Authorized purpose: Auto Positioning uses 3D patient-position information and the selected CT protocol to suggest scout range, orientation, table height, and collision checks before a technologist-approved first scout.
Clinical role
CT acquisition, reconstruction, visualization, and release-specific examination planning.
Images and patients
CT/CTA · Multi-region
Key limitation
Keep in mind: Incorporation of this optional feature does not preclude the technologist from performing the existing manual workflow on the CT system, if desired.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: Auto Positioning uses 3D patient-position information and the selected CT protocol to suggest scout range, orientation, table height, and collision checks before a technologist-approved first scout.
Which images or clinical context are in scope?Modality: CT/CTA Anatomy or clinical context: Multi-region Input: 3D spatial information of the individual patient on the table and information from the selected protocol.
Who uses it?Intended users: CT technologist.
What does the software produce?Output: Auto Positioning acquires 3D spatial information of the individual patient on the table and combines it with information from the selected protocol to automatically calculate and visually display the scout's start and end locations. Concurrently it checks for proper patient orientation, determines the table height for optimum patient centering, and checks for potential contact between the patient and the gantry. Upon acceptance by the technologist the patient is automatically moved to the correct scout start location.
What is its most important limitation?Key limitation: Incorporation of this optional feature does not preclude the technologist from performing the existing manual workflow on the CT system, if desired.
Clinical takeaway. Clinical role: Auto Positioning uses 3D patient-position information and the selected CT protocol to suggest scout range, orientation, table height, and collision checks before a technologist-approved first scout.
Evidence takeaway. Evidence from the reviewed source record: K192956 reports manufacturer camera, positioning, software, workflow, and safety validation for supported patient orientations and protocols.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
Auto Positioning uses 3D patient-position information and the selected CT protocol to suggest scout range, orientation, table height, and collision checks before a technologist-approved first scout.
Exact FDA submission
Checked 2026-09-19
Clinical-use display mirrors the exact FDA-supported purpose reconciled from the decision-summary packet.
Auto Positioning acquires 3D spatial information of the individual patient on the table and combines it with information from the selected protocol to automatically calculate and visually display the scout's start and end locations. Concurrently it checks for proper patient orientation, determines the table height for optimum patient centering, and checks for potential contact between the patient and the gantry. Upon acceptance by the technologist the patient is automatically moved to the correct scout start location.
Exact FDA submission
Checked 2026-09-19
The packet's output is the scout setup and safety-check workflow, not cross-sectional images, measurements, or planning datasets.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
K192956 reports manufacturer camera, positioning, software, workflow, and safety validation for supported patient orientations and protocols.
Sensitivity is not the primary endpoint for this acquisition, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.
Reported specificity
Not applicable
Not applicable
Not applicable
Checked 2026-09-03
Specificity is not the primary endpoint for this acquisition, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
K192956 FDA performance package for Auto Positioning
DesignManufacturer positioning-accuracy, software, workflow, and safety validation
PopulationSupported CT patient-positioning and scout-planning workflows
Scopeexact submission
Tested versionAuto Positioning release cleared in K192956
IndependenceSponsor or vendor study
K192956
Manufacturer evidence submitted for clearance or approval; local clinical, technical, and workflow acceptance remains necessary.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K192956 by exact submission identifier. This does not exclude product-family or later safety information.
Public source
Exact FDA submission
Checked 2026-09-03
Exact-identifier review does not prove absence of all product-family safety signals.
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging, procedural, or surgical service line and verify payer, site-of-service, and coding assumptions independently.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: 3D spatial information of the individual patient on the table and information from the selected protocol.
Scope: Exact FDA submission. The packet describes camera-derived 3D spatial information combined with the selected protocol, not X-ray projection data.
Confirm scanner/model, protocol, series completeness and both software builds.
Source context, not route validation: Scanner-console and DICOM workflow to PACS, advanced visualization, and treatment-planning systems; validate worklist, protocol, dose record, geometry, series labels, and routing.
Scope: Exact FDA submission.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Source context, not route validation: Auto Positioning acquires 3D spatial information of the individual patient on the table and combines it with information from the selected protocol to automatically calculate and visually display the scout's start and end locations. Concurrently it checks for proper patient orientation, determines the table height for optimum patient centering, and checks for potential contact between the patient and the gantry. Upon acceptance by the technologist the patient is automatically moved to the correct scout start location.
Scope: Exact FDA submission. The packet's output is the scout setup and safety-check workflow, not cross-sectional images, measurements, or planning datasets.
Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.
Unresolved route: Auto Positioning acquires 3D spatial information of the individual patient on the table and combines it with information from the selected protocol to automatically calculate and visually display the scout's start and end locations. Concurrently it checks for proper patient orientation, determines the table height for optimum patient centering, and checks for potential contact between the patient and the gantry. Upon acceptance by the technologist the patient is automatically moved to the correct scout start location.
Scope: Exact FDA submission. The packet's output is the scout setup and safety-check workflow, not cross-sectional images, measurements, or planning datasets.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
Scanner-console and DICOM workflow to PACS, advanced visualization, and treatment-planning systems; validate worklist, protocol, dose record, geometry, series labels, and routing.
GE HealthCare's Product Security Portal is described as a credentialed global portal giving customers security updates, vulnerability status and required actions for GE HealthCare products potentially impacted by critical vulnerabilities, plus historical patch views; GE HealthCare references its Cybersecurity Risk Management System and publishes a Coordinated Vulnerability Disclosure Statement for reporting vulnerabilities in GE HealthCare products.
Vendor-wide
Checked 2026-10-26
Vendor statement, not an attestation. Identity bridge: GE HealthCare's own offices page on gehealthcare.com names Ge Hangwei Medical Systems Co., Ltd (Beijing). Corporate program statement; portal content is credentialed; exact-release scope is not established.
Only qualified ct technologists and trained interpreting or treating physicians. should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track these product-specific limits: The technologist must review and accept suggested motion and scan setup; the feature does not replace positioning, collision, protocol, or patient-safety checks.
Scanner-console and DICOM workflow to PACS, advanced visualization, and treatment-planning systems; validate worklist, protocol, dose record, geometry, series labels, and routing.
exact submission · checked 2026-09-03
Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.
Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. The technologist must review and accept suggested motion and scan setup; the feature does not replace positioning, collision, protocol, or patient-safety checks.
exact submission · checked 2026-09-03
Define stop-use, escalation, rollback, and revalidation triggers before production use.
Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.
product family · checked 2026-09-03
Vendor productivity claims are scenario inputs, not guaranteed local savings.
Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track these product-specific limits: The technologist must review and accept suggested motion and scan setup; the feature does not replace positioning, collision, protocol, or patient-safety checks.
exact submission · checked 2026-09-03
Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
The classification name is a computed tomography x-ray system; the packet does not state CTA as the modality label. Controlled directory display reconciled from exact FDA wording: Computed tomography x-ray system. -> CT/CTA.
This audit distinguishes completed source review from fields that have not yet been researched.
Exact FDA record reviewedStatus
2026-09-20Last searched
27Fields reviewed
11Source classes checked
14Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains 14 PubMed and 0 ClinicalTrials.gov candidate records. Discovery metadata alone does not establish exact tested-product, version, population, and endpoint applicability, so all candidates remain quarantined and none was promoted. The current discovery snapshot contains 14 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, fda decision material, reimbursement policy, fda device recall, fda decision summary packet, literature index, trial registry, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not publicly available - checked 2026-09-03 - Exact FDA submission
No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, enabled features, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.
Authorized purpose: NAEOTOM Alpha.Peak, Alpha, Alpha.Pro, and Alpha.Prime are photon-counting CT systems for diagnostic cross-sectional imaging, treatment and interventional support, low-dose lung-cancer screening under approved protocols, and radiation-therapy planning.
CT 5300 acquires head and whole-body cross-sectional CT images for diagnosis in adult and pediatric patients and supports low-dose lung-cancer screening under approved protocols.
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.