Authorized purpose: Automatically labels, visualizes, and quantifies the volume and shape of segmentable brain structures from MR images.
Clinical role
Automatically segments, labels and quantifies volume and shape of brain structures on T1-weighted MR images, showing color-coded overlays or 3-D meshes and exporting a PDF report comparing volumes to normative values.
Images and patients
MRI · Brain
Key limitation
Keep in mind: Limited to segmentable brain structures from T1-weighted brain MRI; additional exact FDA limitations are unknown.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: Automatically labels, visualizes, and quantifies the volume and shape of segmentable brain structures from MR images.
Which images or clinical context are in scope?Modality: MRI Anatomy or clinical context: Brain Input: DICOM T1-weighted brain MR scans.
Who uses it?Intended users: Medical professionals such as radiologists, neurologists, and neuroradiologists.
What does the software produce?Output: Automated labels, color-coded overlays or 3D meshes, volume and shape measurements, DICOM results, and PDF reports with normative comparisons.
What is its most important limitation?Key limitation: Limited to segmentable brain structures from T1-weighted brain MRI; additional exact FDA limitations are unknown.
Clinical takeaway. Clinical role: Automatically labels, visualizes, and quantifies the volume and shape of segmentable brain structures from MR images.
Evidence takeaway. Evidence from the reviewed source record: FDA summary reports validation in 101 subjects across 21 structures, with Dice and relative-volume acceptance criteria; third-ventricle mean Dice was below 0.70, and repeat-scan reproducibility was under 10% in the 20Repeat set.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
Automatically labels, visualizes, and quantifies the volume and shape of segmentable brain structures from MR images.
Exact FDA submission
Checked 2026-09-19
Clinical-use display mirrors the exact FDA-supported purpose reconciled in this remediation finding.
Automatically segments, labels and quantifies volume and shape of brain structures on T1-weighted MR images, showing color-coded overlays or 3-D meshes and exporting a PDF report comparing volumes to normative values.
Exact FDA submission
Checked 2026-11-14
Packet wording: "The ClearPoint MaestroTM Brain Model provides automated image processing of brain structures from T1-weighted MR images. Specifically, the device automates the manual process of identifying, labeling, and quantifying the volume and shape of subcortical and cortical structures to simplify the workflow for MRI segmentation." … "The segmented brain structures are color coded and overlayed onto the MR images or be displayed as 3-D triangular mesh representation.". Read from the exact FDA decision document on 2026-11-14.
FDA summary reports validation in 101 subjects across 21 structures, with Dice and relative-volume acceptance criteria; third-ventricle mean Dice was below 0.70, and repeat-scan reproducibility was under 10% in the 20Repeat set.
Exact FDA submission
Checked 2026-09-19
Submission-reported segmentation validation; not a general clinical-outcomes claim.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
FDA summary reports validation in 101 subjects across 21 structures, with Dice and relative-volume acceptance criteria; third-ventricle mean Dice was below 0.70, and repeat-scan reproducibility was under 10% in the 20Repeat set.
FDA source
Exact FDA submission
Checked 2026-09-19
Submission-reported segmentation validation; not a general clinical-outcomes claim.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
ClearPoint Maestro Brain Model regulatory performance study summary
DesignStand-Alone Performance
PopulationPopulation described in the exact-submission ACR model card
Scopeexact submission
Tested versionExact product version not reported; regulatory study summarized for FDA submission K213645
Samplen=101
IndependenceIndependence not established
K213645
ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K213645. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: DICOM T1-weighted brain MR scans.
Scope: Exact FDA submission. FDA device description and comparison table.
Confirm scanner/model, protocol, series completeness and both software builds.
Source context, not route validation: The exact-submission ACR model card explicitly references DICOM in the described input, output, or workflow; exact interfaces, versions, and local compatibility require confirmation.
Scope: Exact FDA submission. Interoperability terms explicitly reported in the exact-submission ACR model card; interface versions and local compatibility still require confirmation.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Unresolved route: Not established for the exact route.
Scope: Exact FDA submission. The reviewed sources do not establish the current exact-release hosting topology, data flow, region, tenancy, or disaster-recovery configuration.
Confirm processing location, queue ownership, exclusions and exact AI build.
Field checked 2026-09-09 ·
Returned objects
Source context, not route validation: Automated labels, color-coded overlays or 3D meshes, volume and shape measurements, DICOM results, and PDF reports with normative comparisons.
Scope: Exact FDA submission. FDA device description.
Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.
Unresolved route: Automated labels, color-coded overlays or 3D meshes, volume and shape measurements, DICOM results, and PDF reports with normative comparisons.
Scope: Exact FDA submission. FDA device description.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
The exact-submission ACR model card explicitly references DICOM in the described input, output, or workflow; exact interfaces, versions, and local compatibility require confirmation.
Exact FDA submission
Checked 2026-09-01
Interoperability terms explicitly reported in the exact-submission ACR model card; interface versions and local compatibility still require confirmation.
ACR AI Central provides an exact-submission model card for ClearPoint Maestro Brain Model; FDA labeling remains controlling for clinical use, limitations, and performance.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Exact FDA record reviewedStatus
2026-09-20Last searched
24Fields reviewed
10Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, fda decision summary packet, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Care setting: Not publicly available - checked 2026-11-14 - Exact FDA submission
The exact FDA decision document was reviewed on 2026-11-14 and does not state a care setting or patient population for the subject device. Confirm intended sites and patients with the vendor and local policy.
Public pricing: Not verified - checked 2026-09-09 - Exact FDA submission
No public price or quote for the exact K213645 configuration was established in this review.
Pricing model: Not yet reviewed
Setup and integration fees: Not yet reviewed
Recurring fees: Not yet reviewed
Minimum commitment: Not yet reviewed
Contract duration: Not yet reviewed
Renewal terms: Not yet reviewed
Cancellation terms: Not yet reviewed
Support and service terms: Not yet reviewed
Trial or pilot terms: Not yet reviewed
Reimbursement: Not publicly available - checked 2026-09-09 - Exact FDA submission
No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.
Deployment: Not verified - checked 2026-09-09 - Exact FDA submission
The reviewed sources do not establish the current exact-release hosting topology, data flow, region, tenancy, or disaster-recovery configuration.
Data hosting location: Not yet reviewed
Data retention: Not yet reviewed
Documented data use: Not yet reviewed
Hardware requirements: Not yet reviewed
Network requirements: Not yet reviewed
Identity and access: Not yet reviewed
Security: Not verified - checked 2026-09-09 - Exact FDA submission
Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.
Installation timeline: Not yet reviewed
Training: Not verified - checked 2026-09-09 - Exact FDA submission
Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.
Monitoring: Not verified - checked 2026-09-09 - Exact FDA submission
Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.
Authorized purpose: ANDI 2.2 processes brain MR images for atlas-assisted visualization, segmentation and volumetric quantification, presenting adjunctive DICOM-readable information for trained medical professionals; it does not predict, diagnose or interpret brain health.
Authorized purpose: Morph automatically labels, visualizes, and volumetrically quantifies segmentable brain structures from MR images and may compare measurements with reference percentile data; it is adjunctive and not the sole basis for final decisions.
Authorized purpose: Brain MRI morphometry and white matter hyperintensity quantification, including normative comparison and longitudinal change assessment for clinician review.
Authorized purpose: Adjunctive quantitative brain MRI analysis that compares patient measures with normative and longitudinal references; a physician reviews all results and makes the clinical decision.
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.
Accepted repository change history
No entry has been added to this scoped change history for this record. This is not proof that the record never changed.