Authorized purpose: D2P is a software interface and image-segmentation system for transferring DICOM imaging information from a medical scanner to an output file and is also preoperative software for surgical planning. The output may be used to produce a physical replica for adjunctive use with other diagnostic tools and expert clinical judgment in cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and musculoskeletal applications.
Clinical role
DICOM visualization, segmentation, editing, output-file generation, and adjunctive surgical planning.
Images and patients
Radiological images; modality varies or is not specified in FDA list · Cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and musculoskeletal anatomy
Key limitation
Keep in mind: Physical replicas are adjunctive and require other diagnostic tools and expert clinical judgment; application, printer, material, anatomy, and output-validation limits apply.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: D2P is a software interface and image-segmentation system for transferring DICOM imaging information from a medical scanner to an output file and is also preoperative software for surgical planning. The output may be used to produce a physical replica for adjunctive use with other diagnostic tools and expert clinical judgment in cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and musculoskeletal applications.
Which images or clinical context are in scope?Modality: Radiological images; modality varies or is not specified in FDA list Anatomy or clinical context: Cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and musculoskeletal anatomy Input: DICOM imaging information from a medical scanner and user-edited segmentation inputs.
Who uses it?Intended users: Surgeons, radiologists, and other trained healthcare professionals.
What does the software produce?Output: Segmented output files, 3D visualizations, physical anatomic replicas, and non-diagnostic VR views.
What is its most important limitation?Key limitation: Physical replicas are adjunctive and require other diagnostic tools and expert clinical judgment; application, printer, material, anatomy, and output-validation limits apply.
Clinical takeaway. Clinical role: D2P is a software interface and image-segmentation system for transferring DICOM imaging information from a medical scanner to an output file and is also preoperative software for surgical planning. The output may be used to produce a physical replica for adjunctive use with other diagnostic tools and expert clinical judgment in cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and musculoskeletal applications.
Evidence takeaway. Evidence from the reviewed source record: The exact FDA document reports non-clinical tests only: a measurement accuracy and calculation 3D study, usability study, and decimation study confirmed to be within specification, and validation of printing of physical replicas showing that anatomic models for cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and/or musculoskeletal cases can be printed accurately on compatible 3D printers and materials. The document gives no quantitative results and describes no clinical testing. Results are as reported in the FDA submission, not independent validation.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
D2P is a software interface and image-segmentation system for transferring DICOM imaging information from a medical scanner to an output file and is also preoperative software for surgical planning. The output may be used to produce a physical replica for adjunctive use with other diagnostic tools and expert clinical judgment in cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and musculoskeletal applications.
Physical replicas are adjunctive and require other diagnostic tools and expert clinical judgment; application, printer, material, anatomy, and output-validation limits apply.
The exact FDA document reports non-clinical tests only: a measurement accuracy and calculation 3D study, usability study, and decimation study confirmed to be within specification, and validation of printing of physical replicas showing that anatomic models for cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and/or musculoskeletal cases can be printed accurately on compatible 3D printers and materials. The document gives no quantitative results and describes no clinical testing. Results are as reported in the FDA submission, not independent validation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "A measurement accuracy and calculation 3D study, usability study, and decimation study were performed and confirmed to be within specification." … "Validation of printing of physical replicas was performed and demonstrated that anatomic models for cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and/or musculoskeletal cases can be printed accurately when using any of the compatible 3D printers and materials.". Read from the exact FDA decision document on 2026-11-14.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
The exact FDA document reports non-clinical tests only: a measurement accuracy and calculation 3D study, usability study, and decimation study confirmed to be within specification, and validation of printing of physical replicas showing that anatomic models for cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and/or musculoskeletal cases can be printed accurately on compatible 3D printers and materials. The document gives no quantitative results and describes no clinical testing. Results are as reported in the FDA submission, not independent validation.
FDA source
Exact FDA submission
Checked 2026-11-14
Packet wording: "A measurement accuracy and calculation 3D study, usability study, and decimation study were performed and confirmed to be within specification." … "Validation of printing of physical replicas was performed and demonstrated that anatomic models for cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and/or musculoskeletal cases can be printed accurately when using any of the compatible 3D printers and materials.". Read from the exact FDA decision document on 2026-11-14.
Sensitivity is not promoted as a product-wide endpoint. Compare only an exact endpoint, threshold, population, version, study design, and source if separately verified.
Reported specificity
Not applicable
Not applicable
Not applicable
Checked 2026-09-04
Specificity is not promoted as a product-wide endpoint. Compare only an exact endpoint, threshold, population, version, study design, and source if separately verified.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
K183489 FDA evidence package for D2P
DesignFDA 510(k) bench and substantial-equivalence evidence for DICOM transfer, visualization, segmentation, and physical-model output.
PopulationCardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and musculoskeletal anatomy
Scopeexact submission
Tested versionD2P Software configuration described in K183489
IndependenceSponsor or vendor study
K183489
Physical replicas are adjunctive and remain subject to expert clinical judgment and local output validation.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K183489 by exact native submission identifier. This does not exclude product-family, later-version, or adverse-event information.
Public source
Exact FDA submission
Checked 2026-09-04
Exact-identifier review does not prove absence of all product-family safety signals.
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging or clinical service line and verify payer, site-of-service, coding, and coverage assumptions independently.
No separate named-product payment identified in reviewed public material
Centers for Medicare & Medicaid Services · United States
This is not a product-specific reimbursement determination. Build a local total-cost and value model and verify payer, site-of-service, coding, and coverage assumptions independently.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: DICOM imaging information from a medical scanner and user-edited segmentation inputs.
Scope: Exact FDA submission.
Confirm scanner/model, protocol, series completeness and both software builds.
Source context, not route validation: The FDA package describes DICOM viewing and analysis, automated segmentation, editing and pre-printing, integration with 3D Systems printers and software packages, and virtual-reality visualization for non-diagnostic use.
Scope: Exact FDA submission. This is the interface boundary described in the exact FDA package; current local compatibility, routing, identity, archive, and downtime behavior still require acceptance testing.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Source context, not route validation: The FDA package describes a stand-alone modular software package with output to qualified 3D-printing technologies and materials; current hosting, hardware, printer, service, security, and local configuration remain unverified.
Scope: Exact FDA submission. This is the release-level boundary described in the exact FDA package; current hosting, contracts, security, availability, upgrade path, and site configuration remain separate evidence.
Confirm processing location, queue ownership, exclusions and exact AI build.
Unresolved route: Segmented output files, 3D visualizations, physical anatomic replicas, and non-diagnostic VR views.
Scope: Exact FDA submission.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
The FDA package describes a stand-alone modular software package with output to qualified 3D-printing technologies and materials; current hosting, hardware, printer, service, security, and local configuration remain unverified.
Exact FDA submission
Checked 2026-09-04
This is the release-level boundary described in the exact FDA package; current hosting, contracts, security, availability, upgrade path, and site configuration remain separate evidence.
The FDA package describes DICOM viewing and analysis, automated segmentation, editing and pre-printing, integration with 3D Systems printers and software packages, and virtual-reality visualization for non-diagnostic use.
Exact FDA submission
Checked 2026-09-04
This is the interface boundary described in the exact FDA package; current local compatibility, routing, identity, archive, and downtime behavior still require acceptance testing.
Only surgeons, radiologists, and other trained healthcare professionals should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
Monitor K183489 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Physical replicas are adjunctive and require other diagnostic tools and expert clinical judgment; application, printer, material, anatomy, and output-validation limits apply.
Exact FDA submission
Checked 2026-09-04
This is buyer-facing monitoring guidance derived from the exact cleared or authorized scope; it is not a claim that the FDA package provides a live monitoring service.
The FDA package describes a stand-alone modular software package with output to qualified 3D-printing technologies and materials; current hosting, hardware, printer, service, security, and local configuration remain unverified.
exact submission · checked 2026-09-04
FDA-described deployment boundaries do not establish current hosting contracts, security controls, uptime, or local qualification.
The FDA package describes DICOM viewing and analysis, automated segmentation, editing and pre-printing, integration with 3D Systems printers and software packages, and virtual-reality visualization for non-diagnostic use.
exact submission · checked 2026-09-04
Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.
Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Physical replicas are adjunctive and require other diagnostic tools and expert clinical judgment; application, printer, material, anatomy, and output-validation limits apply.
exact submission · checked 2026-09-04
Define stop-use, escalation, rollback, and revalidation triggers before production use.
Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.
product family · checked 2026-09-04
No exact-product price was found in the reviewed public material; productivity or value claims require local evidence.
Monitor K183489 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Physical replicas are adjunctive and require other diagnostic tools and expert clinical judgment; application, printer, material, anatomy, and output-validation limits apply.
exact submission · checked 2026-09-04
Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. DICOM image-segmentation and surgical-planning software limits remain submission-specific.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Human reviewedStatus
2026-09-04Last searched
26Fields reviewed
12Source classes checked
20Screened PubMed leads
2Screened trial leads
0Unreviewed FDA recall leads
Twenty PubMed records and two trial records were screened. Two publications evaluate DICOM to PRINT or D2P segmentation without a tested version or K183489 linkage; the remaining records use D2P for unrelated terms, so all 22 records are quarantined. The current discovery snapshot contains 20 PubMed and 2 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, acr ai central product, fda decision material, fda pma record, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not publicly available - checked 2026-09-04 - Exact FDA submission
No public list price for the exact cleared or authorized configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.
Pricing model: Not yet reviewed
Setup and integration fees: Not yet reviewed
Recurring fees: Not yet reviewed
Minimum commitment: Not yet reviewed
Contract duration: Not yet reviewed
Renewal terms: Not yet reviewed
Cancellation terms: Not yet reviewed
Support and service terms: Not yet reviewed
Trial or pilot terms: Not yet reviewed
Data hosting location: Not yet reviewed
Data retention: Not yet reviewed
Documented data use: Not yet reviewed
Hardware requirements: Not yet reviewed
Network requirements: Not yet reviewed
Identity and access: Not yet reviewed
Security: Not publicly available - checked 2026-09-04 - Exact FDA submission
Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.
Authorized purpose: ADAS 3D supports visualization and analysis of MR and CT images of the heart for individual patients with cardiovascular disease, including patients with myocardial scar from ischemic or non-ischemic heart disease. Qualified professionals may use it for cardiac calculation, quantification, visualization, pre-planning, and electrophysiology procedures; produced data support clinical decision-making and must not be used as irrefutable medical advice or for radiotherapy planning, dose calculation, or treatment delivery.
Modality
Radiological images; modality varies or is not specified in FDA list
Authorized purpose: Cercare Neurosuite with Capillary Function and Oncology Virtual Expert processes CT or MR information to provide perfusion, tissue-blood-volume, diffusion, labeling, and quantification support for adult brain glioma workflows.
Modality
Radiological images; modality varies or is not specified in FDA list
Synapse PACS 7.6.0 provides medical-image management and processing, including measurements, regional segmentation, clinical calculations, structured reports, comparison, fusion, and volume rendering to aid reading, interpretation, reporting, and treatment planning.
Modality
Radiological images; modality varies or is not specified in FDA list
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.
Accepted repository change history
No entry has been added to this scoped change history for this record. This is not proof that the record never changed.