Authorized purpose: The Aquilion Serve acquires and displays cross-sectional volumes of the whole body including the head, provides volume sets for specialized studies using indicated software or hardware, and includes AiCE deep-convolutional noise reduction for abdomen, pelvis, lung, cardiac, extremity, head, and inner-ear applications.
Clinical role
X-ray CT acquisition and reconstruction, with release-specific image-quality, spectral, or deep-learning processing functions where labeled.
Images and patients
CT/CTA · Whole body including head; abdomen, pelvis, lung, cardiac, extremity, and inner-ear applications for AiCE
Key limitation
Keep in mind: Use only the exact labeled scanner, software, dose, reconstruction, patient population, anatomy, and screening criteria; local dose and image-quality decisions remain clinical responsibilities.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: The Aquilion Serve acquires and displays cross-sectional volumes of the whole body including the head, provides volume sets for specialized studies using indicated software or hardware, and includes AiCE deep-convolutional noise reduction for abdomen, pelvis, lung, cardiac, extremity, head, and inner-ear applications.
Which images or clinical context are in scope?Modality: CT/CTA Anatomy or clinical context: Whole body including head; abdomen, pelvis, lung, cardiac, extremity, and inner-ear applications for AiCE Input: X-ray transmission data acquired with the exact labeled CT system, protocols, software, and hardware.
Who uses it?Intended users: Qualified radiologic technologists and physicians or other trained healthcare professionals within the labeled workflow.
What does the software produce?Output: Cross-sectional CT images, volume sets, and labeled reconstructed or processed image outputs for professional review.
What is its most important limitation?Key limitation: Use only the exact labeled scanner, software, dose, reconstruction, patient population, anatomy, and screening criteria; local dose and image-quality decisions remain clinical responsibilities.
Clinical takeaway. Clinical role: The Aquilion Serve acquires and displays cross-sectional volumes of the whole body including the head, provides volume sets for specialized studies using indicated software or hardware, and includes AiCE deep-convolutional noise reduction for abdomen, pelvis, lung, cardiac, extremity, head, and inner-ear applications.
Evidence takeaway. Evidence from the reviewed source record: The exact FDA document reports bench testing: an image quality metric evaluation against the predicate concluded equivalent or improved performance; an AiCE image quality evaluation showed equivalent or improved performance relative to FBP and AIDR 3D; and a phantom comparison study conducted to support a claim of twice the high contrast spatial resolution for AiCE Body at 10% MTF compared to AIDR reconstruction. In clinical image review, a board-certified radiologist confirmed representative subject-device images were of diagnostic quality. Results are as reported in the FDA submission, not independent validation.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
The Aquilion Serve acquires and displays cross-sectional volumes of the whole body including the head, provides volume sets for specialized studies using indicated software or hardware, and includes AiCE deep-convolutional noise reduction for abdomen, pelvis, lung, cardiac, extremity, head, and inner-ear applications.
Use only the exact labeled scanner, software, dose, reconstruction, patient population, anatomy, and screening criteria; local dose and image-quality decisions remain clinical responsibilities.
The exact FDA document reports bench testing: an image quality metric evaluation against the predicate concluded equivalent or improved performance; an AiCE image quality evaluation showed equivalent or improved performance relative to FBP and AIDR 3D; and a phantom comparison study conducted to support a claim of twice the high contrast spatial resolution for AiCE Body at 10% MTF compared to AIDR reconstruction. In clinical image review, a board-certified radiologist confirmed representative subject-device images were of diagnostic quality. Results are as reported in the FDA submission, not independent validation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "Performance Testing - Bench Objective IQ Performance Comparison An image quality metric evaluation comparing performance of the TSX-307A (Aquilion Serve) system relative to the predicate device was conducted." … "It was concluded that the subject device demonstrated equivalent or improved performance, compared to the predicate device, as demonstrated by the results of the above testing.". Read from the exact FDA decision document on 2026-11-14.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
The exact FDA document reports bench testing: an image quality metric evaluation against the predicate concluded equivalent or improved performance; an AiCE image quality evaluation showed equivalent or improved performance relative to FBP and AIDR 3D; and a phantom comparison study conducted to support a claim of twice the high contrast spatial resolution for AiCE Body at 10% MTF compared to AIDR reconstruction. In clinical image review, a board-certified radiologist confirmed representative subject-device images were of diagnostic quality. Results are as reported in the FDA submission, not independent validation.
FDA source
Exact FDA submission
Checked 2026-11-14
Packet wording: "Performance Testing - Bench Objective IQ Performance Comparison An image quality metric evaluation comparing performance of the TSX-307A (Aquilion Serve) system relative to the predicate device was conducted." … "It was concluded that the subject device demonstrated equivalent or improved performance, compared to the predicate device, as demonstrated by the results of the above testing.". Read from the exact FDA decision document on 2026-11-14.
No exact-submission and version-linked sensitivity endpoint is promoted; candidate or product-family values were not substituted.
Reported specificity
Not established in reviewed sources
Reviewed sources checked
Exact FDA submission
Checked 2026-09-04
No exact-submission and version-linked specificity endpoint is promoted; candidate or product-family values were not substituted.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
FDA submission-specific verification and validation evidence for K222819
DesignFDA 510(k) regulatory decision material and submission-specific verification or validation evidence for K222819; this is regulatory evidence, not an independent clinical effectiveness study.
PopulationThe labeled population, anatomy, images, and test materials described in the exact FDA decision material for K222819; a broader clinical population is not established here.
Scopeexact submission
Tested versionAquilion Serve (TSX-307A/1) V1.2 with AiCE-i (K222819)
IndependenceSponsor or vendor study
K222819
The exact FDA material is submission-specific and should not be generalized to other versions, enabled options, sites, protocols, or clinical outcomes.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
Not established in reviewed sources
Reviewed sources checked
Exact FDA submission
Checked 2026-09-04
No native submission-identifier match was found in the checked openFDA radiology recall snapshot. Product-name, manufacturer, and adverse-event searches remain necessary; this does not establish that no recall or safety report exists.
Pricing & contractsPublic price, contract terms, and reimbursement
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: X-ray transmission data acquired with the exact labeled CT system, protocols, software, and hardware.
Scope: Exact FDA submission.
Confirm scanner/model, protocol, series completeness and both software builds.
Unresolved route: Not established for the exact route.
Scope: Exact FDA submission. Exact FDA evidence does not establish the complete local interface, identity, routing, display, report, fallback, or interoperability behavior for this submission.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Field checked 2026-09-04 ·
AI processing
Unresolved route: Not established for the exact route.
Scope: Exact FDA submission. Exact deployment topology, hosting, network boundaries, PHI flow, retention, update mechanism, and local operational requirements are not established for this submission.
Confirm processing location, queue ownership, exclusions and exact AI build.
Field checked 2026-09-04 ·
Returned objects
Source context, not route validation: Cross-sectional CT images, volume sets, and labeled reconstructed or processed image outputs for professional review.
Scope: Exact FDA submission.
Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.
Unresolved route: Cross-sectional CT images, volume sets, and labeled reconstructed or processed image outputs for professional review.
Scope: Exact FDA submission.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
Canon Medical's cybersecurity page states that Canon imaging systems follow a software development life cycle governed by the NIST 800-53 Risk Management Framework adhering to Defense Health Agency certification requirements; FIPS 140-2 certified cipher standards; deny-all firewall configuration and near-real-time SIEM syslog export as premium offerings; a dedicated cybersecurity risk management team for threat intelligence monitoring, advisories and incident response; and published OS and third-party update policies. Vendor statement; no certification or test report was read.
Vendor-wide
Checked 2026-10-26
Vendor statement, not an attestation. Identity bridge: the page is on us.medical.canon, Canon Medical's own site (Canon Medical Systems USA), and describes Canon Medical imaging devices generally; the FDA applicant is Canon Medical Systems Corporation, named in the corporate name-change press release on the same domain. Regional page with general device claims; exact-release scope is not established.
Current vendor material is product-family context for K222819 only; exact release, enabled options, deployment, security, interoperability, pricing, and local workflow require confirmation.
product family · checked 2026-09-04
Vendor material is not exact-release evidence unless that linkage is separately proven.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Exact FDA record reviewedStatus
2026-09-20Last searched
31Fields reviewed
13Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, vendor product page, implementation documentation, reimbursement policy, peer reviewed literature, clinical trial registry, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not publicly available - checked 2026-09-04 - Exact FDA submission
No public exact-release price, license fee, subscription, contract, or implementation cost was established in this review.
Pricing model: Not yet reviewed
Setup and integration fees: Not yet reviewed
Recurring fees: Not yet reviewed
Minimum commitment: Not yet reviewed
Contract duration: Not yet reviewed
Renewal terms: Not yet reviewed
Cancellation terms: Not yet reviewed
Support and service terms: Not yet reviewed
Trial or pilot terms: Not yet reviewed
Reimbursement: Not applicable - checked 2026-09-04 - Not applicable
CMS coverage-index context was checked, but no product-specific coverage or reimbursement determination is inferred.
Deployment: Not publicly available - checked 2026-09-04 - Exact FDA submission
Exact deployment topology, hosting, network boundaries, PHI flow, retention, update mechanism, and local operational requirements are not established for this submission.
Integration: Not publicly available - checked 2026-09-04 - Exact FDA submission
Exact FDA evidence does not establish the complete local interface, identity, routing, display, report, fallback, or interoperability behavior for this submission.
Data hosting location: Not yet reviewed
Data retention: Not yet reviewed
Documented data use: Not yet reviewed
Hardware requirements: Not yet reviewed
Network requirements: Not yet reviewed
Identity and access: Not yet reviewed
Installation timeline: Not yet reviewed
Training: Not verified - checked 2026-10-27 - Exact FDA submission
Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.
Monitoring: Not verified - checked 2026-10-27 - Exact FDA submission
Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.
Authorized purpose: The Aquilion Serve SP (TSX-307B/1) V1.3 acquires and displays cross-sectional volumes of the whole body including the head, provides volume sets for specialized studies using indicated software or hardware, and includes AiCE deep-convolutional noise reduction for abdomen, pelvis, lung, cardiac, extremity, head, and inner-ear applications.
Modality
CT/CTA
Anatomy
Whole body including head; abdomen, pelvis, lung, cardiac, extremity, and inner-ear applications for AiCE
Authorized purpose: The Aquilion Serve acquires and displays cross-sectional volumes of the whole body including the head, provides volume sets for specialized studies using indicated software or hardware, and includes AiCE deep-convolutional noise reduction for abdomen, pelvis, lung, cardiac, extremity, head, and inner-ear applications.
Modality
CT/CTA
Anatomy
Whole body including head; abdomen, pelvis, lung, cardiac, extremity, and inner-ear applications for AiCE
Authorized purpose: An extension for syngo.via that assists trained physicians with automated and semi-automated analysis, visualization and reporting of cardiac CT angiography volume datasets for coronary-lesion evaluation, documentation, follow-up and treatment planning.
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.
Accepted repository change history
No entry has been added to this scoped change history for this record. This is not proof that the record never changed.