% CI 1.5595-1.8526). Liver-volume measurement mean volume error was 2.7269% (95% CI 2.4308%-3.0230%). Subgroup and scanner analyses were also reported; no sensitivity or specificity disease-classification endpoint was reported.
  • Exact FDA submission
  • Checked 2026-09-08

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
FDA K241543 reports validation on 450 contrast-enhanced CT scans from 150 patients (age 60 +/- 12) at U.S. medical institutions, with ground truth from three U.S. board-certified radiologists. Liver segmentation had mean Dice 0.9649 (95% CI 0.9631-0.9667) and HD95 1.7061 (95% CI 1.5595-1.8526). Liver-volume measurement mean volume error was 2.7269% (95% CI 2.4308%-3.0230%). Subgroup and scanner analyses were also reported; no sensitivity or specificity disease-classification endpoint was reported.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Reported sensitivity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

Reported specificity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

DrAid™ for Liver Segmentation regulatory performance study summary

  • DesignStand-Alone Performance
  • PopulationAdult
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K241543
  • Samplen=450
  • IndependenceIndependence not established

K241543

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K241543. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-08
Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not verified: Pricing information is recorded but not verified against a resolving source.

Reimbursement

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: Multiphase DICOM CT images with slice thickness <= 3.0 mm from patients older than 22 years; contrast-enhanced CT scans were used in the reported validation.

    Scope: Exact FDA submission.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-09-08 · FDA K241543: DrAid™ for Liver Segmentation

  2. Routing

    Source context, not route validation: Web-based application running on standard hospital/clinic workstations with DICOM CT processing, viewer/manual editing, reporting, and results export. The exact PACS, RIS, EHR interface and hosting model are not established by the packet.

    Scope: Exact FDA submission.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-09-08 · FDA K241543: DrAid™ for Liver Segmentation

  3. AI processing

    Source context, not route validation: The exact FDA package describes DrAid for Liver Segmentation as a web-based application running on standard hospital/clinic workstations, with installation and deployment carried out by VinBrain and trained technicians. Current hosting, data flow, and commercial configuration require vendor confirmation.

    Scope: Exact FDA submission. Packet wording: "Web-based application running on standard hospital/clinic workstations" … "The installation and deployment of the software medical device should be carried out by VinBrain and trained technicians.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-11-14 · FDA K241543: DrAid™ for Liver Segmentation

  4. Returned objects

    Source context, not route validation: Editable liver segmentation mask; manual or physician-identified lesion segmentation; liver volume, lesion volume, maximum lesion diameter, lesion position, visualization, multiplanar reconstruction, and reporting outputs.

    Scope: Exact FDA submission.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-08 · FDA K241543: DrAid™ for Liver Segmentation

  5. Viewer/worklist/report

    Unresolved route: Editable liver segmentation mask; manual or physician-identified lesion segmentation; liver volume, lesion volume, maximum lesion diameter, lesion position, visualization, multiplanar reconstruction, and reporting outputs.

    Scope: Exact FDA submission.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-08 · FDA K241543: DrAid™ for Liver Segmentation

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Deployment
The exact FDA package describes DrAid for Liver Segmentation as a web-based application running on standard hospital/clinic workstations, with installation and deployment carried out by VinBrain and trained technicians. Current hosting, data flow, and commercial configuration require vendor confirmation.
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "Web-based application running on standard hospital/clinic workstations" … "The installation and deployment of the software medical device should be carried out by VinBrain and trained technicians.". Read from the exact FDA decision document on 2026-11-14; this is the release-level description, not current commercial configuration.

Integration
Web-based application running on standard hospital/clinic workstations with DICOM CT processing, viewer/manual editing, reporting, and results export. The exact PACS, RIS, EHR interface and hosting model are not established by the packet.
  • Exact FDA submission
  • Checked 2026-09-08

Input or output interoperability

DICOM is explicitly referenced in the described input, output, or workflow.

exact submission · checked 2026-09-01

The model card does not establish every supported DICOM object, transfer method, or local interface.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for DrAid™ for Liver Segmentation; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-08
Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K241543
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
CT/CTA
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Anatomy
Liver and physician-identified liver lesions.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2024-12-06
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
23Fields reviewed
9Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not verified - checked 2026-09-08 - Exact FDA submission

    No public price or quote for the exact K241543 configuration was established in this review.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Reimbursement: Not publicly available - checked 2026-09-08 - Exact FDA submission

    No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Security: Not verified - checked 2026-09-08 - Exact FDA submission

    Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.

  • Installation timeline: Not yet reviewed
  • Training: Not verified - checked 2026-09-08 - Exact FDA submission

    Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

  • Monitoring: Not verified - checked 2026-09-08 - Exact FDA submission

    Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

7 sources - 3 history items

Source ledger

  1. FDA K241543: DrAid™ for Liver Segmentation

    U.S. Food and Drug Administration - Published 2024-12-06 - Accessed 2026-09-08

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation, deployment

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. DrAid™ for Liver Segmentation model card (K241543)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, evidenceSummary, integration, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K241543

    U.S. Food and Drug Administration - Accessed 2026-09-08

    Scope: postmarketSafety

Research history

  1. K241543 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K252992

CT Rembra RT; CT Areta RT; CT Rembra

Philips Healthcare (Suzhou) Co., Ltd.

Authorized purpose: CT Rembra RT, CT Areta RT, and CT Rembra support whole-body diagnostic imaging, oncology, radiotherapy planning, interventional, vascular, neurological, cardiovascular, and qualified lung-screening workflows for all ages.

Modality
CT/CTA
Anatomy
Multi-region
Decision
2026-03-23
FDA-cleared K252970

BriefCase-Triage: CARE Multi-triage CT Body

Aidoc Medical, Ltd.

Authorized purpose: Adult contrast-enhanced and non-contrast CT triage for 11 chest, abdominal, and pelvic findings: diverticulitis, abdominal-pelvic abscess, appendicitis, intestinal ischemia/pneumatosis, obstructive renal stone, small- and large-bowel obstruction, spleen injury, liver injury, kidney injury, and pelvic fracture.

Modality
CT/CTA
Anatomy
Chest, Abdomen, and Pelvis
Decision
2026-01-07
FDA-cleared K252029

AI-CVD

HeartLung Corporation

Authorized purpose: AI-CVD provides automated quantitative CT-derived anatomical and density measurements for clinician review, including calcium scores, cardiac and vascular measurements, lung and liver density, bone mineral density, and muscle/fat composition; it is opportunistic decision support, not diagnostic interpretation or risk prediction.

Modality
CT/CTA
Anatomy
Chest, abdomen, heart, lungs, liver, aorta, pulmonary artery, bone, muscle, and fat
Decision
2025-12-19
FDA-cleared K242624

Medical Image Post-processing Software (uOmnispace.CT)

Shanghai United Imaging Healthcare Co., Ltd.

Authorized purpose: uOmnispace.CT provides healthcare institutions with tools to view, manipulate, evaluate, and analyze medical CT images, including colon, lung, vessel, brain, cardiac, liver, rib, spine, dual-energy, perfusion, and cardiovascular applications described in the exact FDA submission.

Modality
CT/CTA
Anatomy
Colon, lungs and airway, vessels, brain, heart, liver, ribs, spine, and other CT regions
Decision
2025-05-14

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination

Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.

Accepted repository change history

No entry has been added to this scoped change history for this record. This is not proof that the record never changed.

Correction policy and limits · Public repository change history