Authorized purpose: AI-Rad Companion Pulmonary processes CT DICOM studies to support lung and lobe segmentation, measurements, density thresholds, and pulmonary-lesion segmentation, measurement, and lobe assignment.
Clinical role
CT segmentation, measurement, quantitative analysis, and reporting support.
Images and patients
CT/CTA · Lungs, lobes, pulmonary lesions, and adult thorax
Key limitation
Keep in mind: The FDA validation scope is adult CT DICOM data with specified scanner and reconstruction conditions; lung and lesion outputs require professional review.
Clinical detailsPurpose, workflow fit, evidence, and safety record
Workflow navigation pilot
Workflow role has not yet been reviewed for this navigation pilot.
A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources
What does this software do?Authorized purpose: AI-Rad Companion Pulmonary processes CT DICOM studies to support lung and lobe segmentation, measurements, density thresholds, and pulmonary-lesion segmentation, measurement, and lobe assignment.
Which images or clinical context are in scope?Modality: CT/CTA Anatomy or clinical context: Lungs, lobes, pulmonary lesions, and adult thorax Input: Supported adult thoracic or cardiac CT DICOM studies and labeled reconstruction kernels.
Who uses it?Intended users: Trained radiologists and qualified physicians.
What does the software produce?Output: Segmentations, density values, measurements, scores, overlays, and reports.
What is its most important limitation?Key limitation: The FDA validation scope is adult CT DICOM data with specified scanner and reconstruction conditions; lung and lesion outputs require professional review.
Clinical takeaway. Clinical role: AI-Rad Companion Pulmonary processes CT DICOM studies to support lung and lobe segmentation, measurements, density thresholds, and pulmonary-lesion segmentation, measurement, and lobe assignment.
Evidence takeaway. Evidence from the reviewed source record: The exact FDA document reports software validation, bench testing, and a retrospective lung lobe segmentation study (n>4,500 CT data sets from multiple clinical sites from within and outside United States) comparing algorithm output to manually established ground truth. Average DICE coefficients for individual lung lobes ranged between 0.95 and 0.98 (SD <= 0.07); mean surface distance ranged between 0.5 and 1.0 mm; volume error was between 1.5 and 3.5 %. Results are as reported in the FDA submission, not independent validation.
Evidence context
Evidence context not reviewed for this FDA record.
Public-label constraints
Label constraints have not been reviewed for this exact FDA record. Unknown is not all ages or evidence of unsuitability.
Public-label constraints, not patient-level eligibility or a medical recommendation. Training and evaluation populations are separate. Read the original labeling and confirm the exact release and protocol.
Purpose
AI-Rad Companion Pulmonary processes CT DICOM studies to support lung and lobe segmentation, measurements, density thresholds, and pulmonary-lesion segmentation, measurement, and lobe assignment.
The FDA validation scope is adult CT DICOM data with specified scanner and reconstruction conditions; lung and lesion outputs require professional review.
The exact FDA document reports software validation, bench testing, and a retrospective lung lobe segmentation study (n>4,500 CT data sets from multiple clinical sites from within and outside United States) comparing algorithm output to manually established ground truth. Average DICE coefficients for individual lung lobes ranged between 0.95 and 0.98 (SD <= 0.07); mean surface distance ranged between 0.5 and 1.0 mm; volume error was between 1.5 and 3.5 %. Results are as reported in the FDA submission, not independent validation.
Exact FDA submission
Checked 2026-11-14
Packet wording: "Software validations, bench testing, and clinical data-based software validations have been conducted to the performance claims as well as the claim of substantial equivalence to the predicate devices." … "Performance of lung lobe segmentation of AI-Rad Companion.Pulmonary device has been validated in a retrospective performance study (n>4,500 CT data sets from multiple clinical sites from within and outside United States).". Read from the exact FDA decision document on 2026-11-14.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
The exact FDA document reports software validation, bench testing, and a retrospective lung lobe segmentation study (n>4,500 CT data sets from multiple clinical sites from within and outside United States) comparing algorithm output to manually established ground truth. Average DICE coefficients for individual lung lobes ranged between 0.95 and 0.98 (SD <= 0.07); mean surface distance ranged between 0.5 and 1.0 mm; volume error was between 1.5 and 3.5 %. Results are as reported in the FDA submission, not independent validation.
FDA source
Exact FDA submission
Checked 2026-11-14
Packet wording: "Software validations, bench testing, and clinical data-based software validations have been conducted to the performance claims as well as the claim of substantial equivalence to the predicate devices." … "Performance of lung lobe segmentation of AI-Rad Companion.Pulmonary device has been validated in a retrospective performance study (n>4,500 CT data sets from multiple clinical sites from within and outside United States).". Read from the exact FDA decision document on 2026-11-14.
Sensitivity is not promoted as a product-wide endpoint. Compare only an exact endpoint, threshold, population, version, and source if later verified.
Reported specificity
Not applicable
Not applicable
Not applicable
Checked 2026-09-04
Specificity is not promoted as a product-wide endpoint. Compare only an exact endpoint, threshold, population, version, and source if later verified.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
K183271 FDA evidence package for AI-Rad Companion (Pulmonary)
DesignFDA 510(k) or PMA decision evidence and product verification described in the exact submission.
PopulationLungs, lobes, pulmonary lesions, and adult thorax
Scopeexact submission
Tested versionAI-Rad Companion (Pulmonary) configuration described in K183271
IndependenceSponsor or vendor study
K183271
Manufacturer evidence submitted for the exact FDA decision. It is not independent clinical validation, and local technical, clinical, privacy, security, and workflow acceptance remains necessary.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K183271 by exact native submission identifier. This does not exclude product-family, later, or adverse-event information.
Public source
Exact FDA submission
Checked 2026-09-04
Exact-identifier review does not prove absence of all product-family safety signals.
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging or clinical service line and verify payer, site-of-service, coding, and coverage assumptions independently.
No separate named-product payment identified in reviewed public material
Centers for Medicare & Medicaid Services · United States
This is not a product-specific reimbursement determination. Build a local total-cost and value model and verify payer, site-of-service, coding, and coverage assumptions independently.
IT implementationDeployment, interfaces, data handling, and support
System and version relationships
Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.
No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.
Public source context, not tested interoperability. Each hop needs its own release/configuration evidence. Missing or unavailable routes are not documented incompatibility. All local tests are planned, not passed; no installed-system acceptance is recorded.
Acquisition/source → routing → AI processing → returned objects → viewer/worklist/report → support owner
No specific route context has been source-reviewed for this record. Obtain both endpoint builds and documents; this is not a finding of incompatibility.
Acquisition/source
Source context, not route validation: Supported adult thoracic or cardiac CT DICOM studies and labeled reconstruction kernels.
Scope: Exact FDA submission.
Confirm scanner/model, protocol, series completeness and both software builds.
Source context, not route validation: The FDA package describes processing of CT DICOM studies; exact local routing, archive, and output behavior remain unverified.
Scope: Exact FDA submission. This is the interface boundary described in the exact FDA package; current local compatibility, routing, identity, archive, and downtime behavior still require acceptance testing.
Name gateways, destinations, transfer rules and failure visibility; a DICOM statement does not prove a route.
Unresolved route: Not established for the exact route.
Scope: Exact FDA submission. The exact public FDA decision material does not establish the complete deployment topology, hosting responsibility, availability, upgrade path, or service contract.
Confirm processing location, queue ownership, exclusions and exact AI build.
Field checked 2026-09-04 ·
Returned objects
Source context, not route validation: Segmentations, density values, measurements, scores, overlays, and reports.
Scope: Exact FDA submission.
Confirm object classes, units, references and whether findings, overlays or measurements reach their intended destination.
Unresolved route: Segmentations, density values, measurements, scores, overlays, and reports.
Scope: Exact FDA submission.
Confirm the named viewer/worklist/report build, display behavior and report transfer. Output support alone does not establish this endpoint. The displayed source describes output, not validated end-to-end delivery.
Unresolved route: Not established for the exact route.
Scope: Scope not established.
Name the local, platform and AI supplier support owners and escalation path.
Planned local route tests
No tests below have been performed or accepted.
Patient/study identity (planned). Use approved non-patient test data to verify identifiers, accession and result-to-study linkage at every hop.
Study completeness (planned). Test missing views/series, partial transfer, late arrivals, priors and unsupported acquisition inputs; confirm explicit refusal or fallback.
Ordering and duplicates (planned). Send out-of-order, repeated and corrected studies; verify worklist order, deduplication and no misleading stale results.
Report transfer (planned). Verify output object types, annotations, units, report content, destination and availability of original images.
Access (planned). Verify authorized roles, audit trail, denied access and support ownership at each destination.
Latency and failures (planned). Agree a workload and turnaround target; test queues, timeouts, missing results and fallback reading.
Updates and rollback (planned). Define version change control, revalidation triggers, stop-use criteria and a tested rollback route.
Local implementation checklist
Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.
The FDA package describes processing of CT DICOM studies; exact local routing, archive, and output behavior remain unverified.
Exact FDA submission
Checked 2026-09-04
This is the interface boundary described in the exact FDA package; current local compatibility, routing, identity, archive, and downtime behavior still require acceptance testing.
Siemens Healthineers' corporate cybersecurity page states a Secure Development Lifecycle addressing pre-market and post-market phases for products under development and a range of existing offerings; a Cybersecurity Management Service maintaining a product security inventory (over 1,000 product versions and over 21,000 monitored software components) with security whitepapers, SBOMs and vulnerability assessments available to customers via teamplay Fleet; incident and vulnerability management channels; and that Siemens Healthineers AG maintains an independent certification of its global Cybersecurity Management System to ISO/IEC 27001:2022 extended by ISO/IEC 27701:2019, with the certificate referenced to the SGS directory. Vendor statement; no certificate or report was read.
Vendor-wide
Checked 2026-10-26
Vendor statement, not an attestation. Identity bridge: Siemens Healthineers' own corporate-info page names Siemens Medical Solutions USA, Inc. (Malvern, PA), and its 2024 governance press release states Siemens Healthcare GmbH's business was transferred to Siemens Healthineers AG. Corporate program statement; exact-release scope is not established.
Only trained radiologists and qualified physicians should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
Monitor K183271 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: The FDA validation scope is adult CT DICOM data with specified scanner and reconstruction conditions; lung and lesion outputs require professional review.
Exact FDA submission
Checked 2026-09-04
Monitoring guidance is buyer-facing implementation guidance derived from the exact cleared scope; it is not a claim that the FDA package provides a live monitoring service.
The exact public deployment topology remains incomplete; establish hosting responsibility, data flow, availability, upgrade, service, and downtime boundaries locally.
exact submission · checked 2026-09-04
FDA-described deployment boundaries do not establish current hosting contracts, security controls, uptime, or local qualification.
Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. The FDA validation scope is adult CT DICOM data with specified scanner and reconstruction conditions; lung and lesion outputs require professional review.
exact submission · checked 2026-09-04
Define stop-use, escalation, rollback, and revalidation triggers before production use.
Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.
product family · checked 2026-09-04
No exact-product price was found in the reviewed public material; productivity or value claims require local evidence.
Monitor K183271 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: The FDA validation scope is adult CT DICOM data with specified scanner and reconstruction conditions; lung and lesion outputs require professional review.
exact submission · checked 2026-09-04
Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. thoracic CT analysis software limits remain submission-specific.
Sources & review detailsFDA identity, research record, source ledger, and history
Regulatory identity
FDA record and catalog context
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
This audit distinguishes completed source review from fields that have not yet been researched.
Human reviewedStatus
2026-09-20Last searched
27Fields reviewed
13Source classes checked
20Screened PubMed leads
1Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains 20 PubMed and 1 ClinicalTrials.gov candidate records. Discovery metadata alone does not establish exact tested-product, version, population, and endpoint applicability, so all candidates remain quarantined and none was promoted. The current discovery snapshot contains 20 PubMed and 1 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, acr ai central product, fda decision material, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall
Field review details
Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.
Public pricing: Not publicly available - checked 2026-09-04 - Exact FDA submission
No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.
Pricing model: Not yet reviewed
Setup and integration fees: Not yet reviewed
Recurring fees: Not yet reviewed
Minimum commitment: Not yet reviewed
Contract duration: Not yet reviewed
Renewal terms: Not yet reviewed
Cancellation terms: Not yet reviewed
Support and service terms: Not yet reviewed
Trial or pilot terms: Not yet reviewed
Deployment: Not publicly available - checked 2026-09-04 - Exact FDA submission
The exact public FDA decision material does not establish the complete deployment topology, hosting responsibility, availability, upgrade path, or service contract.
Authorized purpose: Auto-Seg and Spine Auto-Seg accept spine and pelvis CT images and provide three-dimensional segmentation and labels to support spine-surgery planning and assessment by spine surgeons.
Authorized purpose: Acorn 3D Software and Acorn 3DP Model process CT, synthetic CT, and CT angiography for segmentation, visualization, measurement, treatment planning, and three-dimensional printing, including adult and pediatric spine trajectory planning.
Modality
CT/CTA
Anatomy
Spine; craniomaxillofacial; musculoskeletal anatomy; and patient-specific printed models
Copy this public context and add the field, proposed wording and original source before sending. Nothing here is submitted automatically. Form-specific prefill has not been verified.
Permission to display a submission is not evidence verification. A submission enters editorial review; it does not approve a public change. Form availability, data entry, receipt, review and publication are separate steps. Data entry and receipt remain unverified.
Accepted repository change history
No entry has been added to this scoped change history for this record. This is not proof that the record never changed.