FDA-cleared profile

ClariPulmo

ClariPi Inc. · CT/CTA · FDA-cleared (510(k))

Authorized purpose: Non-invasive CT image analysis that quantifies lung images and supports physician review of pulmonary tissue images and abnormalities.

Clinical role
Image analysis software that automatically segments the lungs on thoracic CT and quantifies low- and high-attenuation areas using predefined Hounsfield unit thresholds, with color overlay reports supporting physician diagnosis and documentation; not intended for diagnosing pneumonia or COVID-19.
Images and patients
CT/CTA · Chest/Lung; lungs/pulmonary tissue
Key limitation
Keep in mind: Not intended to diagnose pneumonia or COVID-19; intended population is adults receiving lung CT, with physician interpretation required.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: Non-invasive CT image analysis that quantifies lung images and supports physician review of pulmonary tissue images and abnormalities.
  • Which images or clinical context are in scope?Modality: CT/CTA Anatomy or clinical context: Chest/Lung; lungs/pulmonary tissue Input: Adult thoracic lung CT datasets.
  • Who uses it?Intended users: Physicians.
  • What does the software produce?Output: Automated lung segmentation, low/high attenuation quantification using predefined HU thresholds, color overlays, summary tables and charts, and optional denoised or kernel-normalized images.
  • What is its most important limitation?Key limitation: Not intended to diagnose pneumonia or COVID-19; intended population is adults receiving lung CT, with physician interpretation required.

Clinical takeaway. Clinical role: Non-invasive CT image analysis that quantifies lung images and supports physician review of pulmonary tissue images and abnormalities.

Evidence takeaway. Evidence from the reviewed source record: FDA-reported nonclinical verification and validation covered attenuation analysis, lung segmentation, denoising, and kernel normalization using internal and two external datasets; no clinical study was required for substantial equivalence.

Evidence context

Evidence context not reviewed for this FDA record.

Public-label constraints

Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.

Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.

Purpose
Non-invasive CT image analysis that quantifies lung images and supports physician review of pulmonary tissue images and abnormalities.
  • Exact FDA submission
  • Checked 2026-09-19

Clinical-use display mirrors the exact FDA-supported purpose reconciled in this remediation finding.

Workflow role
Image analysis software that automatically segments the lungs on thoracic CT and quantifies low- and high-attenuation areas using predefined Hounsfield unit thresholds, with color overlay reports supporting physician diagnosis and documentation; not intended for diagnosing pneumonia or COVID-19.
  • Exact FDA submission
  • Checked 2026-11-14

Packet wording: "The software is designed to support the physician in the diagnosis and documentation of pulmonary tissue images (e.g., abnormalities) from the CT thoracic datasets. (The software is not intended for the diagnosis of pneumonia or COVID-19)." … "The software provides automated segmentation of the lungs and quantification of low-attenuation and high-attenuation areas within the segmented lungs by using predefined Hounsfield unit thresholds.". Read from the exact FDA decision document on 2026-11-14.

Required input
Adult thoracic lung CT datasets.
  • Exact FDA submission
  • Checked 2026-09-19

FDA packet input and population scope.

Output and clinician action
Automated lung segmentation, low/high attenuation quantification using predefined HU thresholds, color overlays, summary tables and charts, and optional denoised or kernel-normalized images.
  • Exact FDA submission
  • Checked 2026-09-19

FDA device-description output scope.

Care setting
Adult
  • Exact FDA submission
  • Checked 2026-09-01

Patient-population text reported in the exact-submission ACR model card; institutional setting may still require vendor confirmation.

Intended users
Physicians.
  • Exact FDA submission
  • Checked 2026-09-19

FDA packet supports physician use.

Limitations
Not intended to diagnose pneumonia or COVID-19; intended population is adults receiving lung CT, with physician interpretation required.
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA limitation boundary.

Evidence summary
FDA-reported nonclinical verification and validation covered attenuation analysis, lung segmentation, denoising, and kernel normalization using internal and two external datasets; no clinical study was required for substantial equivalence.
  • Exact FDA submission
  • Checked 2026-09-19

Submission evidence only; not a clinical-performance claim.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
FDA-reported nonclinical verification and validation covered attenuation analysis, lung segmentation, denoising, and kernel normalization using internal and two external datasets; no clinical study was required for substantial equivalence.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Submission evidence only; not a clinical-performance claim.

Reported sensitivity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

Reported specificity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

ClariPulmo regulatory performance study summary

  • DesignStand-Alone Performance
  • PopulationAdult
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K203783
  • IndependenceIndependence not established

K203783

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K203783. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-09
Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not verified: Pricing information is recorded but not verified against a resolving source.

Reimbursement

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Integration chain: evidence at each hop

Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.

Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner

No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.

  1. Acquisition/source

    Source context, not route validation: Adult thoracic lung CT datasets.

    Scope: Exact FDA submission. FDA packet input and population scope.

    Confirm scanner/model, protocol, series completeness and both software builds.

    Field checked 2026-09-19 · FDA K203783: ClariPulmo

  2. Routing

    Unresolved route: Not established for the exact route.

    Scope: Exact FDA submission. The exact FDA decision document was reviewed on 2026-11-14 and does not state the subject device's DICOM, PACS, RIS, or EHR interfaces. Confirm supported DICOM services and PACS, RIS, EHR, and interface versions with the vendor.

    Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.

    Field checked 2026-11-14 ·

  3. AI processing

    Unresolved route: Not established for the exact route.

    Scope: Exact FDA submission. The exact FDA decision document was reviewed on 2026-11-14 and does not state where the subject device's software runs or how it is hosted. Confirm the deployment model, hosting, data flow, and current commercial configuration with the vendor.

    Confirm processing location, queue ownership, exclusions and exact AI build.

    Field checked 2026-11-14 ·

  4. Returned objects

    Source context, not route validation: Automated lung segmentation, low/high attenuation quantification using predefined HU thresholds, color overlays, summary tables and charts, and optional denoised or kernel-normalized images.

    Scope: Exact FDA submission. FDA device-description output scope.

    Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.

    Field checked 2026-09-19 · FDA K203783: ClariPulmo

  5. Viewer/worklist/report

    Unresolved route: Automated lung segmentation, low/high attenuation quantification using predefined HU thresholds, color overlays, summary tables and charts, and optional denoised or kernel-normalized images.

    Scope: Exact FDA submission. FDA device-description output scope.

    Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.

    Field checked 2026-09-19 · FDA K203783: ClariPulmo

  6. Support owner

    Unresolved route: Not established for the exact route.

    Scope: Scope not established.

    Name the local, platform and AI supplier support owners and escalation path.

Planned local route tests

No tests below have been performed or accepted.

  1. Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
  2. Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
  3. Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
  4. Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
  5. Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
  6. Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
  7. Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Supported acquisition environment

Scanner manufacturers: All

exact submission · checked 2026-09-01

Confirm currently supported software versions, protocols, and site-specific acquisition configurations.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for ClariPulmo; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-09
Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K203783
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
CT/CTA
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

FDA specifies CT; CT/CTA is the controlled catalog enum and CTA is not implied.

Anatomy
Chest/Lung; lungs/pulmonary tissue
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA thoracic lung scope.

Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2022-04-06
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
25Fields reviewed
10Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, fda decision summary packet, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not verified - checked 2026-09-09 - Exact FDA submission

    No public price or quote for the exact K203783 configuration was established in this review.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Reimbursement: Not publicly available - checked 2026-09-09 - Exact FDA submission

    No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

  • Deployment: Not publicly available - checked 2026-11-14 - Exact FDA submission

    The exact FDA decision document was reviewed on 2026-11-14 and does not state where the subject device's software runs or how it is hosted. Confirm the deployment model, hosting, data flow, and current commercial configuration with the vendor.

  • Integration: Not publicly available - checked 2026-11-14 - Exact FDA submission

    The exact FDA decision document was reviewed on 2026-11-14 and does not state the subject device's DICOM, PACS, RIS, or EHR interfaces. Confirm supported DICOM services and PACS, RIS, EHR, and interface versions with the vendor.

  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Security: Not verified - checked 2026-09-09 - Exact FDA submission

    Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.

  • Installation timeline: Not yet reviewed
  • Training: Not verified - checked 2026-09-09 - Exact FDA submission

    Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

  • Monitoring: Not verified - checked 2026-09-09 - Exact FDA submission

    Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

7 sources - 3 history items

Source ledger

  1. FDA K203783: ClariPulmo

    U.S. Food and Drug Administration - Published 2022-04-06 - Accessed 2026-09-09

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation, purpose, intendedUsers, input, output, limitations, evidenceSummary

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. ClariPulmo model card (K203783)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, limitations, evidenceSummary, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K203783

    U.S. Food and Drug Administration - Accessed 2026-09-09

    Scope: postmarketSafety

Research history

  1. K203783 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K260406

Brainomix 360 Hyperdensity

Brainomix Limited

Authorized purpose: Provides volumetric hyperdensity segmentation and quantification as an aid; physicians must review original images.

Modality
CT/CTA
Anatomy
Brain
Decision
2026-06-12
FDA-cleared K260322

Acorn 3D Software (AC-SEG-4009); Acorn 3DP Model (AC-101-XX)

Mighty Oak Medical

Authorized purpose: Acorn 3D Software and Acorn 3DP Model process CT, synthetic CT, and CT angiography for segmentation, visualization, measurement, treatment planning, and three-dimensional printing, including adult and pediatric spine trajectory planning.

Modality
CT/CTA
Anatomy
Spine; craniomaxillofacial; musculoskeletal anatomy; and patient-specific printed models
Decision
2026-06-11

Correct this exact record

Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.

Open existing correction destination

Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.

Accepted repository change history

No entry has been added to this scoped change history for this record. This is not proof that the record never changed.

Correction policy and limits · Public repository change history